Test-Retest Variability of Quick Contrast Sensitivity Function Testing
2017. április 10. frissítette: Nova Southeastern University
The contrast sensitivity function (CSF) provides a comprehensive characterization of spatial vision and predicts functional vision better than visual acuity, but long testing times prevent its psychophysical assessment in clinical applications.
Dr. Luis Lesmes et al. (2010) developed the qCSF method to obtain precise CSF measurements in only 30-50 trials using a computerized software program that presents letters on a large monitor that are identified by a patient similar to typical visual acuity testing with an eye chart.
The quick CSF method is a Bayesian adaptive method that estimates the full shape of the CSF, and the test duration is only about 3-6 minutes.
Some preliminary testing has been performed at other sites with this test in patients with vision loss due to amblyopia and glaucoma, but data from individuals without eye disease who have normal visual acuity has not been systematically collected with the latest version of this test procedure across a wide range of ages.
In addition, we aim to gain a better understanding of the typical test-retest variability that is obtained between-sessions that are about a week apart from individuals with good ocular health and visual acuity.
A tanulmány áttekintése
Állapot
Befejezve
Körülmények
Beavatkozás / kezelés
Tanulmány típusa
Megfigyelő
Beiratkozás (Tényleges)
100
Kapcsolatok és helyek
Ez a rész a vizsgálatot végzők elérhetőségeit, valamint a vizsgálat lefolytatásának helyére vonatkozó információkat tartalmazza.
Tanulmányi helyek
-
-
Florida
-
Fort Lauderdale, Florida, Egyesült Államok, 33328
- Nova Southeastern University; College of Optometry
-
-
Részvételi kritériumok
A kutatók olyan embereket keresnek, akik megfelelnek egy bizonyos leírásnak, az úgynevezett jogosultsági kritériumoknak. Néhány példa ezekre a kritériumokra a személy általános egészségi állapota vagy a korábbi kezelések.
Jogosultsági kritériumok
Tanulmányozható életkorok
20 év (Felnőtt, Idősebb felnőtt)
Egészséges önkénteseket fogad
Igen
Tanulmányozható nemek
Összes
Mintavételi módszer
Nem valószínűségi minta
Tanulmányi populáció
Individuals between the ages of 20-89 without ocular disease will be recruited from among patients seen in the Primary Eye Care Clinic at the Nova Southeastern University's The Eye Care Institute in Davie-Ft.
Lauderdale, FL, as well as faculty, students and staff at this institution.
Leírás
Inclusion Criteria:
- Ages 20-89
- Normal distance visual acuity in each of both eyes (20/20)
- Absence of ocular disease
- Able and willing to complete contrast sensitivity testing
- Provide informed consent
Exclusion Criteria:
- Any ocular disease
- Refractive errors >6D
- Inability to understand study or communicate responses (cognitive impairment)
- Unable to understand, read and speak English fluently
Tanulási terv
Ez a rész a vizsgálati terv részleteit tartalmazza, beleértve a vizsgálat megtervezését és a vizsgálat mérését.
Hogyan készül a tanulmány?
Tervezési részletek
Kohorszok és beavatkozások
Csoport / Kohorsz |
Beavatkozás / kezelés |
|---|---|
|
Normals without ocular disease
|
Mit mér a tanulmány?
Elsődleges eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
|---|---|---|
|
95% coefficient of repeatability across subjects
Időkeret: test-retest at 2 visits about a week apart
|
We will evaluate various contrast sensitivity metrics that are generated by the computer program: (i) AULCSF or area under the curve for the log contrast sensitivity function (CSF), (ii) CSF acuity or the cutoff-frequency at which sensitivity=2.0 is calculated, and (iii) contrast sensitivity at various spatial frequencies (i.e.
1.5, 3, 6, 12, 18.5 cycles per degree), obtained with testing performed binocularly, with the better eye monocularly, and with the better eye using a NoIR 4% transmission filter.
Within subject, the data will be analyzed as differences between the two values obtained at the two sessions, which can then be used to determine a 95% coefficient of repeatability across subjects.
|
test-retest at 2 visits about a week apart
|
Együttműködők és nyomozók
Itt találhatja meg a tanulmányban érintett személyeket és szervezeteket.
Szponzor
Együttműködők
Nyomozók
- Kutatásvezető: Ava K Bittner, OD, PhD, Nova Southeastern University; College of Optometry
Publikációk és hasznos linkek
A vizsgálattal kapcsolatos információk beviteléért felelős személy önkéntesen bocsátja rendelkezésre ezeket a kiadványokat. Ezek bármiről szólhatnak, ami a tanulmányhoz kapcsolódik.
Általános kiadványok
- Dorr M, Lesmes LA, Lu ZL, Bex PJ. Rapid and reliable assessment of the contrast sensitivity function on an iPad. Invest Ophthalmol Vis Sci. 2013 Nov 5;54(12):7266-73. doi: 10.1167/iovs.13-11743.
- Lesmes LA, Lu ZL, Baek J, Albright TD. Bayesian adaptive estimation of the contrast sensitivity function: the quick CSF method. J Vis. 2010 Mar 30;10(3):17.1-21. doi: 10.1167/10.3.17.
Tanulmányi rekorddátumok
Ezek a dátumok nyomon követik a ClinicalTrials.gov webhelyre benyújtott vizsgálati rekordok és összefoglaló eredmények benyújtásának folyamatát. A vizsgálati feljegyzéseket és a jelentett eredményeket a Nemzeti Orvostudományi Könyvtár (NLM) felülvizsgálja, hogy megbizonyosodjon arról, hogy megfelelnek-e az adott minőség-ellenőrzési szabványoknak, mielőtt közzéteszik őket a nyilvános weboldalon.
Tanulmány főbb dátumok
Tanulmány kezdete
2014. június 1.
Elsődleges befejezés (Tényleges)
2016. november 1.
A tanulmány befejezése (Tényleges)
2016. november 1.
Tanulmányi regisztráció dátumai
Először benyújtva
2014. június 26.
Először nyújtották be, amely megfelel a minőségbiztosítási kritériumoknak
2014. június 27.
Első közzététel (Becslés)
2014. június 30.
Tanulmányi rekordok frissítései
Utolsó frissítés közzétéve (Tényleges)
2017. április 11.
Az utolsó frissítés elküldve, amely megfelel a minőségbiztosítási kritériumoknak
2017. április 10.
Utolsó ellenőrzés
2017. április 1.
Több információ
A tanulmányhoz kapcsolódó kifejezések
Kulcsszavak
Egyéb vizsgálati azonosító számok
- 05291468Exp
- R21EY023720 (Az Egyesült Államok NIH támogatása/szerződése)
- Adaptive Sensory Technology (Egyéb támogatási/finanszírozási szám: Adaptive Sensory Technology)
Ezt az információt közvetlenül a clinicaltrials.gov webhelyről szereztük be, változtatás nélkül. Ha bármilyen kérése van vizsgálati adatainak módosítására, eltávolítására vagy frissítésére, kérjük, írjon a következő címre: register@clinicaltrials.gov. Amint a változás bevezetésre kerül a clinicaltrials.gov oldalon, ez a webhelyünkön is automatikusan frissül. .