Heat distribution in the lower leg from pulsed short-wave diathermy and ultrasound treatments
C L Garrett, D O Draper, K L Knight, C L Garrett, D O Draper, K L Knight
Abstract
Objective: To compare tissue temperature rise and decay after 20-minute diathermy and ultrasound treatments.
Design and setting: We inserted 3 26-gauge thermistor microprobes into the medial aspect of the anesthetized triceps surae muscle at a depth of 3 cm and spaced 5 cm apart. Eight subjects received the diathermy treatment first, followed by the ultrasound treatment. This sequence was reversed for the remaining 8 subjects. The diathermy was applied at a frequency of 27.12 MHz at the following settings: 800 bursts per second, 400-microsecond burst duration, 850-microsecond interburst interval, peak root mean square amplitude of 150 W per burst, and an average root mean square output of 48 W per burst. The ultrasound was delivered at a frequency of 1 MHz and an intensity of 1.5 W/cm(2) in the continuous mode for 20 minutes over an area of 40 times the effective radiating area. The study was performed in a ventilated research laboratory.
Subjects: Sixteen (11 men, 5 women) healthy subjects (mean age = 23.56 +/- 4.73 years) volunteered to participate in this study.
Measurements: We recorded baseline, final, and decay temperatures for each of the 3 sites.
Results: The average temperature increases over baseline temperature after pulsed short-wave diathermy were 3.02 degrees C +/- 1.02 degrees C in site 1, 4.58 degrees C +/- 0.87 degrees C in site 2, and 3.28 degrees C +/- 1.64 degrees C in site 3. The average temperature increases over baseline temperature after ultrasound were only 0.17 degrees C +/- 0.40 degrees C, 0.09 degrees C +/- 0.56 degrees C, and -0.43 degrees C +/- 0.41 degrees C in sites 1, 2, and 3, respectively. The temperature dropped only 1 degrees C in 7.65 +/- 4.96 minutes after pulsed short-wave diathermy.
Conclusions: We conclude that pulsed short-wave diathermy was more effective than 1-MHz ultrasound in heating a large muscle mass and resulted in the muscles' retaining heat longer.
References
- J Orthop Sports Phys Ther. 1999 Jan;29(1):13-8; discussion 19-22
- J Athl Train. 1998 Apr;33(2):130-5
- J Athl Train. 1995 Oct;30(4):304-7
- Arch Phys Med Rehabil. 1959 Dec;40:510-2
- Arch Phys Med Rehabil. 1983 Dec;64(12):575-77
- J Athl Train. 1996 Apr;31(2):139-43
- J Orthop Sports Phys Ther. 1995 Oct;22(4):142-50
- Can J Physiol Pharmacol. 1991 May;69(5):704-12
- J Urol. 1984 Jul;132(1):175-8
- J Am Med Assoc. 1955 Nov 26;159(13):1286-7
Source: PubMed