Use of AlignRT System During Stereotactic Radiosurgery
調査の概要
詳細な説明
This was a prospective, single-group clinical study conducted at the Institute of Oncology Ljubljana. The study included adult patients with brain metastases who were scheduled to receive single-fraction stereotactic radiosurgery using the TrueBeam STx linear accelerator.
Each participant underwent two positioning procedures before treatment. Standard manual positioning was performed using the delta couch shift function, followed by partial CBCT verification. The participant was then positioned automatically using the AlignRT surface-guided radiotherapy system, followed by a second partial CBCT verification.
Residual positioning deviations were recorded in three translational directions (vertical, longitudinal, and lateral) and three rotational directions (rotation, pitch, and roll). The average deviations obtained after manual positioning were compared with those obtained after AlignRT-guided positioning. Additional analyses considered participant age and sex. Positioning data were obtained from the ARIA information system and analyzed using Microsoft Excel and IBM SPSS.
A total of 20 participants completed the study. No additional follow-up was required after completion of the positioning measurements.
研究の種類
入学 (実際)
段階
- 適用できない
連絡先と場所
研究場所
-
-
-
Ljubljana、スロベニア、1000
- Institute of Oncology Ljubljana
-
-
参加基準
適格基準
就学可能な年齢
- 大人
- 高齢者
健康ボランティアの受け入れ
説明
Inclusion Criteria:
- Adults aged 18 years or older
- Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
- Treatment performed using the TrueBeam STx linear accelerator
- Able and willing to provide written informed consent.
Exclusion Criteria:
- Patients younger than 18 years
- Patients receiving fractionated stereotactic radiotherapy
- Treatment performed using a linear accelerator other than TrueBeam STx.
研究計画
研究はどのように設計されていますか?
デザインの詳細
- 主な目的:他の
- 割り当て:なし
- 介入モデル:単一グループの割り当て
- マスキング:なし(オープンラベル)
武器と介入
参加者グループ / アーム |
介入・治療 |
|---|---|
|
実験的:Manual and AlignRT-Guided Positioning
Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery.
A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.
|
Participants were manually positioned from the initial setup position to the treatment isocenter using the delta couch shift function.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
Participants were automatically positioned at the treatment isocenter using the AlignRT surface-guided radiotherapy system.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
|
この研究は何を測定していますか?
主要な結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Residual Translational Setup Deviations After Manual and AlignRT-Guided Positioning
時間枠:Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography in the vertical, longitudinal, and lateral directions, expressed in centimeters.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Residual Rotational Setup Deviations After Manual and AlignRT-Guided Positioning
時間枠:Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography for rotation, pitch, and roll, expressed in degrees.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
二次結果の測定
結果測定 |
メジャーの説明 |
時間枠 |
|---|---|---|
|
Difference in Absolute Translational Setup Deviations Between AlignRT-Guided and Manual Positioning
時間枠:Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured in the vertical, longitudinal, and lateral directions after AlignRT-guided positioning and manual positioning.
Deviations were expressed in centimeters.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Difference in Absolute Rotational Setup Deviations Between AlignRT-Guided and Manual Positioning
時間枠:Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured for rotation, pitch, and roll after AlignRT-guided positioning and manual positioning.
Deviations were expressed in degrees.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
協力者と研究者
捜査官
- 主任研究者:Klemen Salmič, MSc、Institute of Oncology Ljubljana
研究記録日
主要日程の研究
研究開始 (実際)
一次修了 (実際)
研究の完了 (実際)
試験登録日
最初に提出
QC基準を満たした最初の提出物
最初の投稿 (実際)
学習記録の更新
投稿された最後の更新 (実際)
QC基準を満たした最後の更新が送信されました
最終確認日
詳しくは
この情報は、Web サイト clinicaltrials.gov から変更なしで直接取得したものです。研究の詳細を変更、削除、または更新するリクエストがある場合は、register@clinicaltrials.gov。 までご連絡ください。 clinicaltrials.gov に変更が加えられるとすぐに、ウェブサイトでも自動的に更新されます。