Use of AlignRT System During Stereotactic Radiosurgery
Обзор исследования
Статус
Условия
Вмешательство/лечение
Подробное описание
This was a prospective, single-group clinical study conducted at the Institute of Oncology Ljubljana. The study included adult patients with brain metastases who were scheduled to receive single-fraction stereotactic radiosurgery using the TrueBeam STx linear accelerator.
Each participant underwent two positioning procedures before treatment. Standard manual positioning was performed using the delta couch shift function, followed by partial CBCT verification. The participant was then positioned automatically using the AlignRT surface-guided radiotherapy system, followed by a second partial CBCT verification.
Residual positioning deviations were recorded in three translational directions (vertical, longitudinal, and lateral) and three rotational directions (rotation, pitch, and roll). The average deviations obtained after manual positioning were compared with those obtained after AlignRT-guided positioning. Additional analyses considered participant age and sex. Positioning data were obtained from the ARIA information system and analyzed using Microsoft Excel and IBM SPSS.
A total of 20 participants completed the study. No additional follow-up was required after completion of the positioning measurements.
Тип исследования
Регистрация (Действительный)
Фаза
- Непригодный
Контакты и местонахождение
Места учебы
-
-
-
Ljubljana, Словения, 1000
- Institute of Oncology Ljubljana
-
-
Критерии участия
Критерии приемлемости
Возраст, подходящий для обучения
- Взрослый
- Пожилой взрослый
Принимает здоровых добровольцев
Описание
Inclusion Criteria:
- Adults aged 18 years or older
- Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
- Treatment performed using the TrueBeam STx linear accelerator
- Able and willing to provide written informed consent.
Exclusion Criteria:
- Patients younger than 18 years
- Patients receiving fractionated stereotactic radiotherapy
- Treatment performed using a linear accelerator other than TrueBeam STx.
Учебный план
Как устроено исследование?
Детали дизайна
- Основная цель: Другой
- Распределение: Н/Д
- Интервенционная модель: Одногрупповое задание
- Маскировка: Нет (открытая этикетка)
Оружие и интервенции
Группа участников / Армия |
Вмешательство/лечение |
|---|---|
|
Экспериментальный: Manual and AlignRT-Guided Positioning
Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery.
A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.
|
Participants were manually positioned from the initial setup position to the treatment isocenter using the delta couch shift function.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
Participants were automatically positioned at the treatment isocenter using the AlignRT surface-guided radiotherapy system.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
|
Что измеряет исследование?
Первичные показатели результатов
Мера результата |
Мера Описание |
Временное ограничение |
|---|---|---|
|
Residual Translational Setup Deviations After Manual and AlignRT-Guided Positioning
Временное ограничение: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography in the vertical, longitudinal, and lateral directions, expressed in centimeters.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Residual Rotational Setup Deviations After Manual and AlignRT-Guided Positioning
Временное ограничение: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography for rotation, pitch, and roll, expressed in degrees.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Вторичные показатели результатов
Мера результата |
Мера Описание |
Временное ограничение |
|---|---|---|
|
Difference in Absolute Translational Setup Deviations Between AlignRT-Guided and Manual Positioning
Временное ограничение: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured in the vertical, longitudinal, and lateral directions after AlignRT-guided positioning and manual positioning.
Deviations were expressed in centimeters.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Difference in Absolute Rotational Setup Deviations Between AlignRT-Guided and Manual Positioning
Временное ограничение: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured for rotation, pitch, and roll after AlignRT-guided positioning and manual positioning.
Deviations were expressed in degrees.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Соавторы и исследователи
Спонсор
Следователи
- Главный следователь: Klemen Salmič, MSc, Institute of Oncology Ljubljana
Даты записи исследования
Изучение основных дат
Начало исследования (Действительный)
Первичное завершение (Действительный)
Завершение исследования (Действительный)
Даты регистрации исследования
Первый отправленный
Впервые представлено, что соответствует критериям контроля качества
Первый опубликованный (Действительный)
Обновления учебных записей
Последнее опубликованное обновление (Действительный)
Последнее отправленное обновление, отвечающее критериям контроля качества
Последняя проверка
Дополнительная информация
Термины, связанные с этим исследованием
Дополнительные соответствующие термины MeSH
Другие идентификационные номера исследования
- ERIDNPVO-0017/2025
Планирование данных отдельных участников (IPD)
Планируете делиться данными об отдельных участниках (IPD)?
Информация о лекарствах и устройствах, исследовательские документы
Изучает лекарственный продукт, регулируемый FDA США.
Изучает продукт устройства, регулируемый Управлением по санитарному надзору за качеством пищевых продуктов и медикаментов США.
Эта информация была получена непосредственно с веб-сайта clinicaltrials.gov без каких-либо изменений. Если у вас есть запросы на изменение, удаление или обновление сведений об исследовании, обращайтесь по адресу register@clinicaltrials.gov. Как только изменение будет реализовано на clinicaltrials.gov, оно будет автоматически обновлено и на нашем веб-сайте. .