Home Use of Day-and-Night Hybrid Closed-Loop Insulin Delivery in Suboptimally Controlled Adolescents With Type 1 Diabetes: A 3-Week, Free-Living, Randomized Crossover Trial

Martin Tauschmann, Janet M Allen, Malgorzata E Wilinska, Hood Thabit, Carlo L Acerini, David B Dunger, Roman Hovorka, Martin Tauschmann, Janet M Allen, Malgorzata E Wilinska, Hood Thabit, Carlo L Acerini, David B Dunger, Roman Hovorka

Abstract

Objective: This study evaluated the feasibility, safety, and efficacy of day-and-night hybrid closed-loop insulin delivery in adolescents with type 1 diabetes under free-living conditions.

Research design and methods: In an open-label randomized crossover study, 12 suboptimally controlled adolescents on insulin pump therapy (mean ± SD age 14.6 ± 3.1 years; HbA1c 69 ± 8 mmol/mol [8.5 ± 0.7%]; duration of diabetes 7.8 ± 3.5 years) underwent two 21-day periods in which hybrid closed-loop insulin delivery was compared with sensor-augmented insulin pump therapy in random order. During the closed-loop intervention, a model predictive algorithm automatically directed insulin delivery between meals and overnight. Participants used a bolus calculator to administer prandial boluses.

Results: The proportion of time that sensor glucose was in the target range (3.9-10 mmol/L; primary end point) was increased during the closed-loop intervention compared with sensor-augmented insulin pump therapy by 18.8 ± 9.8 percentage points (mean ± SD; P < 0.001), the mean sensor glucose level was reduced by 1.8 ± 1.3 mmol/L (P = 0.001), and the time spent above target was reduced by 19.3 ± 11.3 percentage points (P < 0.001). The time spent with sensor glucose levels below 3.9 mmol/L was low and comparable between interventions (median difference 0.4 [interquartile range -2.2 to 1.3] percentage points; P = 0.33). Improved glucose control during closed-loop was associated with increased variability of basal insulin delivery (P < 0.001) and an increase in the total daily insulin dose (53.5 [39.5-72.1] vs. 51.5 [37.6-64.3] units/day; P = 0.006). Participants expressed positive attitudes and experience with the closed-loop system.

Conclusions: Free-living home use of day-and-night closed-loop in suboptimally controlled adolescents with type 1 diabetes is safe, feasible, and improves glucose control without increasing the risk of hypoglycemia. Larger and longer studies are warranted.

Trial registration: ClinicalTrials.gov NCT01873066.

© 2016 by the American Diabetes Association.

Figures

Figure 1
Figure 1
Median (interquartile range) of sensor glucose (A) and insulin delivery (B) during closed-loop (solid red line and red shaded area) and control period (dashed black line and gray shaded area) from midnight to midnight. The glucose range of 3.9 to 10.0 mmol/L is denoted by horizontal dashed lines (A).
Figure 2
Figure 2
Individual values of mean sensor glucose during day-and-night closed-loop study. The size of the bubble indicates the proportion of time spent with low glucose below 2.8 mmol/L.