Comparison of drug-eluting and bare metal stents for extracranial vertebral artery stenting

Damian R Maciejewski, Piotr Pieniazek, Lukasz Tekieli, Piotr Paluszek, Tadeusz Przewlocki, Tomasz Tomaszewski, Roman Machnik, Mariusz Trystula, Jacek Legutko, Anna Kablak-Ziembicka, Damian R Maciejewski, Piotr Pieniazek, Lukasz Tekieli, Piotr Paluszek, Tadeusz Przewlocki, Tomasz Tomaszewski, Roman Machnik, Mariusz Trystula, Jacek Legutko, Anna Kablak-Ziembicka

Abstract

Introduction: Drug-eluting stents of the first (DES I) and second generation (DES II) proved superior to bare metal stents (BMS) in the coronary territory. However, there are limited data on whether they have any advantage over BMS in vertebral artery stenosis (VAS).

Aim: To compare outcomes of DES (DES I, DES II) and BMS in the treatment of symptomatic extracranial VAS.

Material and methods: During 13-year study period (2003-2016), 392 consecutive patients underwent VAS angioplasty in 428 arteries, including implantation of 148 DES (DES I: 21; DES II: 127 lesions), and 280 BMS.

Results: The technical success rates for DES and BMS groups were 96.7% and 94.6% (p = 0.103), with similar periprocedural complication rates (1.4% vs. 2.2%; p = 0.565). VAS degree was reduced from 86 ±9.7 to 2.7 ±5.0% in DES (p < 0.001) and from 84.1 ±9.4 to 4.3 ±6.9% in BMS (p < 0.001). Angiography confirmed in-stent restenosis/occlusion (ISR/ISO) 50-99% in 53 (14.2%) and 21 (5.6%) out of 373 patients (409 arteries) with at least 6-month follow-up. ISR/ISO rates were similar in DES vs. BMS (22.8% vs. 19.4%; p = 0.635), as well as in DES I vs. DES II (6/19; 31.6% vs. 25/92; 27.2%, p = 0.325). Stainless steel (24/135; 17.8%) and cobalt-chromium (23/121;19%) BMS had significantly lower incidence of ISR/ISO, as compared to platinum-chromium (7/18; 38.9%), p = 0.034. ISR/ISO was associated with age (p = 0.01) and CRP level > 5 mg/l (p = 0.043), while greater stent length was associated with ISR only in the DES group (p = 0.024).

Conclusions: Our results do not support significant differences in ISR/ISO rates between DES and BMS, although differences between particular stent types and ISR rates require further investigation.

Keywords: bare metal stent; drug-eluting stent; in-stent restenosis; vertebral artery stenting.

Conflict of interest statement

The authors declare no conflict of interest.

Copyright: © 2019 Termedia Sp. z o. o.

Figures

Figure 1
Figure 1
A – Right vertebral artery critical stenosis (white arrow). B – Balloon post-dilatation after balloon-expandable Multi-Link Vision (BMS – cobalt-chromium; Abbott) 3.5 mm × 12 mm direct stenting. C – Final angiography showing optimal effect of VA stenting. D – Angiography confirming in-stent restenosis (white arrow). E – Sprinter (Medtronic) 3.0 mm × 15 mm balloon pre-dilatation. F – Dior (Eurocor) 3.5 mm × 15 mm drug-eluting balloon (DEB) dilatation. G – Final angiography of re-angioplasty
Figure 2
Figure 2
A – Survival according to the level of creatinine. B – Survival related to MI or stroke. C – Overall survival curves
Figure 3
Figure 3
Survival without ISR/ISO according to the level of CRP

References

    1. Cloud GC, Markus HS. Diagnosis and management of vertebral artery stenosis. QJM. 2003;96:27–54.
    1. Savitz SI, Caplan LR. Vertebrobasilar disease. N Engl J Med. 2005;352:2618–26.
    1. Kocak B, Korkmazer B, Islak C, et al. Endovascular treatment of extracranial vertebral artery stenosis. World J Radiol. 2012;4:391–400.
    1. Wityk RJ, Chang HM, Rosengart A, et al. Proximal extracranial vertebral artery disease in the New England medical center posterior circulation registry. Arch Neurol. 1998;55:470–8.
    1. Markus HS, Larsson SC, Kuker W, et al. Stenting for symptomatic vertebral artery stenosis: the Vertebral Artery Ischaemia Stenting Trial. Neurology. 2017;89:1229–36.
    1. Berguer R, Flynn L, Kline R, et al. Surgical reconstruction of the extracranial vertebral artery: management and outcome. J Vasc Surg. 2000;31:9–18.
    1. Borhanim Haghighi A, Edgell RC, Cruz-Flores S, et al. Vertebral artery origin stenosis and its treatment. J Stroke Cerebrovasc Dis. 2011;20:369–76.
    1. Stayman AN, Nogueira RG, Gupta R. A systematic review of stenting and angioplasty of symptomatic extracranial vertebral artery stenosis. Stroke. 2011;42:2212–6.
    1. Navarese EP, Kowalewski M, Kandzari D, et al. First-generation versus second-generation drug-eluting stents in current clinical practice: updated evidence from a comprehensive meta-analysis of randomised clinical trials comprising 31 379 patients. Open Heart. 2014;1:e000064.
    1. Abu Sharar H, Gomes B, Chorianopoulos E, et al. Procedural advantages of a novel coronary stent design with ultra-thin struts and bioabsorbable abluminal polymer coating in an all-comers registry. Adv Interv Cardiol. 2018;14:240–6.
    1. Gulli G, Marquardt L, Rothwell PM, Markus HS. Stroke risk after posterior circulation stroke/transient ischemic attack and its relationship to site of vertebrobasilar stenosis: pooled data analysis from prospective studies. Stroke. 2013;44:598–604.
    1. Aboyans V, Ricco JB, Bartelink MEL, et al. 2017 ESC Guidelines on the Diagnosis and Treatment of Peripheral Arterial Diseases, in collaboration with the European Society for Vascular Surgery (ESVS): document covering atherosclerotic disease of extracranial carotid and vertebral, mesenteric, renal, upper and lower extremity arteries. Endorsed by: the European Stroke Organization (ESO) The Task Force for the Diagnosis and Treatment of Peripheral Arterial Diseases of the European Society of Cardiology (ESC) and of the European Society for Vascular Surgery (ESVS) Eur Heart J. 2018;39:763–816.
    1. Ogilvy CS, Yang X, Natarajan SK, et al. Restenosis rates following vertebral artery origin stenting: does stent type make a difference? J Invasive Cardiol. 2010;22:119–24.
    1. Langwieser N, Buyer D, Schuster T, et al. Bare metal vs. drug-eluting stents for extracranial vertebral artery disease: a meta-analysis of nonrandomized comparative studies. J Endovasc Ther. 2014;21:683–92.
    1. Lugmayr H, Kastner M, Fröhler W, et al. Sirolimus-eluting stents for the treatment of symptomatic extracranial vertebral artery stenoses: early experience and 6-month follow-up. Rofo. 2004;176:1431–5.
    1. Taylor RA, Siddiq F, Suri MF, et al. Risk factors for in-stent restenosis after vertebral ostium stenting. J Endovasc Ther. 2008;15:203–12.
    1. Eberhardt O, Naegele T, Raygrotzki S, et al. Stenting of vertebrobasilar arteries in symptomatic atherosclerotic disease and acute occlusion: case series and review of the literature. J Vasc Surg. 2006;43:1145–54.
    1. Tank H, Ghosh R, Gupta V, et al. Drug eluting stents versus bare metal stents for the treatment of extracranial vertebral artery disease: a meta-analysis. J Neurointerv Surg. 2016;8:770–4.
    1. Gąsior P, Gierlotka M, Szczurek-Katanski K, et al. Safety and efficacy of biodegradable polymer-coated thin strut sirolimus-eluting stent vs. durable polymer-coated everolimus-eluting stent in patients with acute myocardial infarction. Adv Interv Cardiol. 2018;14:347–55.
    1. Che WQ, Dong H, Jiang XJ, et al. Clinical outcomes and influencing factors of in-stent restenosis after stenting for symptomatic stenosis of the vertebral V1 segment. J Vasc Surg. 2018;68:1406–13.
    1. Song L, Li J, Gu Y, et al. Drug-eluting vs. bare metal stents for symptomatic vertebral artery stenosis. J Endovasc Ther. 2012;19:231–8.
    1. He Y, Li T, Bai W, et al. Cerebrovascular drug-eluting stent versus bare-metal stent in the treatment of vertebral artery stenosis: a non-inferiority randomized clinical trial. J Stroke. 2019;21:101–4.
    1. Langwieser N, Prothmann S, Buyer D, et al. Safety and efficacy of different stent types for the endovascular therapy of extracranial vertebral artery disease. Clin Res Cardiol. 2014;103:353–62.
    1. Jenkins JS, Stewart M. Endovascular treatment of vertebral artery stenosis. Prog Cardiovasc. 2017;59:619–25.
    1. Paluszek P, Pieniążek P, Musiałek P, et al. Symptomatic vertebral artery stenting with the use of bare metal and drug eluting stents. Postep Kardiol Interw. 2009;5:1–6.
    1. Zhou Z, Yin Q, Xu G, et al. Influence of vessel size and tortuosity on in-stent restenosis after stent implantation in the vertebral artery ostium. Cardiovasc Intervent Radiol. 2011;34:481–7.
    1. Chen W, Huang F, Li M, et al. Incidence and predictors of the in-stent restenosis after vertebral artery ostium stenting. J Stroke Cerebrovasc Dis. 2018;27:3030–5.
    1. Tekieli L, Pieniazek P, Musialek P, et al. Zotarolimus-eluting stent for the treatment of recurrent, severe carotid artery in-stent stenosis in the TARGET-CAS population. J Endovasc Ther. 2012;19:316–24.
    1. Gacoń J, Przewłocki T, Podolec J, et al. Prospective study on the prognostic value of repeated carotid intima-media thickness assessment in patients with coronary and extra coronary steno-occlusive arterial disease. Pol Arch Intern Med. 2019;129:12–21.
    1. Przewlocki T, Wrotniak L, Kablak-Ziembicka A, et al. Determinants of long-term outcome in patients after percutaneous stent-assisted management of symptomatic subclavian or innominate artery stenosis or occlusion. EuroIntervention. 2017;13:1355–64.
    1. Maciejewski D, Tekieli Ł, Kabłak-Ziembicka A, et al. Transradial approach for vertebral artery stenting. Postep Kardiol Interw. 2015;11:32–6.
    1. Werner M, Braunlich S, Ulrich M, et al. Drug eluting stents for the treatment of vertebral artery origin stenosis. J Endovasc Ther. 2010;17:232–40.

Source: PubMed

3
Subscribe