Validity of the Fitbit activity tracker for measuring steps in community-dwelling older adults

Serene S Paul, Anne Tiedemann, Leanne M Hassett, Elisabeth Ramsay, Catherine Kirkham, Sakina Chagpar, Catherine Sherrington, Serene S Paul, Anne Tiedemann, Leanne M Hassett, Elisabeth Ramsay, Catherine Kirkham, Sakina Chagpar, Catherine Sherrington

Abstract

Background: Commercially available activity monitors, such as the Fitbit, may encourage physical activity. However, the accuracy of the Fitbit in older adults remains unknown. This study aimed to determine (1) the criterion validity of Fitbit step counts compared to visual count and ActiGraph accelerometer step counts and (2) the accuracy of ActiGraph step counts compared to visual count in community-dwelling older people.

Methods: Thirty-two community-dwelling adults aged over 60 wore Fitbit and ActiGraph devices simultaneously during a 2 min walk test (2MWT) and then during waking hours over a 7-day period. A physiotherapist counted the steps taken during the 2MWT.

Results: There was excellent agreement between Fitbit and visually counted steps (intraclass correlation coefficient (ICC2,1)=0.88, 95% CI 0.76 to 0.94) from the 2MWT, and good agreement between Fitbit and ActiGraph (ICC2,1=0.66, 95% CI 0.41 to 0.82), and between ActiGraph and visually counted steps (ICC2,1=0.60, 95% CI 0.33 to 0.79). There was excellent agreement between the Fitbit and ActiGraph in average steps/day over 7 days (ICC2,1=0.94, 95% CI 0.88 to 0.97). Percentage agreement was closest for Fitbit steps compared to visual count (mean 0%, SD 4%) and least for Fitbit average steps/day compared to the ActiGraph (mean 13%, SD 25%).

Conclusions: The Fitbit accurately tracked steps during the 2MWT, but the ActiGraph appeared to underestimate steps. There was strong agreement between Fitbit and ActiGraph counted steps. The Fitbit tracker is sufficiently accurate to be used among community-dwelling older adults to monitor and give feedback on step counts.

Keywords: accelerometer; aged; pedometer; physical activity; validity.

Figures

Figure 1
Figure 1
Bland-Altman plots comparing step counts over a 2 min walk test for the Fitbit tracker, ActiGraph accelerometer and visual count by a physiotherapist. (A) Comparison between the Fitbit tracker and visual count. (B) Comparison between the Fitbit tracker and ActiGraph accelerometer. (C) Comparison between the ActiGraph accelerometer and visual count. Actigraph_2MWT: ActiGraph step counts from the 2 min walk test. Fitbit_2MWT: Fitbit step counts from the 2 min walk test. Visual_2MWT: Physiotherapist-counted steps from the 2 min walk test. Solid line indicates the mean difference between the two measures, dashed lines indicate the limits of agreement (1.96 SDs of the mean difference).
Figure 2
Figure 2
Bland-Altman plot comparing average steps/day over a 7-day period for the Fitbit tracker and ActiGraph accelerometer. Fitbit_week: Fitbit average steps/day from a 7-day period. Actigraph_week: ActiGraph average steps/day from a 7-day period. Solid line indicates the mean difference between the two measures, dashed lines indicate the limits of agreement (1.96 SDs of the mean difference).

References

    1. Lee IM, Shiroma EJ, Lobelo F et al. . Effect of physical inactivity on major non-communicable diseases worldwide: an analysis of burden of disease and life expectancy. Lancet 2012;380:219–29. 10.1016/S0140-6736(12)61031-9
    1. Peeters GM, Mishra GD, Dobson AJ et al. . Health care costs associated with prolonged sitting and inactivity. Am J Prev Med 2014;46:265–72. 10.1016/j.amepre.2013.11.014
    1. Department of Health. Australia's Physical Activity and Sedentary Behaviour Guidelines. 2014.
    1. Loprinzi PD. Accelerometer-determined sedentary and physical activity estimates among older adults with diabetes: considerations by demographic and comorbidity characteristics. J Aging Phys Activ 2014;22:432–40. 10.1123/JAPA.2013-0019
    1. Tudor-Locke C. Steps to better cardiovascular health: how many steps does it take to achieve good health and how confident are we in this number? Curr Cardiovasc Risk Rep 2010;4:271–6. 10.1007/s12170-010-0109-5
    1. van der Ploeg HP, Chey T, Korda RJ et al. . Sitting time and all-cause mortality risk in 222 497 Australian adults. Arch Intern Med 2012;172:494–500. 10.1001/archinternmed.2011.2174
    1. Ekblom-Bak E, Ekblom B, Vikström M et al. . The importance of non-exercise physical activity for cardiovascular health and longevity. Br J Sports Med 2014;48:233–8. 10.1136/bjsports-2012-092038
    1. Tudor-Locke C, Johnson WD, Katzmarzyk PT. Relationship between accelerometer-determined steps/day and other accelerometer outputs in US adults. J Phys Act Health 2011;8:410–9. 10.1136/bjsports-2012-092038
    1. Kang M, Marshall SJ, Barreira TV et al. . Effect of pedometer-based physical activity interventions: a meta-analysis. Res Q Exerc Sport 2009;80:648–55. 10.1080/02701367.2009.10599546
    1. Diaz KM, Krupka DJ, Chang MJ et al. . Fitbit(R): an accurate and reliable device for wireless physical activity tracking. Int J Cardiol 2015;185:138–40. 10.1016/j.ijcard.2015.03.038
    1. Adam Noah J, Spierer DK, Gu J et al. . Comparison of steps and energy expenditure assessment in adults of Fitbit Tracker and Ultra to the Actical and indirect calorimetry. J Med Eng Technol 2013;37:456–62. 10.3109/03091902.2013.831135
    1. Tully MA, McBride C, Heron L et al. . The validation of Fibit Zip physical activity monitor as a measure of free-living physical activity. BMC Res Notes 2014;7:952 10.1186/1756-0500-7-952
    1. Grant PM, Dall PM, Mitchell SL et al. . Activity-monitor accuracy in measuring step number and cadence in community-dwelling older adults. J Phys Act Health 2008;16:201–14.
    1. Webber SC, Magill SM, Schafer JL et al. . GT3X+ accelerometer, Yamax pedometer and SC-StepMX pedometer step count accuracy in community-dwelling older adults. J Aging Phys Activ 2014;22:334–41. 10.1123/JAPA.2013-0002
    1. Fulk GD, Combs SA, Danks KA et al. . Accuracy of 2 activity monitors in detecting steps in people with stroke and traumatic brain injury. Phys Ther 2014;94:222–9. 10.2522/ptj.20120525
    1. Taraldsen K, Askim T, Sletvold O et al. . Evaluation of a body-worn sensor system to measure physical activity in older people with impaired function. Phys Ther 2011;91:277–85. 10.2522/ptj.20100159
    1. Trost SG, Tudor-Locke C. Advances in the science of objective physical activity monitoring: 3rd International Conference on Ambulatory Monitoring of Physical Activity and Movement. Br J Sports Med 2014;48:1009–10. 10.1136/bjsports-2014-093865
    1. Mokkink LB, Terwee CB, Patrick DL et al. . COSMIN checklist manual. Amsterdam: EMGO Institute for Health and Care Research, VU University Medical Center, 2012.
    1. Plasqui G, Westerterp KR. Physical activity assessment with accelerometers: an evaluation against doubly labeled water. Obesity Research (Silver Spring) 2007;15:2371–9. 10.1038/oby.2007.281
    1. Tiedemann A, Paul S, Ramsay E et al. . What is the effect of a combined physical activity and fall prevention intervention enhanced with health coaching and pedometers on older adults’ physical activity levels and mobility-related goals? Study protocol for a randomised controlled trial. BMC Public Health 2015;15:477 10.1186/s12889-015-1380-7
    1. Lipton RB, Katz MJ, Kuslansky G et al. . Screening for dementia by telephone using the memory impairment screen. J Am Geriatr Soc 2003;51:1382–90. 10.1046/j.1532-5415.2003.51455.x
    1. Hallgren KA. Computing inter-rater reliability for observational data: an overview and tutorial. Tutor Quant Methods Psychol 2012;8:23–34. 10.1136/bjsports-2012-092038
    1. Storm FA, Heller BW, Mazza C. Step detection and activity recognition accuracy of seven physical activity monitors. PLoS ONE 2015;10:e0118723 10.1371/journal.pone.0118723
    1. Lyons EJ, Lewis ZH, Mayrsohn BG et al. . Behavior change techniques implemented in electronic lifestyle activity monitors: a systematic content analysis. J Med Internet Res 2014;16:e192 10.2196/jmir.3469
    1. Lelas JL, Merriman GJ, Riley PO et al. . Predicting peak kinematic and kinetic parameters from gait speed. Gait Posture 2003;17:106–12. 10.1016/S0966-6362(02)00060-7
    1. Tudor-Locke C, Ainsworth BE, Thompson RW et al. . Comparison of pedometer and accelerometer measures of free-living physical activity. Med Sci Sport Exer 2002;34:2045–51. 10.1097/00005768-200212000-00027
    1. Lee IM, Shiroma EJ. Using accelerometers to measure physical activity in large-scale epidemiological studies: issues and challenges. Br J Sports Med 2014;48:197–201. 10.1136/bjsports-2013-093154
    1. Trost SG, O'Neil M. Clinical use of objective measures of physical activity. Br J Sports Med 2014;48:178–81. 10.1136/bjsports-2013-093173

Source: PubMed

3
Subscribe