Use of AlignRT System During Stereotactic Radiosurgery

15 juli 2026 uppdaterad av: Institute of Oncology Ljubljana
This prospective study evaluated whether the AlignRT surface-guided radiotherapy system improves patient positioning accuracy during stereotactic radiosurgery (SRS) for brain metastases compared with standard manual positioning. Twenty adults scheduled for single-fraction SRS were positioned twice before treatment. Participants were first positioned using the standard manual method with delta couch shift and were then positioned automatically using the AlignRT system. After each positioning method, a partial cone-beam computed tomography (CBCT) scan was performed to measure residual setup deviations in three translational and three rotational directions. The primary objective was to compare the average setup deviations between the two positioning methods and determine whether AlignRT provided more accurate patient positioning. The additional positioning and imaging procedure prolonged the treatment visit by up to 15 minutes and involved a low additional imaging dose.

Studieöversikt

Detaljerad beskrivning

This was a prospective, single-group clinical study conducted at the Institute of Oncology Ljubljana. The study included adult patients with brain metastases who were scheduled to receive single-fraction stereotactic radiosurgery using the TrueBeam STx linear accelerator.

Each participant underwent two positioning procedures before treatment. Standard manual positioning was performed using the delta couch shift function, followed by partial CBCT verification. The participant was then positioned automatically using the AlignRT surface-guided radiotherapy system, followed by a second partial CBCT verification.

Residual positioning deviations were recorded in three translational directions (vertical, longitudinal, and lateral) and three rotational directions (rotation, pitch, and roll). The average deviations obtained after manual positioning were compared with those obtained after AlignRT-guided positioning. Additional analyses considered participant age and sex. Positioning data were obtained from the ARIA information system and analyzed using Microsoft Excel and IBM SPSS.

A total of 20 participants completed the study. No additional follow-up was required after completion of the positioning measurements.

Studietyp

Interventionell

Inskrivning (Faktisk)

20

Fas

  • Inte tillämpbar

Kontakter och platser

Det här avsnittet innehåller kontaktuppgifter för dem som genomför studien och information om var denna studie genomförs.

Studieorter

      • Ljubljana, Slovenien, 1000
        • Institute of Oncology Ljubljana

Deltagandekriterier

Forskare letar efter personer som passar en viss beskrivning, så kallade behörighetskriterier. Några exempel på dessa kriterier är en persons allmänna hälsotillstånd eller tidigare behandlingar.

Urvalskriterier

Åldrar som är berättigade till studier

  • Vuxen
  • Äldre vuxen

Tar emot friska volontärer

Nej

Beskrivning

Inclusion Criteria:

  • Adults aged 18 years or older
  • Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
  • Treatment performed using the TrueBeam STx linear accelerator
  • Able and willing to provide written informed consent.

Exclusion Criteria:

  • Patients younger than 18 years
  • Patients receiving fractionated stereotactic radiotherapy
  • Treatment performed using a linear accelerator other than TrueBeam STx.

Studieplan

Det här avsnittet ger detaljer om studieplanen, inklusive hur studien är utformad och vad studien mäter.

Hur är studien utformad?

Designdetaljer

  • Primärt syfte: Övrig
  • Tilldelning: N/A
  • Interventionsmodell: Enskild gruppuppgift
  • Maskning: Ingen (Open Label)

Vapen och interventioner

Deltagargrupp / Arm
Intervention / Behandling
Experimentell: Manual and AlignRT-Guided Positioning
Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery. A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.
Participants were manually positioned from the initial setup position to the treatment isocenter using the delta couch shift function. A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
Participants were automatically positioned at the treatment isocenter using the AlignRT surface-guided radiotherapy system. A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.

Vad mäter studien?

Primära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Residual Translational Setup Deviations After Manual and AlignRT-Guided Positioning
Tidsram: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Residual setup deviations measured by partial cone-beam computed tomography in the vertical, longitudinal, and lateral directions, expressed in centimeters. Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Residual Rotational Setup Deviations After Manual and AlignRT-Guided Positioning
Tidsram: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Residual setup deviations measured by partial cone-beam computed tomography for rotation, pitch, and roll, expressed in degrees. Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.

Sekundära resultatmått

Resultatmått
Åtgärdsbeskrivning
Tidsram
Difference in Absolute Translational Setup Deviations Between AlignRT-Guided and Manual Positioning
Tidsram: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Within-participant comparison of the absolute residual setup deviations measured in the vertical, longitudinal, and lateral directions after AlignRT-guided positioning and manual positioning. Deviations were expressed in centimeters. Smaller absolute deviations indicate greater positioning accuracy.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Difference in Absolute Rotational Setup Deviations Between AlignRT-Guided and Manual Positioning
Tidsram: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Within-participant comparison of the absolute residual setup deviations measured for rotation, pitch, and roll after AlignRT-guided positioning and manual positioning. Deviations were expressed in degrees. Smaller absolute deviations indicate greater positioning accuracy.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.

Samarbetspartners och utredare

Det är här du hittar personer och organisationer som är involverade i denna studie.

Utredare

  • Huvudutredare: Klemen Salmič, MSc, Institute of Oncology Ljubljana

Studieavstämningsdatum

Dessa datum spårar framstegen för inlämningar av studieposter och sammanfattande resultat till ClinicalTrials.gov. Studieposter och rapporterade resultat granskas av National Library of Medicine (NLM) för att säkerställa att de uppfyller specifika kvalitetskontrollstandarder innan de publiceras på den offentliga webbplatsen.

Studera stora datum

Studiestart (Faktisk)

1 april 2025

Primärt slutförande (Faktisk)

31 augusti 2025

Avslutad studie (Faktisk)

31 augusti 2025

Studieregistreringsdatum

Först inskickad

15 juli 2026

Först inskickad som uppfyllde QC-kriterierna

15 juli 2026

Första postat (Faktisk)

21 juli 2026

Uppdateringar av studier

Senaste uppdatering publicerad (Faktisk)

21 juli 2026

Senaste inskickade uppdateringen som uppfyllde QC-kriterierna

15 juli 2026

Senast verifierad

1 juli 2026

Mer information

Termer relaterade till denna studie

Plan för individuella deltagardata (IPD)

Planerar du att dela individuella deltagardata (IPD)?

NEJ

Läkemedels- och apparatinformation, studiedokument

Studerar en amerikansk FDA-reglerad läkemedelsprodukt

Nej

Studerar en amerikansk FDA-reglerad produktprodukt

Nej

Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .

Prenumerera