Repeatability of intraocular pressure measurements with Icare PRO rebound, Tono-Pen AVIA, and Goldmann tonometers in sitting and reclining positions

Caterina Schweier, James V M Hanson, Jens Funk, Marc Töteberg-Harms, Caterina Schweier, James V M Hanson, Jens Funk, Marc Töteberg-Harms

Abstract

Background: Icare PRO (ICP) is a new Rebound tonometer that is able to measure intraocular pressure (IOP) in both sitting and reclining positions. In this study, the gold standard Goldmann tonometer (GAT) was compared to ICP and Tono-Pen AVIA (TPA). Hypothesis was that repeatability of GAT is superior to ICP and TPA.

Methods: 36 eyes of 36 healthy caucasian individuals, 13 male and 26 females, 17 right and 19 left eyes have been included in this prospective, randomized, cross-sectional study. The study was conducted at a single site (Dept. of Ophthalmology, University Hospital Zurich, Switzerland). Primary outcome measures were Intraclass correlation coefficients (ICC) and coefficients of variation (COV) and test-retest repeatability as visualized by Bland-Altman analysis. Secondary outcome measures were IOP in sitting (GAT, ICP and TPA) and in reclining (ICP and TPA) position.

Results: Mean IOP measured by GAT was 14.9 ± 3.5 mmHg. Mean IOP measured by ICP was 15.6 ± 3.1 mmHg (with TPA 14.8 ± 2.7 mmHg) in sitting and 16.5 ± 3.5 mmHg (with TPA 17.0 ± 3.0 mmHg) in reclining positions. COVs ranged from 2.9% (GAT) to 6.9% (ICP reclining) and ICCs from 0.819 (ICP reclining) to 0.972 (GAT).

Conclusions: Repeatability is good with all three devices. GAT has higher repeatability compared to the two tested hand-held devices with lowest COVs and highest ICCs. IOP was higher in the reclining compared to the sitting position.

Trial registration: The study was registered to the Clinical Trials Register of the US National Institute of Health, NCT01325324.

Figures

Figure 1
Figure 1
Bland-Altman plots to demonstrate repeatability in measuring IOP (mmHg) between reading 1 and 2 with all devices (GAT a, ICP b and c and TPA d and e) and positions (sitting b and d and reclining c and e). Limits of agreement were provided as 1.96-times standard deviation with upper and lower limit of the differences. Units for both axes are mmHg. (GAT = Goldmann-Applanation-Tonometer, ICP = Icare PRO, TPA = Tono-Pen AVIA, SD = Standard deviation, S = sitting, R = reclining).
Figure 2
Figure 2
Bland-Altman plots to demonstrate differences in mean IOP (mmHg) between Goldmann-Applanation-Tonometer and the two hand-held devices separately for both positions (sitting and reclining), Icare PRO (a, b) and Tono-Pen AVIA (c, d). Limits of agreement were provided as 1.96-times standard deviation with upper and lower limit of the differences. Units for both axes are mmHg. (GAT = Goldmann-Applanation-Tonometer, ICP = Icare PRO, TPA = Tono-Pen AVIA, SD = Standard deviation, M = mean, S = sitting, R = reclining).
Figure 3
Figure 3
Bland-Altman plots to demonstrate differences in IOP (mmHg) between sitting and reclining positions within the two hand-held tonometers, Icare PRO (a) and Tono-Pen AVIA (b). Limits of agreement were provided as 1.96-times standard deviation with upper and lower limit of the differences. Units for both axes are mmHg. (ICP = Icare PRO, TPA = Tono-Pen AVIA, SD = Standard deviation, M = mean, S = sitting, R = reclining).

References

    1. Goldmann H, Schmidt T. Applanation tonometry. Ophthalmologica. 1957;134(4):221–242. doi: 10.1159/000303213.
    1. Punjabi OS, Ho HK, Kniestedt C, Bostrom AG, Stamper RL, Lin SC. Intraocular pressure and ocular pulse amplitude comparisons in different types of glaucoma using dynamic contour tonometry. Curr Eye Res. 2006;31(10):851–862. doi: 10.1080/02713680600899887.
    1. Kaufmann C, Bachmann LM, Thiel MA. Comparison of dynamic contour tonometry with goldmann applanation tonometry. Invest Ophthalmol Vis Sci. 2004;45(9):3118–3121. doi: 10.1167/iovs.04-0018.
    1. Medeiros FA, Sample PA, Weinreb RN. Comparison of dynamic contour tonometry and goldmann applanation tonometry in African American subjects. Ophthalmology. 2007;114(4):658–665. doi: 10.1016/j.ophtha.2006.06.065.
    1. Huang Y, Tham CC, Zhang M. Central corneal thickness and applanation tonometry. J Cataract Refract Surg. 2008;34(3):347.
    1. Holladay JT, Allison ME, Prager TC. Goldmann applanation tonometry in patients with regular corneal astigmatism. Am J Ophthalmol. 1983;96(1):90–93.
    1. Pepose JS, Feigenbaum SK, Qazi MA, Sanderson JP, Roberts CJ. Changes in corneal biomechanics and intraocular pressure following LASIK using static, dynamic, and noncontact tonometry. Am J Ophthalmol. 2007;143(1):39–47. doi: 10.1016/j.ajo.2006.09.036.
    1. Broman AT, Congdon NG, Bandeen-Roche K, Quigley HA. Influence of corneal structure, corneal responsiveness, and other ocular parameters on tonometric measurement of intraocular pressure. J Glaucoma. 2007;16(7):581–588. doi: 10.1097/IJG.0b013e3180640f40.
    1. Iester M, Mermoud A, Achache F, Roy S. New Tonopen XL: comparison with the Goldmann tonometer. Eye (Lond) 2001;15(Pt 1):52–58.
    1. Bhartiya S, Bali SJ, Sharma R, Chaturvedi N, Dada T. Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry. Int Ophthalmol. 2011;31(4):297–302. doi: 10.1007/s10792-011-9458-4.
    1. Mackay RS, Marg E. Fast, automatic, electronic tonometers based on an exact theroy. Acta Ophthalmol (Copenh) 1959;37:495–507.
    1. Mackay RS, Marg E, Oechsli R. Automatic tonometer with exact theory: various biological applications. Science. 1960;131:1668–1669. doi: 10.1126/science.131.3414.1668.
    1. Dekking HM, Coster HD. Dynamic tonometry. Ophthalmologica. 1967;154(1):59–74. doi: 10.1159/000305149.
    1. Kontiola A. A new electromechanical method for measuring intraocular pressure. Doc Ophthalmol. 1996;93(3):265–276.
    1. Kontiola AI. A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol Scand. 2000;78(2):142–145. doi: 10.1034/j.1600-0420.2000.078002142.x.
    1. Kontiola A, Puska P. Measuring intraocular pressure with the pulsair 3000 and rebound tonometers in elderly patients without an anesthetic. Graefes Arch Clin Exp Ophthalmol. 2004;242(1):3–7. doi: 10.1007/s00417-003-0671-3.
    1. Rabe-Hesketh S, Skrondal A. In: Multilevel and Longitudinal Modeling Using STATA. 2. Rabe-Hesketh S, Skrondal A, editor. College Station, TX: STATA Press; 2008. Estimation using xtmixed; pp. 433–436.
    1. Bland JM, Altman DG. Measuring agreement in method comparison studies. Stat Methods Med Res. 1999;8(2):135–160. doi: 10.1191/096228099673819272.
    1. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–310.
    1. Buchanan RA, Williams TD. Intraocular pressure, ocular pulse pressure, and body position. Am J Optom Physiol Opt. 1985;62(1):59–62. doi: 10.1097/00006324-198501000-00008.
    1. Jorge J, Ramoa-Marques R, Lourenco A, Silva S, Nascimento S, Queiros A, Gonzalez-Meijome JM. IOP variations in the sitting and supine positions. J Glaucoma. 2010;19(9):609–612. doi: 10.1097/IJG.0b013e3181ca7ca5.
    1. Prata TS, De-Moraes CG, Kanadani FN, Ritch R, Paranhos A Jr. Posture-induced intraocular pressure changes: considerations regarding body position in glaucoma patients. Surv Ophthalmol. 2010;55(5):445–453. doi: 10.1016/j.survophthal.2009.12.002.
    1. Chui WS, Lam A, Chen D, Chiu R. The influence of corneal properties on rebound tonometry. Ophthalmology. 2008;115(1):80–84. doi: 10.1016/j.ophtha.2007.03.061.
    1. Iliev ME, Goldblum D, Katsoulis K, Amstutz C, Frueh B. Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol. 2006;90(7):833–835. doi: 10.1136/bjo.2005.089870.
    1. Martinez-de-la-Casa JM, Garcia-Feijoo J, Castillo A, Garcia-Sanchez J. Reproducibility and clinical evaluation of rebound tonometry. Invest Ophthalmol Vis Sci. 2005;46(12):4578–4580. doi: 10.1167/iovs.05-0586.
    1. Ismail AR, Lamont M, Perera S, Khan-Lim D, Mehta R, Macleod JD, Anderson DF. Comparison of IOP measurement using GAT and DCT in patients with penetrating keratoplasties. Br J Ophthalmol. 2007;91(7):980–981. doi: 10.1136/bjo.2006.099564.
    1. Moreno-Montanes J, Olmo N, Zarranz-Ventura J, Heras-Mulero H. Dynamic contour tonometry in eyes after penetrating keratoplasty. Cornea. 2009;28(7):836–837.
    1. Rootman DS, Insler MS, Thompson HW, Parelman J, Poland D, Unterman SR. Accuracy and precision of the Tono-Pen in measuring intraocular pressure after keratoplasty and epikeratophakia and in scarred corneas. Arch Ophthalmol. 1988;106(12):1697–1700. doi: 10.1001/archopht.1988.01060140869030.
    1. Ceruti P, Morbio R, Marraffa M, Marchini G. Comparison of dynamic contour tonometry and goldmann applanation tonometry in deep lamellar and penetrating keratoplasties. Am J Ophthalmol. 2008;145(2):215–221. doi: 10.1016/j.ajo.2007.10.003.
    1. Kniestedt C, Lin S, Choe J, Bostrom A, Nee M, Stamper RL. Clinical comparison of contour and applanation tonometry and their relationship to pachymetry. Arch Ophthalmol. 2005;123(11):1532–1537. doi: 10.1001/archopht.123.11.1532.
    1. Kniestedt C, Nee M, Stamper RL. Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol. 2005;243(4):359–366. doi: 10.1007/s00417-004-1024-6.
    1. Frenkel RE, Hong YJ, Shin DH. Comparison of the Tono-Pen to the Goldmann applanation tonometer. Arch Ophthalmol. 1988;106(6):750–753. doi: 10.1001/archopht.1988.01060130820030.
    1. Kao SF, Lichter PR, Bergstrom TJ, Rowe S, Musch DC. Clinical comparison of the Oculab Tono-Pen to the Goldmann applanation tonometer. Ophthalmology. 1987;94(12):1541–1544.
    1. Tonnu PA, Ho T, Sharma K, White E, Bunce C, Garway-Heath D. A comparison of four methods of tonometry: method agreement and interobserver variability. Br J Ophthalmol. 2005;89(7):847–850. doi: 10.1136/bjo.2004.056614.

Source: PubMed

3
Sottoscrivi