Radiation survey around a Liac mobile electron linear accelerator for intraoperative radiation therapy

Mario Ciocca, Guido Pedroli, Roberto Orecchia, Andrea Guido, Federica Cattani, Raffaella Cambria, Umberto Veronesi, Mario Ciocca, Guido Pedroli, Roberto Orecchia, Andrea Guido, Federica Cattani, Raffaella Cambria, Umberto Veronesi

Abstract

The aim of this study was to perform a detailed analysis of the air kerma values around a Liac mobile linear accelerator working in a conventional operating room (OR) for IORT. The Liac delivers electron beams at 4, 6, 8 and 10 MeV. A radiation survey to determine photon leakage and scatter consisted of air kerma measurements on a spherical surface of 1.5 m radius, centered on the titanium exit window of the accelerating structure. Measurements were taken using a 30 cm3 calibrated cylindrical ion chamber in three orthogonal planes, at the maximum electron energy. For each point, 10 Gy was delivered. At selected points, the quality of x-ray radiation was determined by using lead sheets, and measurements were performed for all energies to investigate the energy dependence of stray radiation. The photon scatter contribution from the metallic internal patient-shielding in IORT, used to protect normal tissues underlying the target, was also evaluated. At seven locations outside the OR, the air kerma values derived from in-room measurements were compared to measurements directly performed using a survey meter. The results, for a delivered dose of 10 Gy, showed that the air kerma values ranged from approximately 6 microGy (upper and rear sides of the Liac) to 320 microGy (lateral to beam stopper) in the two orthogonal vertical planes, while values lower than 18 microGy were found in the horizontal plane. At 10 MeV, transmission behind 1 cm lead shield was found to be 42%. The use of internal shielding appeared to increase the photon scatter only slightly. Air kerma values outside the OR were generally lower than 1 mGy for an annual workload of 200 patients. Thus, the Liac can safely work in a conventional OR, while the need for additional shielding mainly depends on patient workload. Our data can be useful for centers planning to implement an IORT program using a mobile linear accelerator, permitting radiation safety personnel to estimate in advance the shielding required for a particular workload.

Figures

Figure 1
Figure 1
Side (a) and front views (b) of the Liac in parking position inside the operating room. Measurement planes (A, B, C), together with position n. 1 (0°) and direction from position n. 1 forward, are reported (not to scale). In the experimental set‐up, the Liac was positioned to simulate a standard treatment geometry, with the radiation head raised up with respect to the parking position shown. The beam stopper and an example of a movable vertical shield in front of the Liac, as well as an applicator docked to the head, are also shown.
Figure 2
Figure 2
Floor plan of the operating room and the adjacent areas. The typical working position of the Liac (outlined) as well as the points (A to F) where the air kerma values were calculated, are shown.

References

    1. Palta JR, Biggs PJ, Hazle JD, et al. Intraoperative electron beam radiation therapy: technique, dosimetry, and dose specification: report of Task Force 48 of the Radiation Therapy Committee, American Association of Physicists in Medicine. Int J Radiat Oncol Biol Phys. 1995;33(3):725–46.
    1. Beddar AS, Biggs PJ, Chang S, et al. Intraoperative radiation therapy using mobile electron linear accelerators: Report of AAPM Radiation Therapy Committee Task Group No. 72. Med Phys. 2006;33(5):1476–89.
    1. Istituto Superiore di Sanità . Guidelines for Quality Assurance in intra‐operative radiation therapy. Rosi A. and Viti V., editors. Rapporti Istisan 03/1. Roma, Italy; 2003.
    1. Mills MD, Almond PR, Boyer AL, et al. Shielding considerations for an operating room based intraoperative electron radiotherapy unit. Int J Radiat Oncol Biol Phys. 1990;18(5):1215–21.
    1. Daves JL, Mills MD. Shielding assessment of a mobile electron accelerator for intraoperative radiotherapy. J Appl Clin Med Phys. 2001;2(3):165–73.
    1. Strigari L, Soriani A, Landoni V, Teodoli S, Bruzzaniti V, Benassi M. Radiation exposure of personnel during intraoperative radiotherapy (IORT): radiation protection aspects. J Exp Clin Cancer Res. 2004;23(3):489–94.
    1. Loi G, Dominietto M, Cannillo B, et al. Neutron production from a mobile linear accelerator operating in electron mode for intraoperative radiation therapy. Phys Med Biol. 2006;51(3):695–702.
    1. Jaradat AK, Biggs PJ. Measurement of the neutron leakage from a dedicated intraoperative radiation therapy electron linear accelerator and a conventional linear accelerator for 9, 12, 15(16), and 18(20) MeV electron energies. Med Phys. 2008;35(5):1711–17.
    1. Orecchia R, Ciocca M, Lazzari R, et al. Intraoperative radiation therapy with electrons (ELIOT) in early‐stage breast cancer. Breast. 2003;12(6):483–90.
    1. Veronesi U, Orecchia R, Luini A, et al. Full‐dose intraoperative radiotherapy with electrons during breast‐conserving surgery: experience with 590 cases. Ann Surg. 2005;242(1):101–06.
    1. Ciocca M, Orecchia R, Garibaldi C, et al. In vivo dosimetry using radiochromic films during intraoperative electron beam radiation therapy in early‐stage breast cancer. Radiother Oncol. 2003;69(3):285–89.
    1. Ciocca M, Piazzi V, Lazzari R, et al. Real‐time in vivo dosimetry using micro‐MOSFET detectors during intraoperative electron beam radiation therapy in early‐stage breast cancer. Radiother Oncol. 2006;78(2):213–16.
    1. Ivaldi GB, Leonardi MC, Orecchia R, et al. Preliminary results of electron intraoperative therapy boost and hypofractionated external beam radiotherapy after breast‐conserving surgery in premenopausal women. Int J Radiat Oncol Biol Phys. 2008;72(2):485–93.
    1. National Council on Radiation Protection and Measurements . Structural shielding design and evaluation for megavoltage X‐ and gamma‐ray radiotherapy facilities. Report No. 151. Bethesda, MD: NCRP; 2005.

Source: PubMed

3
Sottoscrivi