End-tidal capnography and upper airway ultrasonography in the rapid confirmation of endotracheal tube placement in patients requiring intubation for general anaesthesia

Chintamani Abhishek, Kartik Munta, S Manimala Rao, C N Chandrasekhar, Chintamani Abhishek, Kartik Munta, S Manimala Rao, C N Chandrasekhar

Abstract

Background and aims: Confirmation of correct endotracheal tube placement is essential immediately after intubation for general anaesthesia. In this study, we have compared upper airway ultrasonography (USG) with reference to capnography for rapid confirmation of endotracheal tube placement after general anaesthesia.

Methods: A prospective, single centre, observational study was conducted on 100 patients requiring tracheal intubation for general anaesthesia. Both capnography and upper airway USG were performed immediately after intubation to confirm the endotracheal tube (ETT) placement. Sensitivity, specificity, and positive and negative predictive values of upper airway USG were determined against capnography as the reference method. Agreement between the methods and time required to determine ETT placement by the two methods were assessed with kappa statistics and Student's t-test.

Results: Upper airway USG detected all five cases of oesophageal intubation, but could not detect five patients with correct tracheal intubation. Upper airway USG had a sensitivity of 96.84% (95% confidence interval [CI]: 94.25%-96.84%), specificity of 100% (95% CI: 50.6%-100%), positive predictive value of 100% (95% CI: 97.3%-100%) and negative predictive value of 62.5% (95% CI: 31.6%-62.5%). Kappa value was found to be 0.76, indicating a good agreement between upper airway USG and capnography for confirmation of ETT placement. Time taken for confirmation of ETT by capnography was 8.989 ± 1.043 s vs. 12.0 ± 1.318 s for upper airway USG (P < 0.001).

Conclusion: Both capnography and upper airway USG may be used as primary procedures for the confirmation of ETT placement.

Keywords: Capnography; general anaesthesia; tracheal intubation; upper airway ultrasonography.

Conflict of interest statement

There are no conflicts of interest.

Figures

Figure 1
Figure 1
Ultrasound image after tracheal intubation showing posterior shadowing in trachea and oesophagus which is empty
Figure 2
Figure 2
Ultrasound image after tracheal intubation in longitudinal view showing the presence of endotracheal tube

References

    1. Mort TC. Unplanned tracheal extubation outside the operating room: A quality improvement audit of hemodynamic and tracheal airway complications associated with emergency tracheal reintubation. Anesth Analg. 1998;86:1171–6.
    1. Cook TM, Woodall N, Frerk C. Fourth National Audit Project. Major complications of airway management in the UK: Results of the Fourth National Audit Project of the Royal College of Anaesthetists and the Difficult Airway Society. Part 1: Anaesthesia. Br J Anaesth. 2011;106:617–31.
    1. Bozeman WP, Hexter D, Liang HK, Kelen GD. Esophageal detector device versus detection of end-tidal carbon dioxide level in emergency intubation. Ann Emerg Med. 1996;27:595–9.
    1. Grmec S. Comparison of three different methods to confirm tracheal tube placement in emergency intubation. Intensive Care Med. 2002;28:701–4.
    1. Kundra P, Mishra SK, Ramesh A. Ultrasound of the airway. Indian J Anaesth. 2011;55:456–62.
    1. Chou EH, Dickman E, Tsou PY, Tessaro M, Tsai YM, Ma MH, et al. Ultrasonography for confirmation of endotracheal tube placement: A systematic review and meta-analysis. Resuscitation. 2015;90:97–103.
    1. Chou H-C, Chong K-M, Sim S-S, Ma MH-M, Liu S-H, Chen N-C, et al. Real-time tracheal ultrasonography for confirmation of endotracheal tube placement during cardiopulmonary resuscitation. Resuscitation. 2013;84:1708–12.
    1. Chenkin J, McCartney CJ, Jelic T, Romano M, Heslop C, Bandiera G, et al. Defining the learning curve of point-of-care ultrasound for confirming endotracheal tube placement by emergency physicians. Crit Ultrasound J. 2015;7:14.
    1. Adi O, Chuan TW, Rishya M. A feasibility study on bedside upper airway ultrasonography compared to waveform capnography for verifying endotracheal tube location after intubation. Crit Ultrasound J. 2013;5:7.
    1. Muslu B, Sert H, Kaya A, Demircioglu RI, Gozdemir M, Usta B, Boynukalin KS. Use of sonography for rapid identification of esophageal and tracheal intubations in adult patients. J Ultrasound Med. 2011;30:671–6.
    1. Chou HC, Tseng WP, Wang CH, Ma MH, Wang HP, Huang PC, et al. Tracheal rapid ultrasound exam (T.R.U.E) for confirming endotracheal tube placement during emergency intubation. Resuscitation. 2011;82:1279–84.
    1. Takeda T, Tanigawa K, Tanaka H, Hayashi Y, Goto E, Tanaka K. The assessment of three methods to verify tracheal tube placement in the emergency setting. Resuscitation. 2003;56:153–7.
    1. Werner SL, Smith CE, Goldstein JR, Jones RA, Cydulka RK. Pilot study to evaluate the accuracy of ultrasonography in confirming endotracheal tube placement. Ann Emerg Med. 2007;49:75–80.
    1. Sustic A. Role of ultrasound in the airway management of critically ill patients. Crit Care Med. 2007;35(5 Suppl):S173–7.
    1. Rajan S, Surendran J, Paul J, Kumar L. Rapidity and efficacy of ultrasonographic sliding lung sign and auscultation in confirming endotracheal intubation in overweight and obese patients. Indian J Anaesth. 2017;61:230–4.