Evaluation of non-invasive continuous physiological monitoring devices for neonates in Nairobi, Kenya: a research protocol

Amy Sarah Ginsburg, Evangelyn Nkwopara, William Macharia, Roseline Ochieng, Mary Waiyego, Guohai Zhou, Roman Karasik, Shuai Xu, J Mark Ansermino, Amy Sarah Ginsburg, Evangelyn Nkwopara, William Macharia, Roseline Ochieng, Mary Waiyego, Guohai Zhou, Roman Karasik, Shuai Xu, J Mark Ansermino

Abstract

Introduction: Continuous physiological monitoring devices are often not available for monitoring high-risk neonates in low-resource settings. Easy-to-use, non-invasive, multiparameter, continuous physiological monitoring devices could be instrumental in providing appropriate care and improving outcomes for high-risk neonates in these low-resource settings.

Methods and analysis: The purpose of this prospective, observational, facility-based evaluation is to provide evidence to establish whether two existing non-invasive, multiparameter, continuous physiological monitoring devices developed by device developers, EarlySense and Sibel, can accurately and reliably measure vital signs in neonates (when compared with verified reference devices). We will also assess the feasibility, usability and acceptability of these devices for use in neonates in low-resource settings in Africa. Up to 500 neonates are enrolled in two phases: (1) a verification and accuracy evaluation phase at Aga Khan University-Nairobi and (2) a clinical feasibility evaluation phase at Pumwani Maternity Hospital in Nairobi, Kenya. Both quantitative and qualitative data are collected and analysed. Agreement between the investigational and reference devices is determined using a priori-defined accuracy thresholds.

Ethics and dissemination: This trial was approved by the Aga Khan University Nairobi Research Ethics Committee and the Western Institutional Review Board. We plan to disseminate research results in peer-reviewed journals and international conferences.

Trial registration number: NCT03920761.

Keywords: neonatology; physiology; quality in health care.

Conflict of interest statement

Competing interests: RK is employed by EarlySense and SX is employed by Sibel.

© Author(s) (or their employer(s)) 2020. Re-use permitted under CC BY. Published by BMJ.

References

    1. United Nations Inter-agency Group for Child Mortality Estimation (UN IGME) ‘Levels & Trends in Child Mortality: Report 2018, Estimates developed by the United Nations Inter-agency Group for Child Mortality Estimation’ United Nations Children’s Fund. New York, 2018.
    1. Sustainable development goals knowledge platform Division for sustainable development goals. New York, NY: United Nations Headquarters, 2019.
    1. Hug L, Alexander M, You D, et al. . National, regional, and global levels and trends in neonatal mortality between 1990 and 2017, with scenario-based projections to 2030: a systematic analysis. Lancet Glob Health 2019;7:e710–20. 10.1016/S2214-109X(19)30163-9
    1. Zhu Z, Liu T, Li G, et al. . Wearable sensor systems for infants. Sensors 2015;15:3721–49. 10.3390/s150203721
    1. Sahni R. Continuous noninvasive monitoring in the neonatal ICU. Curr Opin Pediatr 2017;29:141–8. 10.1097/MOP.0000000000000459
    1. EarlySense InSight User Guide - USA version, Doc No. MK-3602-IS-US, Revision. 3, 2017.
    1. ANNE System product requirements document, SH-PRD-Rev 2, 2019.
    1. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res 1999;8:135–60. 10.1177/096228029900800204

Source: PubMed

3
Suscribir