Evaluation of biomechanically corrected intraocular pressure using Corvis ST and comparison of the Corvis ST, noncontact tonometer, and Goldmann applanation tonometer in patients with glaucoma

Yoshitaka Nakao, Yoshiaki Kiuchi, Hideaki Okumichi, Yoshitaka Nakao, Yoshiaki Kiuchi, Hideaki Okumichi

Abstract

Purpose: The aim of the study was to investigate the effects of various anatomical structures on intraocular pressure (IOP) measurements obtained by the Corneal Visualization Scheimpflug Technology (Corvis ST), Goldmann applanation tonometer (GAT), and noncontact tonometer (NCT), as well as to assess the interchangeability among the four types of IOP measurement: IOP-GAT, IOP-NCT, IOP-Corvis, and biomechanically corrected IOP (bIOP-Corvis), with a particular focus on bIOP-Corvis.

Materials and methods: We included 71 patients with primary open-angle glaucoma and assessed their IOP measurements obtained with the GAT, NCT, and Corvis ST using a repeated measures ANOVA, a paired t-test with Bonferroni correction, stepwise multiple regression analyses and Bland-Altman plots.

Results: IOP-GAT showed the highest values (13.5 ± 2.1 mmHg [mean ± standard deviation]), followed by IOP-NCT (13.2 ± 2.7 mmHg), IOP-Corvis (10.6 ± 2.8 mmHg), and bIOP-Corvis (10.0 ± 2.3 mmHg). With exceptions of bIOP-Corvis and IOP-GAT, all IOP variations were explained by regression coefficients involving the central corneal thickness. Bland-Altman plots showed a mean difference between IOP-GAT and the other IOP measurements (IOP-Corvis, bIOP-Corvis, and IOP-NCT), which were -2.90, -3.48, and -0.29 mmHg, respectively. The widths of the 95% limits of agreement between all pairs of IOP measurements were greater than 3 mmHg.

Conclusion: IOP values obtained with the Corvis ST, NCT, and GAT were not interchangeable. The bIOP-Corvis measurement corrected for the ocular structure.

Conflict of interest statement

The authors have declared that no competing interests exist.

Figures

Fig 1. Bland–Altman plots between IOPs obtained…
Fig 1. Bland–Altman plots between IOPs obtained with the Corvis ST, GAT, and CT-90A.
A, IOP-Corvis and IOP-GAT. B, bIOP-Corvis and IOP-GAT. C, IOP-NCT and IOP-GAT. The mean values and 95% LOA are indicated by bold lines and solid lines, respectively. IOP, intraocular pressure; IOP-Corvis, the IOP obtained by Corvis ST; bIOP-Corvis, corrected IOP-Corvis; IOP-GAT, the IOP obtained by Goldmann applanation tonometry; IOP-NCT, the IOP obtained by CT-90A tonometer; LOA, limits of agreement.

References

    1. Heijl A, Leske MC, Bengtsson B, Hyman L, Bengtsson B, Hussein M. Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial. Arch Ophthalmol. 2002;120(10):1268–79. Epub 2002/10/09. 10.1001/archopht.120.10.1268 .
    1. Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol (Copenh). 1975;53(1):34–43. Epub 1975/03/01. 10.1111/j.1755-3768.1975.tb01135.x .
    1. Graf M. [Significance of the corneal thickness in non-contact tonometry]. Klin Monbl Augenheilkd. 1991;199(3):183–6. Epub 1991/09/01. 10.1055/s-2008-1046068 .
    1. Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol. 1993;115(5):592–6. Epub 1993/05/15. 10.1016/s0002-9394(14)71455-2 .
    1. Matsumoto T, Makino H, Uozato H, Saishin M, Miyamoto S. The influence of corneal thickness and curvature on the difference between intraocular pressure measurements obtained with a non-contact tonometer and those with a Goldmann applanation tonometer. Jpn J Ophthalmol. 2000;44(6):691 Epub 2000/11/30. 10.1016/s0021-5155(00)00250-1 .
    1. Recep OF, Hasiripi H, Cagil N, Sarikatipoglu H. Relation between corneal thickness and intraocular pressure measurement by noncontact and applanation tonometry. J Cataract Refract Surg. 2001;27(11):1787–91. Epub 2001/11/16. 10.1016/s0886-3350(01)00900-2 .
    1. Tonnu PA, Ho T, Newson T, El Sheikh A, Sharma K, White E, et al. The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry. Br J Ophthalmol. 2005;89(7):851–4. Epub 2005/06/21. 10.1136/bjo.2004.056622
    1. Medeiros FA, Weinreb RN. Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J Glaucoma. 2006;15(5):364–70. Epub 2006/09/22. 10.1097/01.ijg.0000212268.42606.97 .
    1. Garcia-Medina M, Garcia-Medina JJ, Garrido-Fernandez P, Galvan-Espinosa J, Martin-Molina J, Garcia-Maturana C, et al. Central corneal thickness, intraocular pressure, and degree of myopia in an adult myopic population aged 20 to 40 years in southeast Spain: determination and relationships. Clin Ophthalmol. 2011;5:249–58. Epub 2011/04/07. 10.2147/OPTH.S16594
    1. Salvetat ML, Zeppieri M, Tosoni C, Felletti M, Grasso L, Brusini P. Corneal Deformation Parameters Provided by the Corvis-ST Pachy-Tonometer in Healthy Subjects and Glaucoma Patients. J Glaucoma. 2015;24(8):568–74. Epub 2014/10/17. 10.1097/IJG.0000000000000133 .
    1. Nakao Y, Kiuchi Y, Okimoto S. A Comparison of the Corrected Intraocular Pressure Obtained by the Corvis ST and Reichert 7CR Tonometers in Glaucoma Patients. PloS one. 2017;12(1):e0170206 Epub 2017/01/18. 10.1371/journal.pone.0170206 .
    1. Ye Y, Yang Y, Fan Y, Lan M, Yu K, Yu M. Comparison of Biomechanically Corrected Intraocular Pressure Obtained by Corvis ST and Goldmann Applanation Tonometry in Patients With Open-angle Glaucoma and Ocular Hypertension. J Glaucoma. 2019;28(10):922–8. Epub 2019/08/28. 10.1097/IJG.0000000000001348 .
    1. Mark HH, Robbins KP, Mark TL. Axial length in applanation tonometry. J Cataract Refract Surg. 2002;28(3):504–6. Epub 2002/04/26. 10.1016/s0886-3350(01)01091-4 .
    1. Altan C, Kara N, Baz O, Satana B, Demirok A, Yilmaz OF. Corneal biomechanical properties and intraocular pressure measurement in patients with nanophthalmos. Br J Ophthalmol. 2012;96(6):806–10. Epub 2012/03/09. 10.1136/bjophthalmol-2011-300557 .
    1. Mark HH. Corneal curvature in applanation tonometry. Am J Ophthalmol. 1973;76(2):223–4. Epub 1973/08/01. 10.1016/0002-9394(73)90164-5 .
    1. Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg. 2005;31(1):156–62. Epub 2005/02/22. 10.1016/j.jcrs.2004.10.044 .
    1. Vinciguerra R, Elsheikh A, Roberts CJ, Ambrosio R Jr., Kang DS, Lopes BT, et al. Influence of Pachymetry and Intraocular Pressure on Dynamic Corneal Response Parameters in Healthy Patients. J Refract Surg (Thorofare, NJ: 1995). 2016;32(8):550–61. Epub 2016/08/10. 10.3928/1081597x-20160524-01 .
    1. Matsuura M, Murata H, Fujino Y, Yanagisawa M, Nakao Y, Nakakura S, et al. Repeatability of the Novel Intraocular Pressure Measurement From Corvis ST. Transl Vis Sci Technol. 2019;8(3):48 Epub 2019/07/06. 10.1167/tvst.8.3.48
    1. Perez-Rico C, Gutierrez-Ortiz C, Gonzalez-Mesa A, Zandueta AM, Moreno-Salgueiro A, Germain F. Effect of diabetes mellitus on Corvis ST measurement process. Acta Ophthalmol. 2015;93(3):e193–e8. 10.1111/aos.12530 WOS:000353053000003.
    1. Wang C, Li AL, Pang Y, Lei YQ, Yu L. Changes in intraocular pressure and central corneal thickness during pregnancy: a systematic review and Meta-analysis. Int J Ophthalmol. 2017;10(10):1573–9. Epub 2017/10/25. 10.18240/ijo.2017.10.15
    1. Reznicek L, Muth D, Kampik A, Neubauer AS, Hirneiss C. Evaluation of a novel Scheimpflug-based non-contact tonometer in healthy subjects and patients with ocular hypertension and glaucoma. Br J Ophthalmol. 2013;97(11):1410–4. 10.1136/bjophthalmol-2013-303400 .
    1. Vinciguerra R, Rehman S, Vallabh NA, Batterbury M, Czanner G, Choudhary A, et al. Corneal biomechanics and biomechanically corrected intraocular pressure in primary open-angle glaucoma, ocular hypertension and controls. Br J Ophthalmol. 2020;104(1):121–6. Epub 2019/03/30. 10.1136/bjophthalmol-2018-313493 .
    1. Hong J, Xu J, Wei A, Deng SX, Cui X, Yu X, et al. A new tonometer—the Corvis ST tonometer: clinical comparison with noncontact and Goldmann applanation tonometers. Invest Ophthalmol Vis Sci. 2013;54(1):659–65. Epub 2013/01/12. 10.1167/iovs.12-10984 .
    1. Smedowski A, Weglarz B, Tarnawska D, Kaarniranta K, Wylegala E. Comparison of three intraocular pressure measurement methods including biomechanical properties of the cornea. Invest Ophthalmol Vis Sci. 2014;55(2):666–73. Epub 2014/01/16. 10.1167/iovs.13-13172 .
    1. Lanza M, Iaccarino S, Cennamo M, Irregolare C, Romano V, Carnevale UA. Comparison between Corvis and other tonometers in healthy eyes. Cont Lens Anterior Eye. 2015;38(2):94–8. Epub 2014/12/04. 10.1016/j.clae.2014.11.001 .
    1. Jorge J, Diaz-Rey JA, Gonzalez-Meijome JM, Almeida JB, Parafita MA. Clinical performance of the Reichert AT550: a new non-contact tonometer. Ophthalmic Physiol Opt. 2002;22(6):560–4. Epub 2002/12/13. 10.1046/j.1475-1313.2002.00077.x .
    1. Jorge J, Gonzalez-Meijome JM, Diaz-Rey JA, Almeida JB, Ribeiro P, Parafita MA. Clinical performance of non-contact tonometry by Reichert AT550 in glaucomatous patients. Ophthalmic Physiol Opt. 2003;23(6):503–6. Epub 2003/11/19. 10.1046/j.1475-1313.2003.00139.x .
    1. Schiano Lomoriello D, Lombardo M, Tranchina L, Oddone F, Serrao S, Ducoli P. Repeatability of intra-ocular pressure and central corneal thickness measurements provided by a non-contact method of tonometry and pachymetry. Graefes Arch Clin Exp Ophthalmol. 2011;249(3):429–34. Epub 2010/10/29. 10.1007/s00417-010-1550-3 .
    1. Niesel P, Flammer J. Correlations between intraocular pressure, visual field and visual acuity, based on 11 years of observations of treated chronic glaucomas. Int J Ophthalmol. 1980;3(1):31–5. Epub 1980/12/01. 10.1007/bf00136211 .
    1. Seidman C, De Moraes CG, Je RF, Liebmann R, Ritch R. The Relationship Between Visual Acuity, Intraocular Pressure, and Rates of Visual Field Progression Using 10–2 Perimetry. Invest Ophthalmol Vis Sci. 2013;54(15). WOS:000436232709029.
    1. Azuara-Blanco A, Katz LJ, Spaeth GL, Wilson RP, Moster MR, Flartey KJ. Effect of latanoprost on intraocular pressure in patients with glaucoma on maximal tolerated medical treatment. Br J Ophthalmol. 1997;81(12):1116 Epub 1998/03/14. 10.1136/bjo.81.12.1115b
    1. Lee DA, Gornbein JA. Effectiveness and safety of brimonidine as adjunctive therapy for patients with elevated intraocular pressure in a large, open-label community trial. J Glaucoma. 2001;10(3):220–6. Epub 2001/07/10. 10.1097/00061198-200106000-00014 .

Source: PubMed

3
Subscribe