Internal plate fixation of fractures: short history and recent developments

Hans K Uhthoff, Philippe Poitras, David S Backman, Hans K Uhthoff, Philippe Poitras, David S Backman

Abstract

Metal plates for internal fixation of fractures have been used for more than 100 years. Although initial shortcomings such as corrosion and insufficient strength have been overcome, more recent designs have not solved all problems. Further research is needed to develop a plate that accelerates fracture healing while not interfering with bone physiology. The introduction of rigid plates had by far the greatest impact on plate fixation of fractures. However, it led to cortical porosis, delayed bridging, and refractures after plate removal. These unwarranted effects were said to be caused by bone-plate contact interfering with cortical perfusion. Consequently, further plate modifications aimed to reduce this contact area to minimize necrosis and subsequent porosis. The advocates of limited-contact plates have not published measurements of the contact area or proof of the temporary nature of the porosis. Moreover, clinical studies of newer plate types have failed to show a superior outcome. Histomor-phometric measurements of the cortex showed no difference in the extent of necrosis under plates having different contact areas. Necrosis was predominant in the periosteal cortical half, whereas porosis occurred mostly in the endosteal cortical half. No positive correlation was found between either. The scientific evidence to date strongly suggests that bone loss is caused by stress shielding and not interference with cortical perfusion secondary to bone-plate contact. Consequently, an axially compressible plate (ACP) incorporating polylactide (PLA) inserts press-fit around screw holes was designed. The bioresorbable inserts should allow for (1) increased micromotion in the axial plane to promote healing during the union phase and (2) gradual degradation over time to decrease stress shielding during the remodeling phase. Results of ongoing experimental results are encouraging. Only plates allowing dynamic compression in the axial plane can lead to a revolution in fracture fixation.

Figures

Fig. 1
Fig. 1
Lane’s plate abandoned because of corrosion (1895). (From Bechtol CO, Fergusson AB, Laing PE. Metals and Engineering in Bone and Joint Surgery. Williams Wilkins; Baltimore: 1959. p. 20, with permission)
Fig. 2
Fig. 2
Lambotte’s plate (1909) is thin, round, and tapered at both ends
Fig. 3
Fig. 3
Structural instability of Eggers’ plate
Fig. 4
Fig. 4
Danis’ plate (1949) called “coapteur” suppresses interfragmentary motion and increases stability of fixation through interfragmentary compression achieved by tightening the side screw
Fig. 5
Fig. 5
Primary bone healing
Fig. 6
Fig. 6
Bagby and Janes’ (1956) oval holes designed for interfragmentary compression during screw tightening. (From Uhthoff HK. Current Concepts of Internal Fixation of Fractures. Heidelberg: Springer-Verlag; 1980. p. 175, with permission of Springer Science and Business Media)
Fig. 7
Fig. 7
Müller’s plate design (1965) achieves interfragmentary compression by tightening a tensioner that is temporarily anchored to the bone and the plate. (From Sequin F, Texhammar R. AO/ASIF Instrumentation. Heidelberg: Springer-Verlag; 1981. p. 72, with permission of Springer Science and Business Media)
Fig. 8
Fig. 8
Dynamic compression plate (DCP) incorporates specially designed oval holes similar to Bagby and Janes’ invention to compress bony fragments during screw tightening
Fig. 9
Fig. 9
Cortical bone loss under rigid plating
Fig. 10
Fig. 10
Railed plates (left) have limited bone-plate contact area compared to contact plates (right)
Fig. 11
Fig. 11
Decreased cortical perfusion is more pronounced with increased bone-plate contact with contact plates (left) versus railed plates (right)
Fig. 12
Fig. 12
Porosis is more pronounced in the endosteal half of the cortex of a beagle femur plated with six-hole stainless steel rigid plates. HPS ×25
Fig. 13
Fig. 13
Point-contact fixator (PC-Fix) acts like a fixator as the monocortical screws lock into the plate achieving angular stability and preventing the bone to be pulled toward the plate. (From Fernandez Dell’Oca AA, Tepic S, Frigg R, Meisser A, Haas N, Perren SM. Treating forearm fractures using an internal fixator: a prospective study. Clin Orthop 2001;389:196–205, with permission of Lippincott Williams & Wilkins)
Fig. 14
Fig. 14
Radiograph of mongrel canine femur plated with sixhole limited-contact dynamic compression plate (LC-DCP) at 16 weeks following harvest. Endosteal buildup at both ends of the plate is indicative of a high percentage load transferred through the plate
Fig. 15
Fig. 15
Axially flexible plate (AFP) for beagle femors incorporates polymethylmethacrylate (PMMA) cushions between plate and screw head to allow increased micromotion in the axial direction only
Fig. 16
Fig. 16
a Axially compressible plate (ACP). Screws (A) and outer shell (B) are titanium, and inserts (C) are polylactic acid (PLA). Inserts allow increased micromotion in the axial plane during union and reduced stress shielding during remodeling as the inserts degrade. b Cross section of ACP screw (A), outer shell (B), and insert (C). Design allows screw head to subside slightly in bone to reduce plate-bone contact

References

    1. Lane WA. Some remarks on the treatment of fractures. BMJ. 1895;1:861–3. doi: 10.1136/bmj.1.1790.861.
    1. Lambotte A. Technique et indication des prothèses dans le traitement des fractures. Presse Med. 1909;17:321.
    1. Sherman WO. Vanadium steel bone plates and screws. Surg Gynecol Obstet. 1912;14:629–34.
    1. Eggers GWN. Internal contact splint. J Bone Joint Surg Am. 1948;30:40–52.
    1. Danis R. Théorie et pratique de l’ostéosynthèse. Paris: Masson; 1949.
    1. Bagby GW, Janes JM. The effect of compression on the rate of fracture healing using a special plate. Am J Surg. 1958;95:761–71. doi: 10.1016/0002-9610(58)90625-1.
    1. Müller ME, Allgöwer M, Willenegger H. Technique of internal fixation of fractures. Berlin: Springer; 1965. Compression fixation with plates; pp. 47–51.
    1. Perren SM, Cordey J, Rahn BA, Gautier E, Schneider E. Early temporary porosis of bone induced by internal fixation implants: a reaction to necrosis, not to stress protection? Clin Orthop. 1988;232:139–51.
    1. Perren SM, Russenberger M, Steinemann S, Müller ME, Allgöwer M. A dynamic compression plate. Acta Orthop Scand. 1969;Suppl 125:31–41.
    1. Schenk R, Willenegger H. Morphological findings in primary fracture healing. In: Krompecher S, Kerner E, editors. Callus formation symposium on the biology of fracture healing. Budapest: Akadémiai Kiadó; 1967. pp. 75–86.
    1. Kessler SB, Deiler S, Schiffl-Deiler M, Uhthoff HK, Schweiberer L. Refractures: a consequence of impaired local bone viability. Arch Orthop Trauma Surg. 1992;111:96–101. doi: 10.1007/BF00443475.
    1. Berkin CR, Marshall DV. Three-sided plate fixation for fractures of the tibial and femoral shafts: a follow-up note. J Bone Joint Surg Am. 1972;54:1105–13.
    1. Gautier E, Perren SM. Die limited contact dynamic compression plate (LC-DCP): Biomechanische Forschung als Grundlage des neuen Plattendesigns. Orthopade. 1992;21:11–23.
    1. Tepic S, Perren SM. The biomechanics of the PC-Fix internal fixator. Injury. 1995;26(Suppl 2):S-B5–10.
    1. Field JR, Hearn TC, Caldwell CB. Bone plate fixation: an evaluation of interface contact area and force of the dynamic compression plate (DCP) and the limited contact-dynamic compression plate (LC-DCP) applied to cadaveric bone. J Orthop Trauma. 1997;11:368–73. doi: 10.1097/00005131-199707000-00013.
    1. Jain R, Podworny N, Hupel TM, Weinberg J, Schemitsch EH. Influence of plate design on cortical bone perfusion and fracture healing in canine segmental tibial fractures. J Orthop Trauma. 1999;13:178–86. doi: 10.1097/00005131-199903000-00005.
    1. Akeson WH, Woo SLY, Rutherford L, Coutts RD, Gonsalves M, Amiel D. The effects of rigidity of internal fixation plates on long bone remodeling. Acta Orthop Scand. 1976;47:241–9.
    1. Slätis P, Paavolainen P, Karaharju E, Holström T. Structural and biomechanical changes in bone after rigid plate fixation. In: Uhthoff HK, editor. Current concepts of internal fixation of fractures. Berlin: Springer; 1980. p. 291.
    1. Strömberg L, Dalén N, Låftman P, Sigurdsson F. Atrophy of cortical bone caused by rigid plates and its recovery. In: Uhthoff HK, editor. Current concepts of internal fixation of fractures. Berlin: Springer; 1980. pp. 289–90.
    1. Uhthoff HK, Bardos DI, Liskova-Kiar M. The advantages of titanium alloy over stainless steel plates for the internal fixation of fractures: an experimental study in dogs. J Bone Joint Surg Br. 1981;63:427–34.
    1. Uhthoff HK, Dubuc FL. Bone structure changes in the dog under rigid internal fixation. Clin Orthop. 1971;81:165–70. doi: 10.1097/00003086-197111000-00026.
    1. Uhthoff HK, Finnegan M. The effects of metal plates on post-traumatic remodelling and bone mass. J Bone Joint Surg Br. 1983;65:66–71.
    1. Gunst MA, Suter C, Rahn BA. Bone perfusion after plate osteosynthesis: a study of the intact rabbit tibia with disulfin blue vital staining. Helv Chir Acta. 1979;46:171–5.
    1. Uhthoff HK, Boisvert D, Finnegan M. Cortical porosis under plates: reaction to unloading or to necrosis? J Bone Joint Surg Am. 1994;76:1507–12.
    1. Jaworski ZFG, Liskova-Kiar M, Uhthoff HK. Regional disuse osteoporosis and factors influencing its reversal. In: Uhthoff HK, editor. Current concepts of internal fixation of fractures. Berlin: Springer; 1980. pp. 17–26.
    1. Uhthoff HK, Jaworski ZFG. Bone loss in response to long-term immobilization. J Bone Joint Surg Br. 1978;60:420–9.
    1. Korvick DL, Newbrey JW, Bagby GW, Pettit GD, Lincoln JD. Stress shielding reduced by a silicon plate-bone interface. Acta Orthop Scand. 1989;60:611–16.
    1. Perren SM. Evolution of the internal fixation of long bone fractures: the scientific basis of biological internal fixation; choosing a new balance between stability and biology. J Bone Joint Surg Br. 2002;84:1093–110. doi: 10.1302/0301-620X.84B8.13752.
    1. Andersen LD, Sisk D, Tooms RE, Park WI., III Compression-plate fixation in acute diaphyseal fractures of the radius and ulna. J Bone Joint Surg Am. 1975;57:287–97.
    1. Chapman MW, Gordon JE, Zissimos AG. Compression-plate fixation of acute fractures of the diaphyses of the radius and ulna. J Bone Joint Surg Am. 1989;71:159–69.
    1. Hertel R, Pisan M, Lambert S, Ballmer FT. Plate osteosynthesis of diaphyseal fractures of the radius and ulna. Injury. 1996;27:545–8. doi: 10.1016/S0020-1383(96)00091-5.
    1. McKee MD, Seiler JG, Jupiter JB. The application of the limited contact dynamic compression plate in the upper extremity: an analysis of 114 consecutive cases. Injury. 1995;26:661–6. doi: 10.1016/0020-1383(95)00148-4.
    1. Gupta R, Raheja A, Sharma V. Limited contact dynamic compression in diaphyseal fractures of the humerus: good outcome in 51 patients. Acta Orthop Scand. 2000;71:471–4.
    1. Gerber C, Mast JW, Ganz R. Biological internal fixation of fractures. Arch Orthop Trauma Surg. 1990;109:295–303. doi: 10.1007/BF00636165.
    1. Van Frank Haasnoot E, Münch TWH, Matter P, Perren SM. Radiological sequences of healing in internal plates and splints of different contact surface to bone (DCP, LC-DCP and PC-Fix) Injury. 1995;26(Suppl 2):S-B28–36.
    1. Haas N, Hauke C, Schütz M, Kääb M, Perren SM. Treatment of diaphyseal fractures of the forearm using the point contact fixator (PC-Fix): results of 387 fractures of a prospective multicentric study (PC-Fix II) Injury. 2001;32:S-B-51–62. doi: 10.1016/S0020-1383(01)00126-7.
    1. Hertel R, Eijer H, Meisser A, Hauke C, Perren SM. Biomechanical and biological considerations relating to the clinical use of the point contact-fixator: evaluation of the device handling test in the treatment of diaphyseal fractures of the radius and/or ulna. Injury. 2001;32:S-B-10–14. doi: 10.1016/S0020-1383(01)00121-8.
    1. Fernandez Dell’Oca AA, Masliah Galante R. Osteosynthesis of diaphyseal fractures of the radius and ulna using an internal fixator (PC-Fix): a prospective study. Injury. 2001;32:S-B-44–50.
    1. Perren SM, Buchanan JS. Basic concepts relevant to the design and development of the point contact fixator (PC-Fix) Injury. 1995;26(Suppl 2):S-B-1–4.
    1. Leung F, Chow SP. A prospective, randomized trial comparing the limited contact dynamic compression plate with the point contact fixator for forearm fractures. J Bone Joint Surg Am. 2003;85:2343–8.
    1. Schütz M, Südkamp NP. Revolution in plate osteosynthesis: new internal fixator systems. J Orthop Sci. 2003;8:252–8. doi: 10.1007/s007760300044.
    1. Ilizarov GA. Clinical application of the tension-stress effect for limb lengthening. Clin Orthop. 1990;250:8–26.
    1. Goodship AE, Kenwright J. The influence of induced micromovement upon the healing of experimental tibial fractures. J Bone Joint Surg Br. 1985;67:650–5.
    1. Foux A, Yeadon AJ, Uhthoff HK. Improved fracture healing with less rigid plates: a biomechanical study in dogs. Clin Orthop. 1997;339:232–45. doi: 10.1097/00003086-199706000-00032.
    1. Backman D, Uhthoff H, Poitras P, Schwamberger A. Mechanical performance of a fracture plate incorporating bioresorbable inserts. J Bone Joint Surg Proceed Br. 2004;86(Suppl III):300.
    1. Poitras P, Backman D, Uhthoff HK, Schwammberger A. Creep properties of a bioresorbable polymer for an axially compressible fracture plate. Presented at the 5th Combined Meeting of the Orthopaedic Research Societies of Canada, USA, Japan and Europe. Banff, AB. October 2004

Source: PubMed

3
Suscribir