A pilot randomized controlled trial of EKG for neonatal resuscitation

Anup Katheria, Kathy Arnell, Melissa Brown, Kasim Hassen, Mauricio Maldonado, Wade Rich, Neil Finer, Anup Katheria, Kathy Arnell, Melissa Brown, Kasim Hassen, Mauricio Maldonado, Wade Rich, Neil Finer

Abstract

Background: The seventh edition of the American Academy of Pediatrics Neonatal Resuscitation Program recommends the use of a cardiac monitor in infants that need resuscitation. Previous trials have shown that EKG heart rate is available before pulse rate from a pulse oximeter. To date no trial has looked at how the availability of electrocardiogram (EKG) affects clinical interventions in the delivery room.

Objective: To determine whether the availability of an EKG heart rate value and tracing to the clinical team has an effect on physiologic measures and related interventions during the stabilization of preterm infants.

Design/methods: Forty (40) premature infants enrolled in a neuro-monitoring study (The Neu-Prem Trial: NCT02605733) who had an EKG monitor available were randomized to have the heart rate information from the bedside EKG monitor either displayed or not displayed to the clinical team. Heart rate, oxygen saturation, FiO2 and mean airway pressure from a data acquisition system were recorded every 2 seconds. Results were averaged over 30 seconds and the differences analyzed using two-tailed t-test. Interventions analyzed included time to first change in FiO2, first positive pressure ventilation, first increase in airway pressure, and first intubation.

Results: There were no significant differences in time to clinical interventions between the blinded and unblinded group, despite the unblinded group having access to a visible heart rate at 66 +/- 20 compared to 114 +/- 39 seconds for the blinded group (p < .0001). Pulse rate from oximeter was lower than EKG heart rate during the first 2 minutes of life, but this was not significant.

Conclusion(s): EKG provides an earlier, and more accurate heart rate than pulse rate from an oximeter during stabilization of preterm infants, allowing earlier intervention. All interventions were started earlier in the unblinded EKG group but these numbers were not significant in this small trial. Earlier EKG placement before pulse oximeter placement may affect other interventions, but this needs further study.

Conflict of interest statement

Competing Interests: The authors have declared that no competing interests exist.

Figures

Fig 1. Consort diagram.
Fig 1. Consort diagram.
Fig 2. Heart rate in the first…
Fig 2. Heart rate in the first 5 minutes.
Comparison of EKG and Pulse Oximeter Heart Rate in the first 5 minutes. (Beats Per Minute).

References

    1. Wyckoff MH, Aziz K, Escobedo MB, Kapadia VS, Kattwinkel J, Perlman JM, et al. Part 13: Neonatal Resuscitation: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2015;132(18 Suppl 2):S543–60. Epub 2015/10/17. doi: .
    1. Phillipos E, Solevag AL, Pichler G, Aziz K, van Os S, O'Reilly M, et al. Heart Rate Assessment Immediately after Birth. Neonatology. 2016;109(2):130–8. Epub 2015/12/20. doi: .
    1. Kamlin CO, O'Donnell CP, Everest NJ, Davis PG, Morley CJ. Accuracy of clinical assessment of infant heart rate in the delivery room. Resuscitation. 2006;71(3):319–21. Epub 2006/09/23. doi: .
    1. Hawkes GA, Hawkes CP, Kenosi M, Demeulemeester J, Livingstone V, Ryan CA, et al. Auscultate, palpate and tap: time to re-evaluate. Acta paediatrica (Oslo, Norway: 1992). 2016;105(2):178–82. Epub 2015/09/01. doi: .
    1. Iglesias B, Rodriguez MJ, Aleo E, Criado E, Herranz G, Moro M, et al. [Pulse oximetry versus electrocardiogram for heart rate assessment during resuscitation of the preterm infant]. Anales de pediatria (Barcelona, Spain: 2003). 2016;84(5):271–7. Epub 2015/11/22. doi: .
    1. Dyson A, Jeffrey M, Kluckow M. Measurement of neonatal heart rate using handheld Doppler ultrasound. Arch Dis Child Fetal Neonatal Ed. 2016. Epub 2016/06/12. doi: .
    1. Marchionni P, Scalise L, Ercoli I, Tomasini EP. An optical measurement method for the simultaneous assessment of respiration and heart rates in preterm infants. The Review of scientific instruments. 2013;84(12):121705 Epub 2014/01/07. doi: .
    1. Olsson E, Ugnell H, Oberg PA, Sedin G. Photoplethysmography for simultaneous recording of heart and respiratory rates in newborn infants. Acta paediatrica (Oslo, Norway: 1992). 2000;89(7):853–61. Epub 2000/08/16. .
    1. Katheria A, Rich W, Finer N. Electrocardiogram provides a continuous heart rate faster than oximetry during neonatal resuscitation. Pediatrics. 2012;130(5):e1177–81. Epub 2012/10/24. doi: .
    1. Mizumoto H, Tomotaki S, Shibata H, Ueda K, Akashi R, Uchio H, et al. Electrocardiogram shows reliable heart rates much earlier than pulse oximetry during neonatal resuscitation. Pediatr Int. 2012;54(2):205–7. Epub 2011/11/03. doi: .
    1. van Vonderen JJ, Hooper SB, Kroese JK, Roest AA, Narayen IC, van Zwet EW, et al. Pulse oximetry measures a lower heart rate at birth compared with electrocardiography. The Journal of pediatrics. 2015;166(1):49–53. Epub 2014/12/03. doi: .
    1. Kattwinkel J, Niermeyer S, Nadkarni V, Tibballs J, Phillips B, Zideman D, et al. ILCOR advisory statement: resuscitation of the newly born infant. An advisory statement from the pediatric working group of the International Liaison Committee on Resuscitation. Circulation. 1999;99(14):1927–38. Epub 1999/04/13. .
    1. 2005 American Heart Association (AHA) guidelines for cardiopulmonary resuscitation (CPR) and emergency cardiovascular care (ECC) of pediatric and neonatal patients: pediatric basic life support. Pediatrics. 2006;117(5):e989–1004. Epub 2006/05/03. doi: .
    1. Voogdt KG, Morrison AC, Wood FE, van Elburg RM, Wyllie JP. A randomised, simulated study assessing auscultation of heart rate at birth. Resuscitation. 2010;81(8):1000–3. Epub 2010/05/21. doi: .
    1. Chitkara R, Rajani AK, Oehlert JW, Lee HC, Epi MS, Halamek LP. The accuracy of human senses in the detection of neonatal heart rate during standardized simulated resuscitation: implications for delivery of care, training and technology design. Resuscitation. 2013;84(3):369–72. Epub 2012/08/29. doi: .
    1. Owen CJ, Wyllie JP. Determination of heart rate in the baby at birth. Resuscitation. 2004;60(2):213–7. Epub 2004/03/24. doi: .