Perminova Respiratory Effort Accuracy Validation 09-12-2013-Rev 2
Study Overview
Status
Conditions
Detailed Description
After IRB Approval, eighteen volunteer test subjects will be entered into an accuracy study that is designed to validate accuracy of the respiratory rate technology of the Perminova N4 Monitoring system (Device Under Test). The subjects will be selected to represent a range of body types including small, average, muscular, and large with a range of BMI's.
A range of stable respiratory rates will be elicited from each volunteer test subject. The rates will be approximately 5, 10, 15, 20, 25, 30, 35 and 40 breaths per minute; with some natural variation from these exact numbers.
Each subject will be instrumented with an open system mask that allows for measurement of the end tidal carbon dioxide (EtCO2) respiratory rate and tidal volume. The EtCO2 monitor will be used as the Accuracy Reference Device (Reference) in this study. Performance of the currently released ECG Impedance Respiratory Rate monitor (Acceptance Standard) will be used to establish the acceptance criteria in this study by comparing to Reference.
Study Type
Enrollment (Actual)
Contacts and Locations
Study Locations
-
-
Colorado
-
Louisville, Colorado, United States, 80027
- Clinimark Laboratory
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Genders Eligible for Study
Sampling Method
Study Population
Description
Inclusion Criteria:
- male
- female
- 18-45
- range of physiques
- healthy
Exclusion Criteria:
- injury to sensor area
- skin inflammation
- implanted devices such as pacemakers
- pregnant
Study Plan
How is the study designed?
Design Details
Cohorts and Interventions
Group / Cohort |
|---|
|
male female 18-45 years range of body types
18 male female subject mix that are 18-45 years of age and selected to represent a range of body types of varying physiques and body mass indexes
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Time Frame |
|---|---|
|
Respiratory Rate of test subject as compared to end tidal CO2
Time Frame: 1 week
|
1 week
|
Collaborators and Investigators
Sponsor
Study record dates
Study Major Dates
Study Start
Primary Completion (Actual)
Study Completion (Actual)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Estimate)
Study Record Updates
Last Update Posted (Estimate)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Keywords
Other Study ID Numbers
- 78o944o
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Respiratory Rate Accuracy Test
-
Eupnea ASHaraldsplass Deaconess HospitalCompletedRespiratory Rate | Capnography | Wireless Vital Signs Monitoring System | Respiratory Rate Accuracy TestNorway
-
Gabi SmartCareCompleted
-
Binah.ai LTD.Clinimark, LLCCompletedRespiratory and Pulse Rate Accuracy ValidationUnited States
-
ResMedCompletedAccuracy of Respiration Rate Measurement
-
ResMedCompletedAccuracy of Measurement of Respiration Rate
-
Life Meter srlCompletedLaboratory Testing on Healthy Volunteers to Verify SpO2 and Heart Rate (HR) Accuracy of BrOxy MItaly
-
RTM Vital Signs, LLCRecruitingRespiratory Rate | Pulmonary Function Test | Tidal Volume | Ventilators | Breathing SoundsUnited States
-
Fifth Affiliated Hospital, Sun Yat-Sen UniversitySun Yat-sen UniversityUnknownDetection Rate of SLN ; Accuracy Rate of Lymph Node MetastasisChina
-
National Taiwan University HospitalRecruitingColonoscopy | Adenoma Detection Rate | Fecal Immunochemical Test | Three DimensionalTaiwan
-
Cairo UniversityUnknownInaccurate Transfer and Bond Failure of Conventional Indirect Bonding Tray | Accuracy of Bracket Transfer and Bond Failure Rate of Full Arch and Segmented Digitally Printed Trays