Accuracy, reliability, feasibility and nurse acceptance of a subcutaneous continuous glucose management system in critically ill patients: a prospective clinical trial

Tobias Wollersheim, Lilian Jo Engelhardt, Jeanne Pachulla, Rudolf Moergeli, Susanne Koch, Claudia Spies, Michael Hiesmayr, Steffen Weber-Carstens, Tobias Wollersheim, Lilian Jo Engelhardt, Jeanne Pachulla, Rudolf Moergeli, Susanne Koch, Claudia Spies, Michael Hiesmayr, Steffen Weber-Carstens

Abstract

Background: Continuous glucose monitoring (CGM) has not yet been implemented in the intensive care unit (ICU) setting. The purpose of this study was to evaluate reliability, feasibility, nurse acceptance and accuracy of the Medtronic Sentrino(®) CGM system in critically ill patients.

Methods: Sensors were inserted into the subcutaneous tissue of the patient's thigh, quantifying interstitial glucose concentration for up to 72 h per sensor. Reliability and feasibility analysis included frequency of data display, data gaps and reasons for sensor removal. We surveyed nurse acceptance in a questionnaire. For the accuracy analysis, we compared sensor values to glucose values obtained via blood gas analysis. Potential benefits of CGM were investigated in intra-individual analyses of factors, such as glycemic variability or time in target range achieved with CGM compared to that achieved with intermittent glucose monitoring.

Results: The device generated 68,655 real-time values from 31 sensors in 20 critically ill patients. 532 comparative blood glucose values were collected. Data were displayed during 32.5 h [16.0/62.4] per sensor, which is 45.1 % of the expected time of 72 h and 84.8 % of 37.9 h actual monitoring time. 21 out of 31 sensors were removed prematurely. 79.1 % of the nursing staff rated the device as not beneficial; the response rate was one-third. Mean absolute relative difference was 15.3 % (CI 13.5-17.0 %). Clarke error grid: 76.9 % zone A, 21.6 % zone B, 0.2 % zone C, 0.9 % zone D, 0.4 % zone E. Bland-Altman plot: mean bias +0.53 mg/dl, limits of agreement +64.6 and -63.5 mg/dl. Accuracy deteriorated during elevated glycemic variability and in the hyperglycemic range. There was no reduction in dysglycemic events during CGM compared to 72 h before and after CGM. If CGM was measuring accurately, it identified more hyperglycemic events when compared to intermittent measurements. This study was not designed to evaluate potential benefits of CGM on glucose control.

Conclusions: The subcutaneous CGM system did not perform with satisfactory accuracy, feasibility, or nursing acceptance when evaluated in 20 medical-surgical ICU patients. Low point accuracy and prolonged data gaps significantly limited the potential clinical usefulness of the CGM trend data. Accurate continuous data display, with a MARD < 14 %, showed potential benefits in a subgroup of our patients. Trial registration NCT02296372; Ethic vote Charité EA2/095/14.

Keywords: Accuracy; Continuous glucose monitoring; Critically ill patients; Evaluation; Feasibility; ICU; Interstitial; Medtronic Sentrino®; Nurse acceptance; Reliability; Subcutaneous.

Figures

Fig. 1
Fig. 1
Study procedure. We included n = 20 patients during 57 days of recruiting. One patient was excluded from the accuracy analysis due to a lack of comparative blood glucose samples. Ten patients required a second sensor to achieve the minimum number of comparative samples or a minimum running time of 48 h. We used an optional second sensor in one patient
Fig. 2
Fig. 2
Analysis criteria. Detailed criteria for the evaluation of subcutaneous CGM in the ICU
Fig. 3
Fig. 3
Bland–Altman plot, Clarke error grid, Color-coded Surveillance Error-Grid. n = 532 comparative samples. a Bland–Altman plot. The mean bias indicates whether there is a systematic error. Upper and lower limits were calculated by mean bias ±1.96 × standard deviation of the difference between BG and sensor glucose and represent random variations around the mean bias. If there is a Gaussian distribution, 95 % of points are located between these limits. [22, 41]. b Clarke error grid. Distribution: A = 76.9 %, B = 21.6 %, C = 0.2 %, D = 0.9 %, E = 0.4 %. Zones A (CGM data ≤20 % deviation from BG) and B are considered as clinically acceptable zones, whereas values in zones C, D and E are increasingly dangerous for the patient, and zone E may lead to adverse therapeutic decisions. [23]. c Color-coded Surveillance Error-Grid. The Surveillance Error-Grid software is available at http://www.diabetestechnology.org/SEGsoftware/Surveillance-Error-Grid-Analysis.xlsm. Last Accessed: Dec 11 2015 [24]
Fig. 4
Fig. 4
Confounding factors on MARD. a (left): Association between MARD and individual daily blood glucose variability shown in first and second standard deviation of reference glucose. First boxplot The CGM device shows acceptable accuracy* (MARD median 10.9 %) if the blood glucose variability is low (first standard deviation). Second boxplot Accuracy deteriorates (MARD median 24 %) during increased blood glucose variability (second standard deviation). b (right): Association between MARD and blood glucose ranges. Second boxplot The CGM device shows acceptable accuracy* (MARD median 8.8 %) in blood glucose ranges between 80 and 179 mg/dl. First and Third boxplots Accuracy deteriorates in the hypoglycemic range (MARD median 65.8 %) and during severe hyperglycemia (MARD median 16 %). *According to criteria specified within the consensus recommendations [20], MARD should be <14 %

References

    1. McCowen KC, Malhotra A, Bistrian BR. Stress-induced hyperglycemia. Crit Care Clin. 2001;17:107–124. doi: 10.1016/S0749-0704(05)70154-8.
    1. Preiser J-C, Ichai C, Orban J-C, Groeneveld ABJ. Metabolic response to the stress of critical illness. Br J Anaesth. 2014;113:945–954. doi: 10.1093/bja/aeu187.
    1. Weber-Carstens S, Schneider J, Wollersheim T, Assmann A, Bierbrauer J, Marg A, et al. Critical illness myopathy and GLUT4: significance of insulin and muscle contraction. Am J Respir Crit Care Med. 2013;187:387–396. doi: 10.1164/rccm.201209-1649OC.
    1. Krinsley JS. Association between hyperglycemia and increased hospital mortality in a heterogeneous population of critically ill patients. Mayo Clin Proc. 2003;78:1471–1478. doi: 10.4065/78.12.1471.
    1. Krinsley JS. Glycemic variability: a strong independent predictor of mortality in critically ill patients*. Crit Care Med. 2008;36:3008–3013. doi: 10.1097/CCM.0b013e31818b38d2.
    1. NICE-SUGAR Study Investigators. Finfer S, Liu B, Chittock DR, Norton R, Myburgh JA, et al. Hypoglycemia and risk of death in critically ill patients. N Engl J Med. 2012;367:1108–1118. doi: 10.1056/NEJMoa1204942.
    1. van den Berghe G, Wouters P, Weekers F, Verwaest C, Bruyninckx F, Schetz M, et al. Intensive insulin therapy in critically ill patients. N Engl J Med. 2001;345:1359–1367. doi: 10.1056/NEJMoa011300.
    1. Van den Berghe G, Wilmer A, Hermans G, Meersseman W, Wouters PJ, Milants I, et al. Intensive insulin therapy in the medical ICU. N Engl J Med. 2006;354:449–461. doi: 10.1056/NEJMoa052521.
    1. NICE-SUGAR Study Investigators. Finfer S, Chittock DR, Su SY-S, Blair D, Foster D, et al. Intensive versus conventional glucose control in critically ill patients. N Engl J Med. 2009;360:1283–1297. doi: 10.1056/NEJMoa0810625.
    1. Marik PE. Toward understanding tight glycemic control in the ICU: a systematic review and metaanalysis. Chest J. 2010;137:544. doi: 10.1378/chest.09-1737.
    1. Wernerman J, Desaive T, Finfer S, Foubert L, Furnary A, Holzinger U, et al. Continuous glucose control in the ICU: report of a 2013 round table meeting. Crit Care. 2014;18:226. doi: 10.1186/cc13921.
    1. Aragon D. Evaluation of nursing work effort and perceptions about blood glucose testing in tight glycemic control. Am J Crit Care. 2006;15:370–377.
    1. Boom DT, Sechterberger MK, Rijkenberg S, Kreder S, Bosman RJ, Wester JP, et al. Insulin treatment guided by subcutaneous continuous glucose monitoring compared to frequent point-of-care measurement in critically ill patients: a randomized controlled trial. Crit Care. 2014 [cited 2015 Sep 12];18. .
    1. Holzinger U, Warszawska J, Kitzberger R, Wewalka M, Miehsler W, Herkner H, et al. Real-time continuous glucose monitoring in critically ill patients. Diabetes Care. 2010;33:467–472. doi: 10.2337/dc09-1352.
    1. Brunner R, Kitzberger R, Miehsler W, Herkner H, Madl C, Holzinger U. Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients*. Crit Care Med. 2011;39:659–664. doi: 10.1097/CCM.0b013e318206bf2e.
    1. Sechterberger MK, van der Voort PHJ, Strasma PJ, DeVries JH. Accuracy of intra-arterial and subcutaneous continuous glucose monitoring in postoperative cardiac surgery patients in the ICU. J Diabetes Sci Technol. 2015;9:663–667. doi: 10.1177/1932296814564993.
    1. van Hooijdonk RT, Leopold JH, Winters T, Binnekade JM, Juffermans NP, Horn J, et al. Point accuracy and reliability of an interstitial continuous glucose-monitoring device in critically ill patients: a prospective study. Crit Care. 2015 [cited 2015 Sep 12];19. .
    1. Kosiborod M, Gottlieb RK, Sekella JA, Peterman D, Grodzinsky A, Kennedy P, et al. Performance of the medtronic sentrino continuous glucose management (CGM) system in the cardiac intensive care unit. BMJ Open Diabetes Res Care. 2014 [cited 2015 Sep 12];2. .
    1. Punke MA, Decker C, Wodack K, Reuter DA, Kluge S. Continuous glucose monitoring on the ICU using a subcutaneous sensor. Med Klin Intensivmed Notfallmedizin. 2015;110:360–363. doi: 10.1007/s00063-014-0453-1.
    1. Finfer S, Wernerman J, Preiser J-C, Cass T, Desaive T, Hovorka R, et al. Clinical review: consensus recommendations on measurement of blood glucose and reporting glycemic control in critically ill adults. Crit Care. 2013;17:229. doi: 10.1186/cc12537.
    1. Hill NR, Oliver NS, Choudhary P, Levy JC, Hindmarsh P, Matthews DR. Normal reference range for mean tissue glucose and glycemic variability derived from continuous glucose monitoring for subjects without diabetes in different ethnic groups. Diabetes Technol Ther. 2011;13:921–928. doi: 10.1089/dia.2010.0247.
    1. Bland JM, Altman DG. Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet. 1995;346:1085–1087. doi: 10.1016/S0140-6736(95)91748-9.
    1. Clarke WL, Cox D, Gonder-Frederick LA, Carter W, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10:622–628. doi: 10.2337/diacare.10.5.622.
    1. Klonoff DC, Lias C, Vigersky R, Clarke W, Parkes JL, Sacks DB, et al. The surveillance error grid. J Diabetes Sci Technol. 2014;8:658–672. doi: 10.1177/1932296814539589.
    1. Schierenbeck F, Franco-Cereceda A, Liska J. Evaluation of a continuous blood glucose monitoring system using central venous microdialysis. J Diabetes Sci Technol. 2012;6:1365–1371. doi: 10.1177/193229681200600615.
    1. van Hooijdonk RT, Winters T, Fischer JC, van Dongen-Lases EC, Krinsley JS, Preiser J-C, et al. Accuracy and limitations of continuous glucose monitoring using spectroscopy in critically ill patients. Ann Intensive Care. 2014;4:8. doi: 10.1186/2110-5820-4-8.
    1. Crane BC, Barwell NP, Gopal P, Gopichand M, Higgs T, James TD, et al. The development of a continuous intravascular glucose monitoring sensor. J Diabetes Sci Technol. 2015;9:751–761. doi: 10.1177/1932296815587937.
    1. Macken L, Flower OJ, Bird S, Hammond N, Yarad E, Bass F, et al. Continuous intra-arterial blood glucose monitoring using quenched fluorescence sensing in intensive care patients after cardiac surgery: phase II of a product development study. Crit Care Resusc J Australas Acad Crit Care Med. 2015;17:190–196.
    1. Basu A, Dube S, Slama M, Errazuriz I, Amezcua JC, Kudva YC, et al. Time lag of glucose from intravascular to interstitial compartment in humans. Diabetes. 2013;62:4083–4087. doi: 10.2337/db13-1132.
    1. Scuffi C, Lucarelli F, Valgimigli F. Minimizing the impact of time lag variability on accuracy evaluation of continuous glucose monitoring systems. J Diabetes Sci Technol. 2012;6:1383–1391. doi: 10.1177/193229681200600618.
    1. Rebrin K, Sheppard NF, Steil GM. Use of subcutaneous interstitial fluid glucose to estimate blood glucose: revisiting delay and sensor offset. J Diabetes Sci Technol. 2010;4:1087–1098. doi: 10.1177/193229681000400507.
    1. Boyne MS, Silver DM, Kaplan J, Saudek CD. Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor. Diabetes. 2003;52:2790–2794. doi: 10.2337/diabetes.52.11.2790.
    1. Cengiz E, Tamborlane WV. A tale of two compartments: interstitial versus blood glucose monitoring. Diabetes Technol Ther. 2009;11:S-11–S-16. doi: 10.1089/dia.2008.0102.
    1. Lorencio C, Leal Y, Bonet A, Bondia J, Palerm CC, Tache A, et al. Real-time continuous glucose monitoring in an intensive care unit: better accuracy in patients with septic shock. Diabetes Technol Ther. 2012;14:568–575. doi: 10.1089/dia.2012.0008.
    1. Holzinger U, Warszawska J, Kitzberger R, Herkner H, Metnitz PGH, Madl C. Impact of shock requiring norepinephrine on the accuracy and reliability of subcutaneous continuous glucose monitoring. Intensive Care Med. 2009;35:1383–1389. doi: 10.1007/s00134-009-1471-y.
    1. Siegelaar SE, Barwari T, Hermanides J, van der Voort PHJ, Hoekstra JBL, DeVries JH. Microcirculation and its relation to continuous subcutaneous glucose sensor accuracy in cardiac surgery patients in the intensive care unit. J Thorac Cardiovasc Surg. 2013;146:1283–1289. doi: 10.1016/j.jtcvs.2013.06.017.
    1. Igarashi M, Yamatani K, Fukase N, Daimon M, Ohnuma H, Takahashi H, et al. Sepsis inhibits insulin-stimulated glucose transport in isolated rat adipocytes. Diabetes Res Clin Pract. 1992;15:213–218. doi: 10.1016/0168-8227(92)90027-O.
    1. Jansson PA, Fowelin J, Smith U, Lonnroth P. Characterization by microdialysis of intracellular glucose level in subcutaneous tissue in humans. Am J Physiol Endocrinol Metab. 1988;255:E218–E220.
    1. Rebrin K, Steil GM. Can interstitial glucose assessment replace blood glucose measurements? Diabetes Technol Ther. 2000;2:461–472. doi: 10.1089/15209150050194332.
    1. Brunner R, Adelsmayr G, Herkner H, Madl C, Holzinger U. Glycemic variability and glucose complexity in critically ill patients: a retrospective analysis of continuous glucose monitoring data. Crit Care. 2012;16:R175. doi: 10.1186/cc11657.
    1. Martin Bland J, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;327:307–310. doi: 10.1016/S0140-6736(86)90837-8.
    1. Engelhardt L, Wollersheim T, Pachulla J, Mörgeli R, Balzer F, Mai K, Weber-Carstens S. Accuracy of a subcutaneous continuous glucose management system in critically ill patients. Intensive Care Med Exp. 2015;3(Suppl 1):A291. doi: 10.1186/2197-425X-3-S1-A291.

Source: PubMed

3
Abonnieren