- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT05297500
Determination of SpO2 and PR Accuracy Specifications at Rest (71Ag_Vital-0031)
Accuracy of Pulse Oximeters With Profound Hypoxia NIHO 14
Study Overview
Detailed Description
Study Type
Enrollment (Actual)
Contacts and Locations
Study Locations
-
-
California
-
San Francisco, California, United States, 94133
- UCSF Hypoxia Research Laboratory
-
-
Participation Criteria
Eligibility Criteria
Ages Eligible for Study
Accepts Healthy Volunteers
Sampling Method
Study Population
Description
Inclusion Criteria:
- Both male and female subjects who can give written informed consent
- Healthy subjects capable of undergoing controlled hypoxemia to the levels outlined in the desaturation profile
- Meeting the demographic requirements
Exclusion Criteria:
- Pregnant women
- Significant arrhythmia
- Blood pressure above 150 systolic or 90 diastolic
- Carboxyhemoglobin levels over 3%
- Subjects whom the investigator consider ineligible for the study
Study Plan
How is the study designed?
Design Details
Cohorts and Interventions
Group / Cohort |
Intervention / Treatment |
|---|---|
|
Adult healthy subjects
Adult healthy subjects capable of undergoing controlled hypoxemia to the levels outlined in the desaturation profile in an at rest state
|
OLV-4202 pulse oximeter (SW version: 01-12)
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Root-mean-square (Rms) of device SpO2 measurement with room air stabilization is within +-2% [%SpO2] of reference co-oximeter SaO2 measurement at varying desaturation levels (70% to 100%)
Time Frame: 30 minutes
|
Each subject will have an arterial catheter inserted in the left or right wrist and the pulse oximeter study device placed on different fingers of the left or right hand, or the left or right earlobe or the forehead. Measurement position is randomized across subjects. For each subject, a series of desaturation runs are performed. Run 1 stabilized room air period -> stabilized plateaus at saturation level targets: room air, 92%, 87%, 82%, 77%, 72% Run 2 stabilized 100% O2 period -> stabilized plateaus at saturation level targets: 100%, 93%, 88%, 83%, 78%, 73% SpO2 is observed breath by breath arterial saturation, which is in turn computed from end tidal PO2 and PCO2. FiO2 is reduced suddenly at first then adjusted to achieve a stable plateau value at the desired target. Each desaturation plateaus should not exceed 10 minutes. A blood draw is taken at stable FiO2 and +30 seconds. The gas mixture is then changed for the next value and repeated for each plateau value. |
30 minutes
|
|
Root-mean-square (Rms) of device Pulse Rate measurement with room air stabilization is within +-3 [1/min] of reference ECG Pulse Rate measurement at heart rates (30 to 300 [1/min])
Time Frame: 30 minutes
|
Each subject will have an arterial catheter inserted in the left or right wrist and the pulse oximeter study device placed on different fingers of the left or right hand, or the left or right earlobe or the forehead. Measurement position is randomized across subjects. For each subject, a series of desaturation runs are performed. Run 1 stabilized room air period -> stabilized plateaus at saturation level targets: room air, 92%, 87%, 82%, 77%, 72% Run 2 stabilized 100% O2 period -> stabilized plateaus at saturation level targets: 100%, 93%, 88%, 83%, 78%, 73% Pulse rate is recorded from the test device and from the reference ECG monitor. PR data measurements are gathered during the stable plateau value at the desired saturation target. Each desaturation plateaus should not exceed 10 minutes. The gas mixture is then changed for the next value and repeated for each plateau value. |
30 minutes
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Root-mean-square (Rms) of device SpO2 measurement with room air stabilization is within +-2% [%SpO2] of reference co-oximeter SaO2 measurement at desaturation levels (80% to 100%) and within +-3% [%SpO2] at desaturation levels (70% to 80%)
Time Frame: 30 minutes
|
Each subject will have an arterial catheter inserted in the left or right wrist and the pulse oximeter study device placed on different fingers of the left or right hand, or the left or right earlobe or the forehead. Measurement position is randomized across subjects. For each subject, a series of desaturation runs are performed. Run 1 stabilized room air period -> stabilized plateaus at saturation level targets: room air, 92%, 87%, 82%, 77%, 72% Run 2 stabilized 100% O2 period -> stabilized plateaus at saturation level targets: 100%, 93%, 88%, 83%, 78%, 73% SpO2 is observed breath by breath arterial saturation, which is in turn computed from end tidal PO2 and PCO2. FiO2 is reduced suddenly at first then adjusted to achieve a stable plateau value at the desired target. Each desaturation plateaus should not exceed 10 minutes. A blood draw is taken at stable FiO2 and +30 seconds. The gas mixture is then changed for the next value and repeated for each plateau value. |
30 minutes
|
Collaborators and Investigators
Sponsor
Collaborators
Study record dates
Study Major Dates
Study Start (Actual)
Primary Completion (Actual)
Study Completion (Estimated)
Study Registration Dates
First Submitted
First Submitted That Met QC Criteria
First Posted (Actual)
Study Record Updates
Last Update Posted (Actual)
Last Update Submitted That Met QC Criteria
Last Verified
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- 71Ag_Vital-0031
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
IPD Plan Description
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
Studies a U.S. FDA-regulated device product
product manufactured in and exported from the U.S.
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.