Use of AlignRT System During Stereotactic Radiosurgery
Przegląd badań
Status
Warunki
Interwencja / Leczenie
Szczegółowy opis
This was a prospective, single-group clinical study conducted at the Institute of Oncology Ljubljana. The study included adult patients with brain metastases who were scheduled to receive single-fraction stereotactic radiosurgery using the TrueBeam STx linear accelerator.
Each participant underwent two positioning procedures before treatment. Standard manual positioning was performed using the delta couch shift function, followed by partial CBCT verification. The participant was then positioned automatically using the AlignRT surface-guided radiotherapy system, followed by a second partial CBCT verification.
Residual positioning deviations were recorded in three translational directions (vertical, longitudinal, and lateral) and three rotational directions (rotation, pitch, and roll). The average deviations obtained after manual positioning were compared with those obtained after AlignRT-guided positioning. Additional analyses considered participant age and sex. Positioning data were obtained from the ARIA information system and analyzed using Microsoft Excel and IBM SPSS.
A total of 20 participants completed the study. No additional follow-up was required after completion of the positioning measurements.
Typ studiów
Zapisy (Rzeczywisty)
Faza
- Nie dotyczy
Kontakty i lokalizacje
Lokalizacje studiów
-
-
-
Ljubljana, Słowenia, 1000
- Institute of Oncology Ljubljana
-
-
Kryteria uczestnictwa
Kryteria kwalifikacji
Wiek uprawniający do nauki
- Dorosły
- Starszy dorosły
Akceptuje zdrowych ochotników
Opis
Inclusion Criteria:
- Adults aged 18 years or older
- Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
- Treatment performed using the TrueBeam STx linear accelerator
- Able and willing to provide written informed consent.
Exclusion Criteria:
- Patients younger than 18 years
- Patients receiving fractionated stereotactic radiotherapy
- Treatment performed using a linear accelerator other than TrueBeam STx.
Plan studiów
Jak projektuje się badanie?
Szczegóły projektu
- Główny cel: Inny
- Przydział: Nie dotyczy
- Model interwencyjny: Zadanie dla jednej grupy
- Maskowanie: Brak (otwarta etykieta)
Broń i interwencje
Grupa uczestników / Arm |
Interwencja / Leczenie |
|---|---|
|
Eksperymentalny: Manual and AlignRT-Guided Positioning
Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery.
A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.
|
Participants were manually positioned from the initial setup position to the treatment isocenter using the delta couch shift function.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
Participants were automatically positioned at the treatment isocenter using the AlignRT surface-guided radiotherapy system.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
|
Co mierzy badanie?
Podstawowe miary wyniku
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Residual Translational Setup Deviations After Manual and AlignRT-Guided Positioning
Ramy czasowe: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography in the vertical, longitudinal, and lateral directions, expressed in centimeters.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Residual Rotational Setup Deviations After Manual and AlignRT-Guided Positioning
Ramy czasowe: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography for rotation, pitch, and roll, expressed in degrees.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Miary wyników drugorzędnych
Miara wyniku |
Opis środka |
Ramy czasowe |
|---|---|---|
|
Difference in Absolute Translational Setup Deviations Between AlignRT-Guided and Manual Positioning
Ramy czasowe: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured in the vertical, longitudinal, and lateral directions after AlignRT-guided positioning and manual positioning.
Deviations were expressed in centimeters.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Difference in Absolute Rotational Setup Deviations Between AlignRT-Guided and Manual Positioning
Ramy czasowe: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured for rotation, pitch, and roll after AlignRT-guided positioning and manual positioning.
Deviations were expressed in degrees.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Współpracownicy i badacze
Sponsor
Śledczy
- Główny śledczy: Klemen Salmič, MSc, Institute of Oncology Ljubljana
Daty zapisu na studia
Główne daty studiów
Rozpoczęcie studiów (Rzeczywisty)
Zakończenie podstawowe (Rzeczywisty)
Ukończenie studiów (Rzeczywisty)
Daty rejestracji na studia
Pierwszy przesłany
Pierwszy przesłany, który spełnia kryteria kontroli jakości
Pierwszy wysłany (Rzeczywisty)
Aktualizacje rekordów badań
Ostatnia wysłana aktualizacja (Rzeczywisty)
Ostatnia przesłana aktualizacja, która spełniała kryteria kontroli jakości
Ostatnia weryfikacja
Więcej informacji
Terminy związane z tym badaniem
Dodatkowe istotne warunki MeSH
Inne numery identyfikacyjne badania
- ERIDNPVO-0017/2025
Plan dla danych uczestnika indywidualnego (IPD)
Planujesz udostępniać dane poszczególnych uczestników (IPD)?
Informacje o lekach i urządzeniach, dokumenty badawcze
Bada produkt leczniczy regulowany przez amerykańską FDA
Bada produkt urządzenia regulowany przez amerykańską FDA
Te informacje zostały pobrane bezpośrednio ze strony internetowej clinicaltrials.gov bez żadnych zmian. Jeśli chcesz zmienić, usunąć lub zaktualizować dane swojego badania, skontaktuj się z register@clinicaltrials.gov. Gdy tylko zmiana zostanie wprowadzona na stronie clinicaltrials.gov, zostanie ona automatycznie zaktualizowana również na naszej stronie internetowej .
Badania kliniczne na Manual Patient Positioning
-
Fenerbahce UniversityZakończonyPrzepuklina dysku szyjnegoIndyk
-
Medipol UniversityJeszcze nie rekrutacjaBól mięśniowo-twarzowy
-
Cliniques universitaires Saint-Luc- Université...Université Catholique de LouvainJeszcze nie rekrutacjaWirtualna rzeczywistość | Mózg | EEG | Sprawność manualna | fNIRSBelgia
-
Ohio State University Comprehensive Cancer CenterZakończonyAnatomiczny rak piersi w stadium I AJCC v8 | Rak piersi w stadium anatomicznym IA AJCC v8 | Rak piersi w stadium anatomicznym IB AJCC v8 | Anatomiczny rak piersi w stadium II AJCC v8 | Rak piersi w stadium anatomicznym IIA AJCC v8 | Rak piersi w stadium anatomicznym IIB AJCC v8 | Anatomiczny rak piersi... i inne warunkiStany Zjednoczone
-
Brigham and Women's HospitalPatient-Centered Outcomes Research InstituteAktywny, nie rekrutującyOtyłość | NadwagaStany Zjednoczone
-
University of Colorado, DenverNational Institute of Nursing Research (NINR); Kaiser Permanente; Denver Health...ZakończonyOpieka paliatywnaStany Zjednoczone
-
University of California, San FranciscoBristol-Myers Squibb FoundationZakończonyRak płuc w stadium III AJCC v8 | Rak płuc w stadium II AJCC v8 | Rak płuca w stadium IIA AJCC v8 | Rak płuca w stadium IIB AJCC v8 | Rak płuc w stadium IIIA AJCC v8 | Rak płuc w stadium IIIB AJCC v8 | Nowotwór złośliwy | Rak jelita grubego w stadium III AJCC v8 | Stadium IIIA Rak jelita grubego AJCC v8 | Stadium... i inne warunkiStany Zjednoczone
-
Brigham and Women's HospitalNational Institutes of Health (NIH); National Library of Medicine (NLM)NieznanyNowotwór | Utrata słuchu | Kardiomiopatia przerostowaStany Zjednoczone, Kanada
-
The Christie NHS Foundation TrustManchester University NHS Foundation Trust; Aptus Clinical Ltd.; Zenzium Ltd.Zakończony
-
Chinese University of Hong KongJeszcze nie rekrutacjaGruczolak i rak jelita grubegoChiny