A high-value, low-cost bubble continuous positive airway pressure system for low-resource settings: technical assessment and initial case reports

Jocelyn Brown, Heather Machen, Kondwani Kawaza, Zondiwe Mwanza, Suzanne Iniguez, Hans Lang, Alfred Gest, Neil Kennedy, Robert Miros, Rebecca Richards-Kortum, Elizabeth Molyneux, Maria Oden, Jocelyn Brown, Heather Machen, Kondwani Kawaza, Zondiwe Mwanza, Suzanne Iniguez, Hans Lang, Alfred Gest, Neil Kennedy, Robert Miros, Rebecca Richards-Kortum, Elizabeth Molyneux, Maria Oden

Abstract

Acute respiratory infections are the leading cause of global child mortality. In the developing world, nasal oxygen therapy is often the only treatment option for babies who are suffering from respiratory distress. Without the added pressure of bubble Continuous Positive Airway Pressure (bCPAP) which helps maintain alveoli open, babies struggle to breathe and can suffer serious complications, and frequently death. A stand-alone bCPAP device can cost $6,000, too expensive for most developing world hospitals. Here, we describe the design and technical evaluation of a new, rugged bCPAP system that can be made in small volume for a cost-of-goods of approximately $350. Moreover, because of its simple design--consumer-grade pumps, medical tubing, and regulators--it requires only the simple replacement of a <$1 diaphragm approximately every 2 years for maintenance. The low-cost bCPAP device delivers pressure and flow equivalent to those of a reference bCPAP system used in the developed world. We describe the initial clinical cases of a child with bronchiolitis and a neonate with respiratory distress who were treated successfully with the new bCPAP device.

Conflict of interest statement

Competing Interests: Author Robert Miros is CEO of 3rd Stone Design. 3rd Stone is involved in developing the bCPAP device into a commercial product. Thus far, he has been contracted as part of a grant to produce 114 bCPAP devices. This does not alter the authors' adherence to all PLOS ONE policies on sharing data and materials.

Figures

Figure 1. Block diagram and photograph of…
Figure 1. Block diagram and photograph of bCPAP system.
The system consists of: (1) an adjustable flow generator; (2) a pressure-regulated delivery system; and (3) a patient interface. Flow is generated by two air pumps that can be blended with oxygen from a tank or concentrator. The total flow rate and fraction of oxygen delivered are controlled by two flow regulators. The output of the flow generator is connected to the pressure-regulated delivery system. Pressure is controlled by submerging a pressure control tube in a column of water; the mean pressure in the system is determined by the height of the water column. The patient interface is also connected to the pressure control tube, ensuring that the pressure in the patient interface and the pressure control tube are equivalent. The pressurized air mix is delivered to the patient's nostrils via a set of binasal prongs terminated at the distal end.
Figure 2. Pressure vs. time at the…
Figure 2. Pressure vs. time at the nasal prongs for two bCPAP devices.
(Left) a reference standard bCPAP device used clinically in the US and (Right) the low-cost bCPAP device. Dotted lines show the mean and average peak pressures, averaged across 60 seconds of data collection. The pressure waveforms of the two devices are similar, indicating delivery of equivalent therapeutic pressure. In both devices, the mean pressure is controlled by adjusting the height of water in the pressure control tube, and the high frequency oscillations about the mean are associated with the formation of bubbles at the distal tip of the pressure control tube. There were no statistically significant differences between the pressures generated by the two devices (Student t-test, p

Figure 3. Comparison of reference standard and…

Figure 3. Comparison of reference standard and low-cost bCPAP output pressure under different flow and…

Figure 3. Comparison of reference standard and low-cost bCPAP output pressure under different flow and pressure settings.
Each bCPAP system was assembled and nasal prong pressure was measured for 60 seconds of operation and mean pressures were calculated; results were then averaged for 10 independent trials of each system. (A) The mean pressure (mid-point of bar) and peak low and high pressures at a flow rate of 7 L/min at varying pressure settings. (B) The mean pressure (mid-point of bar) and peak low and high pressures at a pressure of 6 cm H2O and varying flow rates.

Figure 4. Vital signs for 6-month old…

Figure 4. Vital signs for 6-month old patient with bronchiolitis (a) and a neonate with…

Figure 4. Vital signs for 6-month old patient with bronchiolitis (a) and a neonate with respiratory distress (b) immediately before and after initiation of bCPAP.
(A) Time course immediately before treatment (large symbols) and after initiation of therapy (small symbols). The patient received CPAP treatment with gradually decreasing oxygen flow for 4 days, was then transitioned to nasal oxygen, and finally transitioned to room air. The patient was discharged on day 6. (B) Time course immediately before treatment (large symbols) and after initiation of therapy (small symbols). The patient received CPAP treatment for 3.5 days. The fraction of oxygen was gradually decreased to room air during the first 2 ½ days.
Figure 3. Comparison of reference standard and…
Figure 3. Comparison of reference standard and low-cost bCPAP output pressure under different flow and pressure settings.
Each bCPAP system was assembled and nasal prong pressure was measured for 60 seconds of operation and mean pressures were calculated; results were then averaged for 10 independent trials of each system. (A) The mean pressure (mid-point of bar) and peak low and high pressures at a flow rate of 7 L/min at varying pressure settings. (B) The mean pressure (mid-point of bar) and peak low and high pressures at a pressure of 6 cm H2O and varying flow rates.
Figure 4. Vital signs for 6-month old…
Figure 4. Vital signs for 6-month old patient with bronchiolitis (a) and a neonate with respiratory distress (b) immediately before and after initiation of bCPAP.
(A) Time course immediately before treatment (large symbols) and after initiation of therapy (small symbols). The patient received CPAP treatment with gradually decreasing oxygen flow for 4 days, was then transitioned to nasal oxygen, and finally transitioned to room air. The patient was discharged on day 6. (B) Time course immediately before treatment (large symbols) and after initiation of therapy (small symbols). The patient received CPAP treatment for 3.5 days. The fraction of oxygen was gradually decreased to room air during the first 2 ½ days.

References

    1. World Bank (2006) Acute Respiratory Infections. In: Jamison DT, editor. Disease and Mortality in Sub-Saharan Africa. Washington (DC): World Bank Publications, 2006. pp. 149–162.
    1. Duke T (2005) Neonatal pneumonia in developing countries. Arch Dis Child Fetal Neonatal Ed 90: F211–F219.
    1. Lawn JE, Kerber K, Enweronu-Laryea C, Cousens S (2010) 3.6 Million Neonatal Deaths – What is Progressing and What Is Not? Semin Perinatol 34: 371–386.
    1. Kamath BD, MacGuire ER, McClure EM, Goldenberg RL, Jobe AH (2011) Neonatal Mortality From Respiratory Distress Syndrome: Lessons for Low-Resource Countries. Pediatrics 127 6: 1139–1146.
    1. March of Dimes, PMNCH, Save the Children, World Health Organization (2012) Born Too Soon: The Global Action Report on Preterm Birth. In: Howson CP, Kinney MV, Lawn JE, editors. Geneva: WHO Press. pp 2–3.
    1. Nowadzky T, Pantoja A, Britton JR (2009) Bubble continuous positive airway pressure, a potentially better practice, reduces the use of mechanical ventilation among very low birth weight infants with respiratory distress syndrome. Pediatr 123 6: 1534–1540.
    1. Watson AD (2010) Premarket Notification Decision for Fisher & Paykel Healthcare Bubble CPAP System. U.S. Food and Drug Administration: 510 (k) Number K100011. Available: . Accessed 2012 Dec 6.
    1. Bonner KM, Mainous RO, Short M, Ikuta L (2008) The Nursing Care of the Infant Receiving Bubble CPAP Therapy. Adv Neonatal Care 8 2: 78–95.
    1. Enarson P, La Vincente S, Gie R, Maganga E, Chokani C (2008) Implementation of an oxygen concentrator system in district hospital paediatric wards throughout Malawi. Bull World Health Organ 86 5 344–348.
    1. Koyamaibole L, Kado J, Qovu JD, Colquhoun S, Duke T (2006) An Evaluation of Bubble-CPAP in a Neonatal Unit in a Developing Country: Effective Respiratory Support That Can Be Applied By Nurses. J Trop Pediatr 52 4: 249–253.
    1. Van Den Heuvel M, Blencowe H, Mittermayer K, Rylance S, Couperus A, et al. (2011) Introduction of bubble CPAP in a teaching hospital in Malawi. Annals of Trop Pediatr 31 1: 59–65.
    1. Urs PS, Khan F, Maiya PP (2009) Bubble CPAP - A Primary Respiratory Support for Respiratory Distress Syndrome in Newborn. Indian Pediatr 46: 409–411.
    1. Kaur C, Sema A, Beri RS, Pulivyel JM (2008) A Simple Circuit to Deliver Bubbling CPAP. Indian Pediatr 45: 312–314.
    1. Diblasi RM, Zignego JC, Smith CV, Hansen TN, Richardson CP (2010) Effective gas exchange in paralyzed juvenile rabbits using simple, inexpensive respiratory support devices. Pediatr Res 68 6: 526–30.
    1. World Health Organization (2006) Neonatal and Perinatal Mortality: Country, Regional and Global Estimates. Geneva: WHO Press. pp 22–57.

Source: PubMed

3
Sottoscrivi