The concept of a cementless isoelastic monoblock cup made of highly cross-linked polyethylene infused with vitamin E: radiological analyses of migration and wear using EBRA and clinical outcomes at mid-term follow-up

Yama Afghanyar, Sebastian Joser, Jonas Tecle, Philipp Drees, Jens Dargel, Philipp Rehbein, Karl Philipp Kutzner, Yama Afghanyar, Sebastian Joser, Jonas Tecle, Philipp Drees, Jens Dargel, Philipp Rehbein, Karl Philipp Kutzner

Abstract

Background: The newest generation of cementless titanium-coated, isoelastic monoblock cup with vitamin E-blended highly cross-linked polyethylene (HXLPE) was introduced to the market in 2009. The aim of the present study was to obtain mid-term follow-up data including migration and wear analyses.

Methods: This prospective study investigated 101 primary total hip arthroplasty (THA) cases in 96 patients treated at a single institution. Patients were allowed full weight-bearing on the first day postoperatively. Harris hip score (HHS) and pain and satisfication on a visual analogue scale (VAS) were assessed at a mean follow-up of 79.0 months. Migration and wear were assessed using Einzel-Bild-Roentgen-Analyse (EBRA) software. Radiological acetabular bone alterations and complications were documented.

Results: At mid-term follow-up (mean 79.0 months, range: 51.8-101.7), 81 cases with complete clinical and radiological data were analyzed. Utilisable EBRA measurements were obtained for 42 hips. The mean HHS was 91.1 (range 38.0-100.0), VAS satisfaction was 9.6 (range 6.0-10.0), VAS rest pain was 0.2 (range 0.0-4.0), and VAS load pain was 0.6 (range 0.0-9.0). Mean migration was 0.86 mm (range: 0.0-2.56) at 24 months and 1.34 mm (range: 0.09-3.14) at 5 years, and the mean annual migration rate was 0.22 (range: - 0.24-1.34). The mean total wear was 0.4 mm (range: 0.03-1.0), corresponding to a mean annual wear rate of 0.06 mm per year (range: 0.0-0.17). Radiographic analysis did not reveal any cases of osteolysis, and no revision surgeries had to be performed.

Conclusions: After using vitamin-E blended HXLPE in cementless isoelastic monoblock cups, there were no obvious signs of osteolysis or aseptic loosening occurred. No patients required revision surgery after mid-term follow-up. Cup migration and wear values were well below the benchmarks considered predictive for potential future failure.

Trial registration: The trial registration number on ClinicalTrials.gov : NCT04322916 (retrospectively registered at 26.03.2020).

Keywords: Cementless acetabular cup; EBRA; Migration; Total hip arthroplasty; Vitamys; Wear.

Conflict of interest statement

KPK and PR serve as instructors for Mathys Ltd., Bettlach, Switzerland. PR also serves as medical advisor. All other authors declare that they have no competing interests.

Figures

Fig. 1
Fig. 1
a RM Classic cup; b RM Pressfit cup; c RM Pressfit vitamys cup. (Mathys Ltd., Bettlach, Switzerland)
Fig. 2
Fig. 2
Radiographs of an RM Pressfit vitamys cup (Mathys Ltd. Bettlach, Switzerland) immediately after surgery (a) and at 5-year follow-up (b). The patient showed signs of a sclerotic line in Zones 1 and 2
Fig. 3
Fig. 3
Study flow diagram

References

    1. Learmonth ID, Young C, Rorabeck C. The operation of the century: total hip replacement. Lancet. 2007;370:1508–1519. doi: 10.1016/S0140-6736(07)60457-7.
    1. Rochcongar G, Buia G, Bourroux E, Dunet J, Chapus V, Hulet C. Creep and Wear in vitamin E-infused highly cross-linked polyethylene cups for Total hip Arthroplasty: a prospective randomized controlled trial. J Bone Joint Surg Am. 2018;100(2):107–114. doi: 10.2106/JBJS.16.01379.
    1. Callaghan JJ, Liu SS, Firestone DE, Yehyawi TM, Goetz DD, Sullivan J, et al. Total hip arthroplasty with cement and use of a collared matte-finish femoral component: nineteen to twenty-year follow-up. J Bone Joint Surg Am. 2008;90:299–306. doi: 10.2106/JBJS.G.00095.
    1. Ihle M, Mai S, Pfluger D, Siebert W. The results of the titanium-coated RM acetabular component at 20 years: a term follow-up of an uncemented primary total hip replacement. J Bone Joint Surg (Br) 2008;90(1):1284–1290. doi: 10.1302/0301-620X.90B10.20274.
    1. Wyatt M, Weidner J, Pfluger D, Beck M. The RM Pressfit vitamys: 5-year Swiss experience of the first 100 cups. Hip Int. 2017;27:368–372. doi: 10.5301/hipint.5000469.
    1. Wyss T, Kägi P, Mayrhofer P, Nötzli H, Pfluger D, Knahr K. Five-year results of the Uncemented RM Pressfit cup clinical evaluation and migration measurements by EBRA. J Arthroplast. 2013;28:1291–1296. doi: 10.1016/j.arth.2012.11.004.
    1. Lafon L, Moubarak H, Druon J, Rosset P. Cementless RM Pressfit® cup. A clinical and radiological study of 91 cases with at least four years follow-up. Orthop Traumatol Surg Res. 2014;100:S225–S229. doi: 10.1016/j.otsr.2014.03.007.
    1. Pakvis D, Biemond L, van Hellemondt G, Spruit M. A cementless elastic monoblock socket in young patients: a ten to 18-year clinical and radiological follow-up. Int Orthop. 2011;35:1445–1451. doi: 10.1007/s00264-010-1120-7.
    1. Hooper N, Sargeant H, Frampton C, Hooper G. Does a titanium-coated polyethylene press-fit cup give reliable midterm results? Clin Orthop Relat Res. 2015;473:3806–3810. doi: 10.1007/s11999-015-4556-7.
    1. Erivan R, Eymond G, Villatte G, Mulliez A, Myriam G, Descamps S, et al. RM Pressfit® cup: good preliminary results at 5 to 8 years follow-up for 189 patients. Hip Int. 2016;26:386–391. doi: 10.5301/hipint.5000359.
    1. Beck M, Delfosse D, Lerf R, Becker R, French G. Oxidation prevention with vitamin E in a HXLPE isoelastic monoblock pressfit cup: preliminary results. In: Knahr K, editor. Total hip arthroplasty. Heidelberg: Springer; 2012. pp. 21–31.
    1. Oral E, Muratoglu OK. Vitamin E diffused, highly crosslinked UHMWPE: a review. Int Orthop. 2011;35:215–223. doi: 10.1007/s00264-010-1161-y.
    1. Oral E, Wannomae KK, Hawkins N, Harris WH, Muratoglu OK. α-Tocopherol-doped irradiated UHMWPE for high fatigue resistance and low wear. Biomaterials. 2004;25:5515–5522. doi: 10.1016/j.biomaterials.2003.12.048.
    1. Scemama C, Anract P, Dumaine V, Babinet A, Courpied JP, Hamadouche M. Does vitamin E-blended polyethylene reduce wear in primary total hip arthroplasty: a blinded randomised clinical trial. Int Orthop. 2017;41:1113–1118. doi: 10.1007/s00264-016-3320-2.
    1. Lerf R, Zurbrügg D, Delfosse D. Use of vitamin E to protect cross-linked UHMWPE from oxidation. Biomaterials. 2010;31:3643–3648. doi: 10.1016/j.biomaterials.2010.01.076.
    1. Bracco P, Oral E. Vitamin E-stabilized UHMWPE for total joint implants: a review. Clin Orthop Relat Res. 2011;469:2286–2293. doi: 10.1007/s11999-010-1717-6.
    1. Halma J, Senaris J, Delfosse D, Lerf R, Oberbach T, van Gaalen S, et al. Edge loading does not increase wear rates of ceramic-on-ceramic and metal-on-polyethylene articulations. J Biomed Mater Res B Appl Biomater. 2014;102:1627–1638. doi: 10.1002/jbm.b.33147.
    1. Halma J, Eshuis R, Vogely HC, van Gaalen S, de Gast A. An uncemented iso-elastic monoblock acetabular component: preliminary results. J Arthroplast. 2015;30:615–621. doi: 10.1016/j.arth.2014.11.017.
    1. Dumbleton JH, Manley MT, Edidin AA. A literature review of the association between wear rate and osteolysis in total hip arthroplasty. J Arthroplast. 2002;17:649–661. doi: 10.1054/arth.2002.33664.
    1. Krismer M, Bauer R, Tschupik J, Mayrhofer P. EBRA: a method to measure migration of acetabular components. J Biomech. 1995;28:1225–1236. doi: 10.1016/0021-9290(94)00177-6.
    1. Ilchmann T, Markovic L, Joshi A, Hardinge K, Murphy J, Wingstrand H. Migration and wear of long-term successful Charnley total hip replacements. J Bone Joint Surg (Br) 1998;80:377–381. doi: 10.1302/0301-620X.80B3.0800377.
    1. Kostakos AT, Macheras GA, Frangakis CE, Stafilas KS, Baltas D, Xenakis TA. Migration of the trabecular metal monoblock acetabular cup system. J Arthroplast. 2010;25:35–40. doi: 10.1016/j.arth.2008.09.027.
    1. Stocks G, Freeman M, Evans S. Acetabular cup migration. Prediction of aseptic loosening. Bone Joint Surg Br. 1995;77:853–861. doi: 10.1302/0301-620X.77B6.7593094.
    1. Wilkinson J, Gordon A, Stockley I. Experiences with the Plasmacup--early stability, wear, remodelling, and outcome. Int Orthop. 2003;27:S16–S19.
    1. Wroblewski B, Siney P, Fleming P. The principle of low frictional torque in the Charnley total hip replacement. Bone Joint Surg Br. 2009;91:855–858. doi: 10.1302/0301-620X.91B7.22027.
    1. Pfeil J, Siebert W, editors. Minimally invasive surgery in Total hip Arthroplasty. Heidelberg: Springer Verlag; 2010.
    1. Engh CA, Jr, Claus AM, Hopper RH, Jr, Engh CA. Long-term results using the anatomic medullary locking hip prosthesis. Clin Orthop Relat Res (1976–2007) 2001;393:137–146. doi: 10.1097/00003086-200112000-00016.
    1. DeLee J, Charnley J. Radiological demarcation of cemented sockets in hip arthroplasty. Clin Orthop Relat Res. 1976;121:20–32.
    1. Brooker AF, Bowerman JW, Robinson RA, Riley LH., Jr Ectopic ossification following total hip replacement: incidence and a method of classification. J Bone Joint Surg Am. 1973;55:1629–1632. doi: 10.2106/00004623-197355080-00006.
    1. Wilkinson J, Hamer A, Elson R, Stockley I, Eastell R. Precision of EBRA-digital software for monitoring implant migration after total hip arthroplasty. J Arthroplast. 2002;17:910–916. doi: 10.1054/arth.2002.34530.
    1. Stoeckl B, Brabec E, Wanner S, Krismer M, Biedermann R. Radiographic evaluation of the Duraloc cup after 4 years. Int Orthop. 2005;29:14–17. doi: 10.1007/s00264-004-0600-z.
    1. Ilchmann T, Mjöberg B, Wingstrand H. Measurement accuracy in acetabular cup wear: three retrospective methods compared with roentgen stereophotogrammetry. J Arthroplast. 1995;10:636–642. doi: 10.1016/S0883-5403(05)80208-6.
    1. McCalden RW, Naudie DD, Yuan X, Bourne RB. Radiographic methods for the assessment of polyethylene wear after total hip arthroplasty. J Bone Joint Surg Am. 2005;87:2323–2334.
    1. Krismer M, Stöckl B, Fischer M, Bauer R, Mayrhofer P, Ogon M. Early migration predicts late aseptic failure of hip sockets. J Bone Joint Surg (Br) 1996;78:422–426. doi: 10.1302/0301-620X.78B3.0780422.
    1. Kurtz S, Dumbleton J, Siskey R, Wang J, Manley M. Trace concentrations of vitamin E protect radiation crosslinked UHMWPE from oxidative degradation. J Biomed Mater Res A. 2009;90:549–563. doi: 10.1002/jbm.a.32122.
    1. van Erp JHJ, Massier JRA, Halma JJ, Snijders TE, de Gast A. 2-year results of an RCT of 2 uncemented isoelastic monoblock acetabular components: lower wear rate with vitamin E blended highly cross-linked polyethylene compared to ultra-high molecular weight polyethylene. Acta Orthop. 2020;91:254–259. doi: 10.1080/17453674.2020.1730073.
    1. Massier JRA, Van Erp JHJ, Snijders TE, ADE G. A vitamin E blended highly cross-linked polyethylene acetabular cup results in less wear: 6-year results of a randomized controlled trial in 199 patients. Acta Orthop. 2020. Online ahead of print. 10.1080/17453674.2020.1807220.

Source: PubMed

3
Subscribe