On the implementation of an automated acoustic output optimization algorithm for subharmonic aided pressure estimation

J K Dave, V G Halldorsdottir, J R Eisenbrey, D A Merton, J B Liu, P Machado, H Zhao, S Park, S Dianis, C L Chalek, K E Thomenius, D B Brown, F Forsberg, J K Dave, V G Halldorsdottir, J R Eisenbrey, D A Merton, J B Liu, P Machado, H Zhao, S Park, S Dianis, C L Chalek, K E Thomenius, D B Brown, F Forsberg

Abstract

Incident acoustic output (IAO) dependent subharmonic signal amplitudes from ultrasound contrast agents can be categorized into occurrence, growth or saturation stages. Subharmonic aided pressure estimation (SHAPE) is a technique that utilizes growth stage subharmonic signal amplitudes for hydrostatic pressure estimation. In this study, we developed an automated IAO optimization algorithm to identify the IAO level eliciting growth stage subharmonic signals and also studied the effect of pulse length on SHAPE. This approach may help eliminate the problems of acquiring and analyzing the data offline at all IAO levels as was done in previous studies and thus, pave the way for real-time clinical pressure monitoring applications. The IAO optimization algorithm was implemented on a Logiq 9 (GE Healthcare, Milwaukee, WI) scanner interfaced with a computer. The optimization algorithm stepped the ultrasound scanner from 0% to 100% IAO. A logistic equation fitting function was applied with the criterion of minimum least squared error between the fitted subharmonic amplitudes and the measured subharmonic amplitudes as a function of the IAO levels and the optimum IAO level was chosen corresponding to the inflection point calculated from the fitted data. The efficacy of the optimum IAO level was investigated for in vivo SHAPE to monitor portal vein (PV) pressures in 5 canines and was compared with the performance of IAO levels, below and above the optimum IAO level, for 4, 8 and 16 transmit cycles. The canines received a continuous infusion of Sonazoid microbubbles (1.5 μl/kg/min; GE Healthcare, Oslo, Norway). PV pressures were obtained using a surgically introduced pressure catheter (Millar Instruments, Inc., Houston, TX) and were recorded before and after increasing PV pressures. The experiments showed that optimum IAO levels for SHAPE in the canines ranged from 6% to 40%. The best correlation between changes in PV pressures and in subharmonic amplitudes (r=-0.76; p=0.24), and between the absolute PV pressures and the subharmonic amplitudes (r=-0.89; p<0.01) were obtained for the optimized IAO and 4 transmit cycles. Only for the optimized IAO and 4 transmit cycles did the subharmonic amplitudes differ significantly (p<0.01) before and after increasing PV pressures. A new algorithm to identify optimum IAO levels for SHAPE has been developed and validated with the best results being obtained for 4 transmit cycles. The work presented in this study may pave the way for real-time clinical applications of estimating pressures using the subharmonic signals from ultrasound contrast agents.

Copyright © 2012 Elsevier B.V. All rights reserved.

Figures

Fig. 1
Fig. 1
Schematic of the experimental setup.
Fig. 2
Fig. 2
The dual grayscale (left) and pulse inversion subharmonic (right: in the region of interest (ROI)) imaging modes. The portal vein (PV) and inferior vena cava (IVC) are marked. At the start of the IAO optimization function, the acoustic output corresponding to the subharmonic mode is set to 0 % (green arrow); due to no insonation, noise artifacts are seen in the subharmonic ROI.
Fig. 3
Fig. 3
Flowchart of the incident acoustic output optimization function.
Fig. 4
Fig. 4
Output of the incident acoustic output (IAO) optimization function (dots). The abscissa corresponds to log transformation of the IAO indicated on the scanner in percentage. The fitting function compensates for motion artifacts (mostly observed due to respiration) during IAO optimization. The optimum IAO was calculated as the IAO level with the maximum slope. In this case, the optimum IAO level (arrow) was 20 % (0.56 MPapeak-to-peak) which corresponds to 1.30 on the abscissa.
Fig 5
Fig 5
Representation of the subharmonic signal at varying incident acoustic output (IAO) levels (green arrows) used in the optimization function. (a) At 5 % IAO level the subharmonic signal begins to appear and resides at the beginning of the growth stage. (b) At 20 % IAO level the subharmonic signal resides in the linear portion of growth stage (ambient pressure sensitive stage). (c) At 60 % IAO level the subharmonic saturation begins, possibly with the occurrence of bubble destruction. (d) At 100 % IAO level the subharmonic signal is not that different from (c) indicating saturation.
Fig. 6
Fig. 6
(a), (b) and (c) depict the relationship between change in portal vein pressures and change in subharmonic signal amplitude at optimized incident acoustic output levels with 4, 8 and 16 transmit cycles – the best fit line is shown (dashed) with correlation coefficient. The best correlation was obtained with 4 transmit cycles and optimized incident acoustic output level (a). (d) The relationship between absolute portal vein pressures and subharmonic signal amplitude is shown with the best fit line (dashed) and the correlation coefficient is indicated.