Use of AlignRT System During Stereotactic Radiosurgery
Studieoversigt
Status
Betingelser
Intervention / Behandling
Detaljeret beskrivelse
This was a prospective, single-group clinical study conducted at the Institute of Oncology Ljubljana. The study included adult patients with brain metastases who were scheduled to receive single-fraction stereotactic radiosurgery using the TrueBeam STx linear accelerator.
Each participant underwent two positioning procedures before treatment. Standard manual positioning was performed using the delta couch shift function, followed by partial CBCT verification. The participant was then positioned automatically using the AlignRT surface-guided radiotherapy system, followed by a second partial CBCT verification.
Residual positioning deviations were recorded in three translational directions (vertical, longitudinal, and lateral) and three rotational directions (rotation, pitch, and roll). The average deviations obtained after manual positioning were compared with those obtained after AlignRT-guided positioning. Additional analyses considered participant age and sex. Positioning data were obtained from the ARIA information system and analyzed using Microsoft Excel and IBM SPSS.
A total of 20 participants completed the study. No additional follow-up was required after completion of the positioning measurements.
Undersøgelsestype
Tilmelding (Faktiske)
Fase
- Ikke anvendelig
Kontakter og lokationer
Studiesteder
-
-
-
Ljubljana, Slovenien, 1000
- Institute of Oncology Ljubljana
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
- Voksen
- Ældre voksen
Tager imod sunde frivillige
Beskrivelse
Inclusion Criteria:
- Adults aged 18 years or older
- Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
- Treatment performed using the TrueBeam STx linear accelerator
- Able and willing to provide written informed consent.
Exclusion Criteria:
- Patients younger than 18 years
- Patients receiving fractionated stereotactic radiotherapy
- Treatment performed using a linear accelerator other than TrueBeam STx.
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Primært formål: Andet
- Tildeling: N/A
- Interventionel model: Enkelt gruppeopgave
- Maskning: Ingen (Åben etiket)
Våben og indgreb
Deltagergruppe / Arm |
Intervention / Behandling |
|---|---|
|
Eksperimentel: Manual and AlignRT-Guided Positioning
Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery.
A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.
|
Participants were manually positioned from the initial setup position to the treatment isocenter using the delta couch shift function.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
Participants were automatically positioned at the treatment isocenter using the AlignRT surface-guided radiotherapy system.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Residual Translational Setup Deviations After Manual and AlignRT-Guided Positioning
Tidsramme: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography in the vertical, longitudinal, and lateral directions, expressed in centimeters.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Residual Rotational Setup Deviations After Manual and AlignRT-Guided Positioning
Tidsramme: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography for rotation, pitch, and roll, expressed in degrees.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Difference in Absolute Translational Setup Deviations Between AlignRT-Guided and Manual Positioning
Tidsramme: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured in the vertical, longitudinal, and lateral directions after AlignRT-guided positioning and manual positioning.
Deviations were expressed in centimeters.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Difference in Absolute Rotational Setup Deviations Between AlignRT-Guided and Manual Positioning
Tidsramme: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured for rotation, pitch, and roll after AlignRT-guided positioning and manual positioning.
Deviations were expressed in degrees.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Samarbejdspartnere og efterforskere
Sponsor
Efterforskere
- Ledende efterforsker: Klemen Salmič, MSc, Institute of Oncology Ljubljana
Datoer for undersøgelser
Studer store datoer
Studiestart (Faktiske)
Primær færdiggørelse (Faktiske)
Studieafslutning (Faktiske)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Faktiske)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Yderligere relevante MeSH-vilkår
Andre undersøgelses-id-numre
- ERIDNPVO-0017/2025
Plan for individuelle deltagerdata (IPD)
Planlægger du at dele individuelle deltagerdata (IPD)?
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Studerer et amerikansk FDA-reguleret enhedsprodukt
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med Hjernemetastaser
-
Virginia Polytechnic Institute and State UniversityTrukket tilbage
-
University of MinnesotaRekruttering
-
Assistance Publique Hopitaux De MarseilleUkendt
-
BraindexClinique de la SauvegardeAfsluttetAnæstesi | Brain MonitorFrankrig
-
National Institute of Mental Health (NIMH)AfsluttetKÆLEDYR | Brain Imaging | Cannabinoid | CB1Forenede Stater
-
GE HealthcareAfsluttetBrain Imaging | Billedbehandling af hele kroppenForenede Stater
-
Mayo ClinicAfsluttetBrain Imaging | Billedbehandling af hele kroppenForenede Stater
-
Tang-Du HospitalIkke rekrutterer endnuPsykisk lidelse | Sociale medier | Brain Imaging
-
University Hospital TuebingenAfsluttetFunktionel dyspepsi | Mad | Brain ImagingTyskland
-
University of MichiganAfsluttetÆndringer i Brain Network ConnectivityForenede Stater
Kliniske forsøg med Manual Patient Positioning
-
Anderson Orthopaedic Research InstituteAfsluttetTotal hofteudskiftning
-
Vanderbilt University Medical CenterNational Institute on Deafness and Other Communication Disorders (NIDCD)Trukket tilbageAkustisk neuromForenede Stater
-
Navotek Medical, Ltd.AfsluttetProstatakræftBelgien, Holland
-
Kirsehir Ahi Evran UniversityAfsluttet
-
Sherry StewartUniversity of British Columbia; University of Victoria; Nova Scotia Health... og andre samarbejdspartnereRekruttering
-
Florida International UniversityJames and Esther King Biomedical Research ProgramAktiv, ikke rekrutterendeTobaksbrugsforstyrrelse | Vaping Ophør | Vaping teenagere | Vaping adfærdForenede Stater
-
National Taiwan University HospitalAktiv, ikke rekrutterendeÆLDRE MENNESKER | Lungekræft (NSCLC) | Avanceret plejeplanlægning (ACP)Taiwan
-
Alexandria UniversityRekrutteringPostoperative smerterEgypten
-
University of ZurichSwiss National Science FoundationAfsluttetDyadisk Peer-vejledning | Vedhæftningspriming | RessourceprimingSchweiz
-
University of La LagunaAktiv, ikke rekrutterende