END-TO-END AUDIT: COMPARISON OF TLD AND LITHIUM FORMATE EPR DOSIMETRY

Emelie Adolfsson, Paulina Wesolowska, Joanna Izewska, Eva Lund, Åsa Carlsson Tedgren, Emelie Adolfsson, Paulina Wesolowska, Joanna Izewska, Eva Lund, Åsa Carlsson Tedgren

Abstract

The aim of this study was to test two different solid state dosimetry systems for the purpose of end-to-end audits of radiotherapy volumetric modulated arc therapy (VMAT) technique; a lithium formate electron paramagnetic resonance system and a lithium fluoride thermoluminescent dosimetry system. As a complement to the solid state systems, ion chamber measurements were performed. A polystyrene phantom with a planning target volume (PTV) and an organ at risk (OAR) structure was scanned using CT. A VMAT dose plan was optimized to deliver 2 Gy to the target volume and to minimize the dose to the OAR. The different detectors were inserted into the phantom and the planned dose distribution was delivered. The measured doses were compared to the treatment planning system (TPS) calculated doses. Good agreement was found between the TPS calculated and the measured doses, well accepted for the dose determinations in remote dosimetry audits of VMAT treatment technique.

© The Author(s) 2019. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Source: PubMed