Comparison of Pulse Wave and Blood Pressure Measurements Before, During and After Ergometry in Healthy Women

Comparison of Pulse Wave and Blood Pressure Parameters Before, During and After Exercise Ergometry in Healthy Female Subjects - a Randomized, Controlled, Single-blind Study

Sponsors

Lead Sponsor: ARCIM Institute Academic Research in Complementary and Integrative Medicine

Collaborator: Universität Tübingen

Source ARCIM Institute Academic Research in Complementary and Integrative Medicine
Brief Summary

The objective of this randomized, controlled single blind study is to compare pulse wave and blood pressure measurements before, during and after exercise ergometry in healthy female subjects.

Detailed Description

In a randomized single-blind study design the investigators explore how well blood pressure parameters calculated using a Pulse Transit Time-based method (device: SOMNOtouchTM RESP) match blood pressure parameters measured by a classic upper arm blood pressure cuff (BOSO-TM-2430 ) under standardized load conditions (ergometry). In 30 healthy female subjects (age 18 - 40 years), cardiovascular parameters before, during and after a standardized load on a bicycle ergometer (half-lying position) are recorded with three different devices (BOSO-TM-2430 versus SOMNOtouchTM RESP versus GeTeMed Vitaguard 3100) and compared. Primary outcome is the difference of the systolic blood pressure values between the devices BOSO-TM-2430 versus SOMNOtouchTM RESP during the three phases. The BOSO-TM-2430 blood pressure cuff is applied randomized to the arm with the previously determined higher or lower systolic blood pressure mean difference. The subjects are blinded to the results of the measurements during ergometry. After a pre-rest period of 12 minutes, the subjects are physically strained on the ergometer for 48 minutes (6 minutes loading phase followed by a 2-minute rest phase; 30 watts increase every 8 minutes up to a maximum load of 180 watts), followed by a 12-minute rest period. Measurements are taken every two minutes with the BOSO-TM-2430; data acquisition with the SOMNOtouchTM RESP and GeTeMed Vitaguard 3100 devices takes place continuously over a total duration of 72 minutes.

Overall Status Completed
Start Date 2019-01-29
Completion Date 2019-05-17
Primary Completion Date 2019-05-17
Phase N/A
Study Type Interventional
Primary Outcome
Measure Time Frame
Difference of the systolic blood pressure Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Secondary Outcome
Measure Time Frame
Diastolic Blood Pressure (SOMNOtouchTM RESP) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Mean Arterial Blood Pressure (SOMNOtouchTM RESP) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Pulse Pressure (SOMNOtouchTM RESP) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Heartrate (SOMNOtouchTM RESP) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Oxygen saturation (SpO2, SOMNOtouchTM RESP) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Pulse-Transit-Time (PTT, SOMNOtouchTM RESP) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Systolic Blood Pressure measured with the device BOSO-TM-2430 Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Diastolic Blood Pressure measured with the device BOSO-TM-2430 Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Mean Arterial Blood Pressure measured with the device BOSO-TM-2430 Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Pulse Pressure measured with the device BOSO-TM-2430 Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Heartrate measured with the device BOSO-TM-2430 Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Pulse Transit Time (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Perfusion Index (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Peak to Peak Time (PPT) (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
SlopeInW1 (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
SlopeInW2 (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
SlopeInW3 (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
SlopeInW4 (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
StiffnessIndex (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
ReflectionIndex (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
TimeTX (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
TimeTY (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
QuotTYTX (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
AreaAX (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
AreaAY (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
QuotAYAX (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
AreaAV (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
AreaAW (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
QuotAWAV (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
QuotTVTW (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
TimeTV (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
TimeTW (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
MinT1 (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
MinT2 (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
MaxT1 (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
MaxT2 (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
QuotHRRespRate (GeTeMed Vitaguard 3100) Physiological parameters are measured continuously (SOMNOtouchTM RESP) and every 2 minutes (BOSO-TM-2430) for a total duration of 72 minutes
Enrollment 30
Condition
Intervention

Intervention Type: Other

Intervention Name: Ergometry H

Description: After a pre-rest period of 12 minutes, the subjects are physically strained on the ergometer for 48 minutes (6 minutes loading phase followed by a 2-minute rest phase; 30 watts increase every 8 minutes up to a maximum load of 180 watts), followed by a 12-minute rest periodmaximum load of 180 watts), followed by a 15-minute post-rest period. BOSO-TM-2430 blood pressure cuff is applied to the upper arm with the previously determined higher systolic blood pressure (reference parameter). SOMNOtouchTM RESP and GeTeMed Vitaguard 3100 are accordingly applied to the arm with the previously determined lower systolic blood pressure.

Arm Group Label: Higher Blood Pressure

Intervention Type: Other

Intervention Name: Ergometry L

Description: After a pre-rest period of 12 minutes, the subjects are physically strained on the ergometer for 48 minutes (6 minutes loading phase followed by a 2-minute rest phase; 30 watts increase every 8 minutes up to a maximum load of 180 watts), followed by a 12-minute rest periodmaximum load of 180 watts), followed by a 15-minute post-rest period. BOSO-TM-2430 blood pressure cuff is applied to the upper arm with the previously determined lower systolic blood pressure (reference parameter). SOMNOtouchTM RESP and GeTeMed Vitaguard 3100 are accordingly applied to the arm with the previously determined higher systolic blood pressure.

Arm Group Label: Lower Blood Pressure

Eligibility

Criteria:

Inclusion Criteria: - Signed informed consent - Healthy at the time of the examinations - Inconspicuous anamnesis regarding cardiovascular diseases and bronchial asthma - resting heart rate 50 to 100 beats per minute - Normal blood pressure or appropriately treated and controlled hypertension (systolic blood pressure ≤ 140 mmHg and diastolic blood pressure ≤ 90 mmHg) Exclusion Criteria: - Acute or chronic (especially cardiovascular) diseases, including cardiac arrhythmia and vegetative regulatory disorders (recorded via anamnesis) - Taking allopathic medication: digitalis, beta blockers, antiarrhythmics - Bronchial asthma (recorded by anamnesis) - Alcohol abuse - Heart diseases NYHA class IV - Blood pressure side difference ≥ 10mmHg - Fever (> 37,5°C)

Gender:

Female

Minimum Age:

18 Years

Maximum Age:

40 Years

Healthy Volunteers:

Accepts Healthy Volunteers

Overall Official
Last Name Role Affiliation
Jan Vagedes, MD Study Director ARCIM Institute, Im Haberschlai 7, 70794 Filderstadt, Germany
Location
Facility: ARCIM Institute
Location Countries

Germany

Verification Date

2019-08-01

Responsible Party

Type: Sponsor

Keywords
Has Expanded Access No
Number Of Arms 2
Arm Group

Label: Higher Blood Pressure

Type: Experimental

Description: BOSO-TM-2430 blood pressure cuff is applied to the upper arm with the previously determined higher systolic blood pressure. Intervention: Ergometry H

Label: Lower Blood Pressure

Type: Experimental

Description: BOSO-TM-2430 blood pressure cuff is applied to the upper arm with the previously determined lower systolic blood pressure. Intervention: Ergometry L

Study Design Info

Allocation: Randomized

Intervention Model: Parallel Assignment

Primary Purpose: Basic Science

Masking: Single (Participant)

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact [email protected]. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.