Test-retest reliability of an insole plantar pressure system to assess gait along linear and curved trajectories

Marco Godi, Anna Maria Turcato, Marco Schieppati, Antonio Nardone, Marco Godi, Anna Maria Turcato, Marco Schieppati, Antonio Nardone

Abstract

Background: Previous studies have assessed reliability of insole technology for evaluating foot pressure distribution during linear walking. Since in natural motion straight walking is intermingled with turns, we determined the test-retest reliability of insole assessment for curved as well as linear trajectories, and estimated the minimum number of steps required to obtain excellent reliability for each output variable.

Methods: Sixteen young healthy participants were recruited. Each performed, two days apart, two sessions of three walking conditions: linear (LIN) and curved, clockwise (CW) and counter-clockwise (CCW). The Pedar-X system was used to collect pressure distribution. Foot print was analyzed both as a whole and as subdivided into eight regions: medial and lateral heel, medial and lateral arch, I metatarsal head, II-V metatarsal heads, hallux, lateral toes. Reliability was assessed by using intraclass correlation coefficient (ICC) for clinically relevant variables from analysis of 50 steps per trajectory: Peak Force (PF); Peak Pressure (PP); Contact Area (CA); Stance Duration (S).

Results: When considering whole-foot, all variables showed an ICC >0.80, therefore highly reliable. This was true for both LIN and curved trajectories. There was no difference in ICC of the four variables between left and right foot. When collapsing foot and trajectories, S had a lower ICC than PP and CA, and PP lower than CA. Mean percent error between the values of first and second session was <5%. When separately considering the eight foot regions, ICCs of PF, PP and CA for all regions and trajectories were generally >0.90, indicating excellent reliability. In curved trajectories, S showed smaller ICCs. Since the least ICC value for S was 0.60 in LIN trajectory, we estimated that to achieve an ICC ≥0.90 more than 200 steps should be collected.

Conclusions: High reliability of insole dynamic variables (PF, PP, CA) is obtained with 50 steps using the Pedar-X system. On the contrary, high reliability of temporal variable (S) requires a larger step number. The negligible differences in ICC between LIN and curved trajectory allow use of this device for gait assessment along mixed trajectories in both clinical and research setting.

Figures

Figure 1
Figure 1
Test-retest repeatability of each of the eight anatomical regions of the foot (A). Intraclass correlation coefficients (ICC) obtained from 50 consecutive steps performed under each trajectory (linear; CW, clockwise; CCW, counter-clockwise) for the 4 output variables (peak force, B; peak pressure, C; contact area, D; stance duration, E). In the bottom insets, the grey foot is the right foot; this is the Foot-In for CW, and the Foot-Out for CCW. ICC values are obtained from the average of left and right foot in the case of linear trajectory, and Foot-In or Foot-Out in the case of CW and CCW. Each colour corresponds to a range of ICC value equal to ± 0.02 around the depicted mean value. L, lateral region of foot; M, medial region of foot.
Figure 2
Figure 2
Number of steps required to reach an ICC of 0.90 (good reliability) for the four variables considered. In the case of Peak Force, 25 steps are sufficient. For Stance Duration, it is necessary to collect more than 200 steps.

References

    1. Rosenbaum D, Becker H. Plantar pressure distribution measurements. Technical background and clinical applications. Foot Ankle Surg. 1997;3:1–14. doi: 10.1046/j.1460-9584.1997.00043.x.
    1. Orlin MN, McPoil TG. Plantar pressure assessment. Phys Ther. 2000;80(4):399–409.
    1. Putti AB, Arnold GP, Cochrane LA, Abboud RJ. Normal pressure values and repeatability of the Emed ST4 system. Gait Posture. 2008;27(3):501–505. doi: 10.1016/j.gaitpost.2007.06.009.
    1. Abboud R. Relevant foot biomechanics. Curr Orthop. 2002;16(3):165–179. doi: 10.1054/cuor.2002.0268.
    1. Ramanathan A, Fadel G, Jain A, Abboud R. Press-fit ceramic implant arthroplasty of the Hallux Metatarsophalangeal joint-evaluation of outcomes. Foot. 2008;18:34–39. doi: 10.1016/j.foot.2007.09.001.
    1. Vela SA, Lavery LA, Armstrong DG, Anaim AA. The effect of increased weight on peak pressures: implications for obesity and diabetic foot pathology. J Foot Ankle Surg. 1998;37(5):416–420. doi: 10.1016/S1067-2516(98)80051-3. [discussion 48–9]
    1. Gurney JK, Marshall PW, Rosenbaum D, Kersting UG. Test-retest reliability of dynamic plantar loading and foot geometry measures in diabetics with peripheral neuropathy. Gait Posture. 2013;37(1):135–137. doi: 10.1016/j.gaitpost.2012.06.007.
    1. Galen SS, Clarke CJ, Allan DB, Conway BA. A portable gait assessment tool to record temporal gait parameters in SCI. Med Eng Phys. 2011;33(5):626–632. doi: 10.1016/j.medengphy.2011.01.003.
    1. Pérennou DA, Hillier SL. Volumes of intact gray matter outside the stroke predict gait performance. Neurology. 2014;82(10):822–823. doi: 10.1212/WNL.0000000000000194.
    1. Lee NK, Kwon JW, Son SM, Nam SH, Choi YW, Kim CS. Changes of plantar pressure distributions following open and closed kinetic chain exercise in patients with stroke. NeuroRehabilitation. 2013;32(2):385–390.
    1. Schuh R, Hofstaetter SG, Adams SB Jr, Pichler F, Kristen KH, Trnka HJ. Rehabilitation after hallux valgus surgery: importance of physical therapy to restore weight bearing of the first ray during the stance phase. Phys Ther. 2009;89(9):934–945. doi: 10.2522/ptj.20080375.
    1. Kernozek TW, LaMott EE, Dancisak MJ. Reliability of an in-shoe pressure measurement system during treadmill walking. Foot Ankle Int. 1996;17(4):204–209. doi: 10.1177/107110079601700404.
    1. Glaister BC, Bernatz GC, Klute GK, Orendurff MS. Video task analysis of turning during activities of daily living. Gait Posture. 2007;25(2):289–294. doi: 10.1016/j.gaitpost.2006.04.003.
    1. Courtine G, Schieppati M. Human walking along a curved path. I. Body trajectory, segment orientation and the effect of vision. Eur J Neurosci. 2003;18(1):177–190. doi: 10.1046/j.1460-9568.2003.02736.x.
    1. Courtine G, Schieppati M. Tuning of a basic coordination pattern constructs straight-ahead and curved walking in humans. J Neurophysiol. 2004;91(4):1524–1535. doi: 10.1152/jn.00817.2003.
    1. Courtine G, Schieppati M. Human walking along a curved path. II. Gait features and EMG patterns. Eur J Neurosci. 2003;18(1):191–205. doi: 10.1046/j.1460-9568.2003.02737.x.
    1. Guglielmetti S, Nardone A, de Nunzio AM, Godi M, Schieppati M. Walking along circular trajectories in Parkinson’s disease. Mov Disord. 2009;24(4):598–604. doi: 10.1002/mds.22424.
    1. Godi M, Nardone A, Schieppati M. Curved walking in hemiparetic patients. J Rehabil Med. 2010;42(9):858–865. doi: 10.2340/16501977-0594.
    1. Schmid M, de Nunzio AM, Beretta MV, Schieppati M. Walking along a curved trajectory. Insight into the generation of the centripetal force. Gait Posture. 2004;20(Suppl 1):S116.
    1. Hurkmans HL, Bussmann JB, Selles RW, Horemans HL, Benda E, Stam HJ, Verhaar JA. Validity of the Pedar Mobile system for vertical force measurement during a seven-hour period. J Biomech. 2006;39(1):110–118. doi: 10.1016/j.jbiomech.2004.10.028.
    1. Kalpen A, Seitz P. Comparison between the force values measured with the Pedar system and Kistler platform. Proceedings of the Fourth EMED User Meeting, Ulm. Gait Posture. 1994;2(4):238–239.
    1. Forner Cordero A, Koopman HJ, van der Helm FC. Use of pressure insoles to calculate the complete ground reaction forces. J Biomech. 2004;37(9):1427–1432. doi: 10.1016/j.jbiomech.2003.12.016.
    1. Fong DT, Chan YY, Hong Y, Yung PS, Fung KY, Chan KM. Estimating the complete ground reaction forces with pressure insoles in walking. J Biomech. 2008;41(11):2597–2601. doi: 10.1016/j.jbiomech.2008.05.007.
    1. Hess RJ, Brach JS, Piva SR, VanSwearingen JM. Walking skill can be assessed in older adults: validity of the Figure-of-8 Walk Test. Phys Ther. 2010;90(1):89–99. doi: 10.2522/ptj.20080121.
    1. Owings TM, Grabiner MD. Measuring step kinematic variability on an instrumented treadmill: how many steps are enough? J Biomech. 2003;36(8):1215–1218. doi: 10.1016/S0021-9290(03)00108-8.
    1. van Uden CJ, Besser MP. Test-retest reliability of temporal and spatial gait characteristics measured with an instrumented walkway system (GAITRite) BMC Musculoskelet Disord. 2004;5:13. doi: 10.1186/1471-2474-5-13.
    1. Randolph AL, Nelson M, Akkapeddi S, Levin A, Alexandrescu R. Reliability of measurements of pressures applied on the foot during walking by a computerized insole sensor system. Arch Phys Med Rehabil. 2000;81(5):573–578. doi: 10.1016/S0003-9993(00)90037-6.
    1. Quesada P, Rash G, Jarboe N. Assessment of pedar and F-Scan revisited. Clin Biomech. 1997;12(3):S15. doi: 10.1016/S0268-0033(97)88328-3.
    1. Hsiao H, Guan J, Weatherly M. Accuracy and precision of two in-shoe pressure measurement systems. Ergonomics. 2002;45(8):537–555. doi: 10.1080/00140130210136963.
    1. Portney LG, Watkins MP. Foundations of Clinical Research: Applications to Research. 2. Upper Saddle, NJ: Prentice Hall Health; 2000.
    1. Nichols DP. Choosing an intraclass correlation coefficient. Principal Support Statistician and Manager of Statistical Support. SPSS Keywords. 1998;67:1–2.
    1. Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19(1):231–240.
    1. Silver NC, Dunlap WP. Averaging correlation coefficients: Should Fisher’s z transformation be used? J Appl Psychol. 1987;72(1):146–148.
    1. Hafer JF, Lenhoff MW, Song J, Jordan JM, Hannan MT, Hillstrom HJ. Reliability of plantar pressure platforms. Gait Posture. 2013;38(3):544–548. doi: 10.1016/j.gaitpost.2013.01.028.
    1. Murphy DF, Beynnon BD, Michelson JD, Vacek PM. Efficacy of plantar loading parameters during gait in terms of reliability, variability, effect of gender and relationship between contact area and plantar pressure. Foot Ankle Int. 2005;26(2):171–179.
    1. Segal AD, Orendurff MS, Czerniecki JM, Shofer JB, Klute GK. Local dynamic stability in turning and straight-line gait. J Biomech. 2008;41(7):1486–1493. doi: 10.1016/j.jbiomech.2008.02.012.

Source: PubMed