The Mobile-Based 6-Minute Walk Test: Usability Study and Algorithm Development and Validation

Dario Salvi, Emma Poffley, Elizabeth Orchard, Lionel Tarassenko, Dario Salvi, Emma Poffley, Elizabeth Orchard, Lionel Tarassenko

Abstract

Background: The 6-min walk test (6MWT) is a convenient method for assessing functional capacity in patients with cardiopulmonary conditions. It is usually performed in the context of a hospital clinic and thus requires the involvement of hospital staff and facilities, with their associated costs.

Objective: This study aimed to develop a mobile phone-based system that allows patients to perform the 6MWT in the community.

Methods: We developed 2 algorithms to compute the distance walked during a 6MWT using sensors embedded in a mobile phone. One algorithm makes use of the global positioning system to track the location of the phone when outdoors and hence computes the distance travelled. The other algorithm is meant to be used indoors and exploits the inertial sensors built into the phone to detect U-turns when patients walk back and forth along a corridor of fixed length. We included these algorithms in a mobile phone app, integrated with wireless pulse oximeters and a back-end server. We performed Bland-Altman analysis of the difference between the distances estimated by the phone and by a reference trundle wheel on 49 indoor tests and 30 outdoor tests, with 11 different mobile phones (both Apple iOS and Google Android operating systems). We also assessed usability aspects related to the app in a discussion group with patients and clinicians using a technology acceptance model to guide discussion.

Results: The mean difference between the mobile phone-estimated distances and the reference values was -2.013 m (SD 7.84 m) for the indoor algorithm and -0.80 m (SD 18.56 m) for the outdoor algorithm. The absolute maximum difference was, in both cases, below the clinically significant threshold. A total of 2 pulmonary hypertension patients, 1 cardiologist, 2 physiologists, and 1 nurse took part in the discussion group, where issues arising from the use of the 6MWT in hospital were identified. The app was demonstrated to be usable, and the 2 patients were keen to use it in the long term.

Conclusions: The system described in this paper allows patients to perform the 6MWT at a place of their convenience. In addition, the use of pulse oximetry allows more information to be generated about the patient's health status and, possibly, be more relevant to the real-life impact of their condition. Preliminary assessment has shown that the developed 6MWT app is highly accurate and well accepted by its users. Further tests are needed to assess its clinical value.

Keywords: cardiology; digital signal processing; exercise test; global positioning system; mobile apps; pulmonary hypertension.

Conflict of interest statement

Conflicts of Interest: None declared.

©Dario Salvi, Emma Poffley, Elizabeth Orchard, Lionel Tarassenko. Originally published in JMIR mHealth and uHealth (http://mhealth.jmir.org), 03.01.2020.

Figures

Figure 1
Figure 1
Architecture of the 6-min walk test (6MWT) system. It includes 4 scenarios: (a) 6MWT at home, where patients perform a 6MWT in their home setting using their mobile phone with the app and a wireless pulse oximeter; (b) 6MWT in the hospital, where patients perform the test while being observed by a physician and with pulse oximetry data being collected through a tablet app; (c) activity tracking data retrieved by Google Fit or HealthKit transmitted for subsequent analysis; (d) data review performed by a physician through a Web interface.
Figure 2
Figure 2
Example of mobile phone azimuth signal. The first seconds show the calibration phase, after which U-turns are detected when the difference between near angles becomes greater than the set threshold within a short time window.
Figure 3
Figure 3
Example of a positioning trace (in red) retrieved from the mobile phone. The walking man figure indicates the starting point of the test; the flag indicates its end. Comparing the trace with the underlying picture shows that the position is sometimes affected by an error, for example, near tall buildings which reflect the signal or because of trees obscuring the global positioning system satellite’s signal.
Figure 4
Figure 4
Maximum and mean error of the distance estimation versus the sampling period of the localization signal selection_period computed on all available tests. The 5 seconds value minimizes both mean and maximum error.
Figure 5
Figure 5
Screenshots of the patients’ app. (a) Home page, (b) instructions about how to perform the test, (c) connection to the pulse oximeter and baseline measurements at rest, (d) estimation of the distance during walk, (e) total distance estimation and recovery at rest, (f) Borg scale questionnaire.
Figure 6
Figure 6
Screenshots of the server Web interface. (a) The form physiologists fill in when observing a 6-min walk test (6MWT), (b) an example of an outdoor 6MWT results (heart rate and oxygen saturation charts are omitted).
Figure 7
Figure 7
Bland-Altman plot of the difference between the estimated distance walked and the absolute distance. The Shapiro-Wilk test confirms the normality of the data (0.91).
Figure 8
Figure 8
Bland-Altman plot of the difference between the estimated distance walked and the ground truth. The Shapiro-Wilk test confirms the normality of the data (0.97).

References

    1. Enright PL. The six-minute walk test. Respir Care. 2003 Aug;48(8):783–5.
    1. Solway S, Brooks D, Lacasse Y, Thomas S. A qualitative systematic overview of the measurement properties of functional walk tests used in the cardiorespiratory domain. Chest. 2001 Jan;119(1):256–70. doi: 10.1378/chest.119.1.256.
    1. Miyamoto S, Nagaya N, Satoh T, Kyotani S, Sakamaki F, Fujita M, Nakanishi N, Miyatake K. Clinical correlates and prognostic significance of six-minute walk test in patients with primary pulmonary hypertension. Comparison with cardiopulmonary exercise testing. Am J Respir Crit Care Med. 2000 Feb;161(2 Pt 1):487–92. doi: 10.1164/ajrccm.161.2.9906015.
    1. Cahalin LP, Mathier MA, Semigran MJ, Dec GW, DiSalvo TG. The six-minute walk test predicts peak oxygen uptake and survival in patients with advanced heart failure. Chest. 1996 Aug;110(2):325–32. doi: 10.1378/chest.110.2.325.
    1. Casanova C, Cote C, Marin JM, Pinto-Plata V, de Torres JP, Aguirre-Jaíme A, Vassaux C, Celli BR. Distance and oxygen desaturation during the 6-min walk test as predictors of long-term mortality in patients with COPD. Chest. 2008 Oct;134(4):746–52. doi: 10.1378/chest.08-0520.
    1. Rasekaba T, Lee AL, Naughton MT, Williams TJ, Holland AE. The six-minute walk test: a useful metric for the cardiopulmonary patient. Intern Med J. 2009 Aug;39(8):495–501. doi: 10.1111/j.1445-5994.2008.01880.x.
    1. ATS Committee on Proficiency Standards for Clinical Pulmonary Function Laboratories ATS statement: guidelines for the six-minute walk test. Am J Respir Crit Care Med. 2002 Jul 1;166(1):111–7. doi: 10.1164/ajrccm.166.1.at1102.
    1. Galiè N, Hoeper MM, Humbert M, Torbicki A, Vachiery JL, Barbera JA, Beghetti M, Corris P, Gaine S, Gibbs JS, Gomez-Sanchez MA, Jondeau G, Klepetko W, Opitz C, Peacock A, Rubin L, Zellweger M, Simonneau G, ESC Committee for Practice Guidelines (CPG) Guidelines for the diagnosis and treatment of pulmonary hypertension: the task force for the diagnosis and treatment of pulmonary hypertension of the European Society of Cardiology (ESC) and the European Respiratory Society (ERS), endorsed by the International Society of Heart and Lung Transplantation (ISHLT) Eur Heart J. 2009 Oct;30(20):2493–537. doi: 10.1093/eurheartj/ehp297.
    1. Brown RG, Hwang PY. Introduction To Random Signal Analysis And Kalman Filtering. New York: Wiley; 1992.
    1. Gray AJ, Jenkins D, Andrews MH, Taaffe DR, Glover ML. Validity and reliability of GPS for measuring distance travelled in field-based team sports. J Sports Sci. 2010 Oct;28(12):1319–25. doi: 10.1080/02640414.2010.504783.
    1. Jennings D, Cormack S, Coutts AJ, Boyd L, Aughey RJ. The validity and reliability of GPS units for measuring distance in team sport specific running patterns. Int J Sports Physiol Perform. 2010 Sep;5(3):328–41. doi: 10.1123/ijspp.5.3.328.
    1. Townshend AD, Worringham CJ, Stewart IB. Assessment of speed and position during human locomotion using nondifferential GPS. Med Sci Sports Exerc. 2008 Jan;40(1):124–32. doi: 10.1249/mss.0b013e3181590bc2.
    1. Wevers LE, Kwakkel G, van de Port IG. Is outdoor use of the six-minute walk test with a global positioning system in stroke patients' own neighbourhoods reproducible and valid? J Rehabil Med. 2011 Nov;43(11):1027–31. doi: 10.2340/16501977-0881.
    1. Tao W, Liu T, Zheng R, Feng H. Gait analysis using wearable sensors. Sensors (Basel) 2012;12(2):2255–83. doi: 10.3390/s120202255.
    1. Nishiguchi S, Yamada M, Nagai K, Mori S, Kajiwara Y, Sonoda T, Yoshimura K, Yoshitomi H, Ito H, Okamoto K, Ito T, Muto S, Ishihara T, Aoyama T. Reliability and validity of gait analysis by android-based smartphone. Telemed J E Health. 2012 May;18(4):292–6. doi: 10.1089/tmj.2011.0132.
    1. Tong K, Granat MH. A practical gait analysis system using gyroscopes. Med Eng Phys. 1999 Mar;21(2):87–94. doi: 10.1016/s1350-4533(99)00030-2.
    1. Shull PB, Jirattigalachote W, Hunt MA, Cutkosky MR, Delp SL. Quantified self and human movement: a review on the clinical impact of wearable sensing and feedback for gait analysis and intervention. Gait Posture. 2014;40(1):11–9. doi: 10.1016/j.gaitpost.2014.03.189.
    1. Schimpl M, Lederer C, Daumer M. Development and validation of a new method to measure walking speed in free-living environments using the actibelt® platform. PLoS One. 2011;6(8):e23080. doi: 10.1371/journal.pone.0023080.
    1. Cheng Q, Juen J, Li Y, Prieto-Centurion V, Krishnan JA, Schatz BR. GaitTrack: Health Monitoring of Body Motion From Spatio-Temporal Parameters of Simple Smart Phones. Proceedings of the International Conference on Bioinformatics, Computational Biology and Biomedical Informatics; BCB'13; September 22-25, 2013; Washington DC, USA. 2013. pp. 897–906.
    1. Schulte J, Duc NA, Hoang DB, Elliott D, McKinley S, Nanda P. A Remote Sensor-based 6-minute Functional Walking Ability Test. Proceedings of the SENSORS, 2012 IEEE conference; SENSORS'12; October 28-31, 2012; Taipei, Taiwan. 2012.
    1. Capela NA, Lemaire ED, Baddour NC. A Smartphone Approach for the 2 and 6-minute Walk Test. Proceedings of the 2014 36th Annual International Conference of the IEEE Engineering in Medicine and Biology Society; EMBC'14; August 26-30, 2014; Chicago, IL, USA. 2014. pp. 958–61.
    1. Capela NA, Lemaire ED, Baddour N. Novel algorithm for a smartphone-based 6-minute walk test application: algorithm, application development, and evaluation. J Neuroeng Rehabil. 2015 Feb 20;12:19. doi: 10.1186/s12984-015-0013-9.
    1. Hartigan C, Kandilakis C, Dalley S, Clausen M, Wilson E, Morrison S, Etheridge S, Farris R. Mobility outcomes following five training sessions with a powered exoskeleton. Top Spinal Cord Inj Rehabil. 2015;21(2):93–9. doi: 10.1310/sci2102-93.
    1. McConnell MV, Shcherbina A, Pavlovic A, Homburger JR, Goldfeder RL, Waggot D, Cho MK, Rosenberger ME, Haskell WL, Myers J, Champagne MA, Mignot E, Landray M, Tarassenko L, Harrington RA, Yeung AC, Ashley EA. Feasibility of obtaining measures of lifestyle from a smartphone app: the MyHeart counts cardiovascular health study. JAMA Cardiol. 2017 Jan 1;2(1):67–76. doi: 10.1001/jamacardio.2016.4395.
    1. Ata R, Gandhi N, Rasmussen H, El-Gabalawy O, Gutierrez S, Ahmad A, Suresh S, Ravi R, Rothenberg K, Aalami O. Clinical validation of smartphone-based activity tracking in peripheral artery disease patients. NPJ Digit Med. 2018;1:66. doi: 10.1038/s41746-018-0073-x.
    1. Doria N, Condran B, Boulos L, Maillet DG, Dowling L, Levy A. Sharpening the focus: differentiating between focus groups for patient engagement vs qualitative research. Res Involv Engagem. 2018;4:19. doi: 10.1186/s40900-018-0102-6.
    1. Staniszewska S, Haywood KL, Brett J, Tutton L. Patient and public involvement in patient-reported outcome measures: evolution not revolution. Patient. 2012;5(2):79–87. doi: 10.2165/11597150-000000000-00000.
    1. Pulido T, Adzerikho I, Channick RN, Delcroix M, Galiè N, Ghofrani H, Jansa P, Jing Z, Le Brun F, Mehta S, Mittelholzer CM, Perchenet L, Sastry B, Sitbon O, Souza R, Torbicki A, Zeng X, Rubin LJ, Simonneau G, SERAPHIN Investigators Macitentan and morbidity and mortality in pulmonary arterial hypertension. N Engl J Med. 2013 Aug 29;369(9):809–18. doi: 10.1056/NEJMoa1213917.
    1. Rosenkranz S, Ghofrani HA, Grünig E, Klose H, Olschewski H, Hoeper MM. Cologne consensus conference on pulmonary hypertension - Update 2018. Int J Cardiol. 2018 Dec 1;272S:1–3. doi: 10.1016/j.ijcard.2018.09.064.
    1. Sun Y, Wang N, Guo X, Peng Z. Understanding the acceptance of mobile health services: a comparison and integration of alternative models. J Electron Commer Re. 2013;14(2):183–200.

Source: PubMed

3
Abonnieren