Utility and feasibility of integrating pulse oximetry into the routine assessment of young infants at primary care clinics in Karachi, Pakistan: a cross-sectional study

Connor A Emdin, Fatima Mir, Shazia Sultana, A M Kazi, Anita K M Zaidi, Michelle C Dimitris, Daniel E Roth, Connor A Emdin, Fatima Mir, Shazia Sultana, A M Kazi, Anita K M Zaidi, Michelle C Dimitris, Daniel E Roth

Abstract

Background: Hypoxemia may occur in young infants with severe acute illnesses or congenital cardiac anomalies, but is not reliably detected on physical exam. Pulse oximetry (PO) can be used to detect hypoxemia, but its application in low-income countries has been limited, and its feasibility in the routine assessment of young infants (aged 0-59 days) has not been previously studied. The aim of this study was to characterize the operational feasibility and parent/guardian acceptability of incorporating PO into the routine clinical assessment of young infants in a primary care setting in a low-income country.

Methods: This was a cross-sectional study of 862 visits by 529 infants at two primary care clinics in Karachi, Pakistan (March to June, 2013). After clinical assessment, oxygen saturation (Sp02) was measured by a handheld PO device (Rad-5v, Masimo Corporation) according to a standardized protocol. Performance time (PT) was the time between sensor placement and attainment of an acceptable PO reading (i.e., stable SpO2 + 1% for at least 10 s, heart rate displayed, and adequate signal indicators). PT included the time for one repeat attempt at a different anatomical site if the first attempt did not yield an acceptable reading within 1 min. Parent/guardian acceptability of PO was based on a questionnaire and unprompted comments about the procedure. All infants underwent physician assessment.

Results: Acceptable PO readings were obtained in ≤ 1 and ≤ 5 min at 94.4% and 99.8% of visits, respectively (n = 862). Median PT was 42 s (interquartile range 37; 50). Parents/guardians overwhelmingly accepted PO (99.6% overall satisfaction, n = 528 first visits). Of 10 infants with at least one visit with Sp02 <92% on a first PO attempt, 3 did not have a significant acute illness on physician assessment. There were no PO-related adverse events.

Discussion: Using a commercially available handheld pulse oximeter, acceptable Sp02 measurements were obtained in nearly all infants in under 1 minute. The procedure was readily integrated into existing assessment pathways and parents/guardians had positive views of the technology.

Conclusions: When incorporated into routine clinical assessment of young infants at primary care clinics in a low-income country, PO was feasible and acceptable to parents/guardians. Future research is needed to determine if the introduction of routine PO screening of young infants will improve outcomes in low-resource settings.

Figures

Fig. 1
Fig. 1
Cumulative proportion of participants with acceptable oxygen saturation (Sp02) values over time, among young infants undergoing routine pulse oximetry at primary health centers in Karachi, Pakistan. Time shown represents pulse oximetry “performance time” (see text for definition)
Fig. 2
Fig. 2
Proportion of infants with measured Sp02 at initial study visits. One infant without an acceptable Sp02 measurement at the initial study visit was excluded from this analysis (n = 528)

References

    1. Zaidi AKM, Ganatra HA, Syed S, Cousens S, Lee ACC, Black R, et al. Effect of case management on neonatal mortality due to sepsis and pneumonia. BMC Public Health. 2011;11(Suppl 3):S13. doi: 10.1186/1471-2458-11-S3-S13.
    1. Darmstadt GL, El Arifeen S, Choi Y, Bari S, Rahman SM, Mannan I, et al. Household surveillance of severe neonatal illness by community health workers in Mirzapur, Bangladesh: coverage and compliance with referral. Health Policy Plan. 2010;25:112–124. doi: 10.1093/heapol/czp048.
    1. Young Infants Clinical Signs Study Group Clinical signs that predict severe illness in children under age 2 months: a multicentre study. Lancet. 2008;371:135–142. doi: 10.1016/S0140-6736(08)60106-3.
    1. Duke T, Blaschke AJ, Sialis S, Bonkowsky JL. Hypoxaemia in acute respiratory and non-respiratory illnesses in neonates and children in a developing country. Arch Dis Child. 2002;86:108–112. doi: 10.1136/adc.86.2.108.
    1. Duke T, Subhi R, Peel D, Frey B. Pulse oximetry: technology to reduce child mortality in developing countries. Ann Trop Paediatr. 2009;29:165–175. doi: 10.1179/027249309X12467994190011.
    1. Mower WR, Sachs C, Nicklin EL, Baraff LJ. Pulse oximetry as a fifth pediatric vital sign. Pediatrics. 1997;99:681–686. doi: 10.1542/peds.99.5.681.
    1. Salyer JW. Neonatal and pediatric pulse oximetry. Respir Care. 2003;48:386–96.
    1. Choi SJ, Ahn HJ, Yang MK, Kim CS, Sim WS, Kim JA, et al. Comparison of desaturation and resaturation response times between transmission and reflectance pulse oximeters. Acta Anaesthesiol Scand. 2010;54:212–217. doi: 10.1111/j.1399-6576.2009.02101.x.
    1. Neuman MI, Monuteaux MC, Scully KJ, Bachur RG. Prediction of pneumonia in a pediatric emergency department. Pediatrics. 2011;128:246–253. doi: 10.1542/peds.2010-3367.
    1. Thangaratinam S, Brown K, Zamora J, Khan KS, Ewer AK. Pulse oximetry screening for critical congenital heart defects in asymptomatic newborn babies: a systematic review and meta-analysis. Lancet. 2012;379:2459–2464. doi: 10.1016/S0140-6736(12)60107-X.
    1. Duke T, Wandi F, Jonathan M, Matai S, Kaupa M, Saavu M, et al. Improved oxygen systems for childhood pneumonia: a multihospital effectiveness study in Papua New Guinea. Lancet. 2008;372:1328–1333. doi: 10.1016/S0140-6736(08)61164-2.
    1. Gove S. Integrated management of childhood illness by outpatient health workers: technical basis and overview. The WHO Working Group on Guidelines for Integrated Management of the Sick Child. Bull World Health Organ. 1997;75(1):7–24.
    1. Louis D, Sundaram V, Kumar P. Pulse oximeter sensor application during neonatal resuscitation: a randomized controlled trial. Pediatrics. 2014;133:476–482. doi: 10.1542/peds.2013-2175.
    1. Pocket Book of Hospital Care for Children . Guidelines for the Management of Common Childhood Illnesses. 2. Geneva: World Health Organization; 2013.
    1. Maneker AJ, Petrack EM, Krug SE. Contribution of routine pulse oximetry to evaluation and management of patients with respiratory illness in a pediatric emergency department. Ann Emerg Med. 1995;25:36–40. doi: 10.1016/S0196-0644(95)70352-7.
    1. Fouzas S, Priftis KN, Anthracopoulos MB. Pulse oximetry in pediatric practice. Pediatrics. 2011;128:740–752. doi: 10.1542/peds.2011-0271.
    1. Zhang L, Mendoza-Sassi R, Santos JCH, Lau J. Accuracy of symptoms and signs in predicting hypoxaemia among young children with acute respiratory infection: a meta-analysis. Int J Tuberc Lung Dis. 2011;15:317–325.
    1. Onyango FE, Steinhoff MC, Wafula EM, Wariua S, Musia J, Kitonyi J. Hypoxaemia in young Kenyan children with acute lower respiratory infection. BMJ. 1993;306:612–615. doi: 10.1136/bmj.306.6878.612.
    1. Mwaniki MK, Nokes DJ, Ignas J, Munywoki P, Ngama M, Newton CR, et al. Emergency triage assessment for hypoxaemia in neonates and young children in a Kenyan hospital: an observational study. Bulletin of the World Health Organization. 2009;87:263–270. doi: 10.2471/BLT.07.049148.
    1. Subhi R, Adamson M, Campbell H, Weber M, Smith K, Duke T, et al. The prevalence of hypoxaemia among ill children in developing countries: a systematic review. Lancet Infect Dis. 2009;9:219–227. doi: 10.1016/S1473-3099(09)70071-4.
    1. Mahle WT, Martin GR, Beekman RH, Morrow WR, Section on Cardiology and Cardiac Surgery Executive Committee Endorsement of Health and Human Services recommendation for pulse oximetry screening for critical congenital heart disease. Pediatrics. 2012;129:190–192. doi: 10.1542/peds.2011-3211.
    1. Ginsburg AS, Van Cleve WC, Thompson MIW, English M. Oxygen and pulse oximetry in childhood pneumonia: a survey of healthcare providers in resource-limited settings. J Trop Pediatr. 2012;58:389–393. doi: 10.1093/tropej/fmr103.
    1. Weber MW, Mulholland EK, Greenwood BM. Respiratory syncytial virus infection in tropical and developing countries. Trop Med Int Health. 1998;3:268–280. doi: 10.1046/j.1365-3156.1998.00213.x.
    1. Kattwinkel J, Perlman JM, Aziz K, Colby C, Fairchild K, Gallagher J, et al. Neonatal resuscitation: 2010 American Heart Association Guidelines for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. In Pediatrics. 2010;126:e1400–13. doi: 10.1542/peds.2010-2972E.
    1. Hay WW, Rodden DJ, Collins SM, Melara DL, Hale KA, Fashaw LM. Reliability of conventional and new pulse oximetry in neonatal patients. J Perinatol. 2002;22:360–366. doi: 10.1038/sj.jp.7210740.
    1. Workie FA, Rais-Bahrami K, Short BL. Clinical use of new-generation pulse oximeters in the neonatal intensive care unit. Am J Perinatol. 2005;22:357–360. doi: 10.1055/s-2005-872704.
    1. Ross PA, Newth CJL, Khemani RG. Accuracy of pulse oximetry in children. Pediatrics. 2014;133:22–29. doi: 10.1542/peds.2013-1760.
    1. Levesque BM, Pollack P, Griffin BE, Nielsen HC. Pulse oximetry: what's normal in the newborn nursery? Pediatr Pulmonol. 2000;30:406–412. doi: 10.1002/1099-0496(200011)30:5<406::AID-PPUL7>;2-C.

Source: PubMed

3
Subscribe