Use of AlignRT System During Stereotactic Radiosurgery
Přehled studie
Postavení
Podmínky
Intervence / Léčba
Detailní popis
This was a prospective, single-group clinical study conducted at the Institute of Oncology Ljubljana. The study included adult patients with brain metastases who were scheduled to receive single-fraction stereotactic radiosurgery using the TrueBeam STx linear accelerator.
Each participant underwent two positioning procedures before treatment. Standard manual positioning was performed using the delta couch shift function, followed by partial CBCT verification. The participant was then positioned automatically using the AlignRT surface-guided radiotherapy system, followed by a second partial CBCT verification.
Residual positioning deviations were recorded in three translational directions (vertical, longitudinal, and lateral) and three rotational directions (rotation, pitch, and roll). The average deviations obtained after manual positioning were compared with those obtained after AlignRT-guided positioning. Additional analyses considered participant age and sex. Positioning data were obtained from the ARIA information system and analyzed using Microsoft Excel and IBM SPSS.
A total of 20 participants completed the study. No additional follow-up was required after completion of the positioning measurements.
Typ studie
Zápis (Aktuální)
Fáze
- Nelze použít
Kontakty a umístění
Studijní místa
-
-
-
Ljubljana, Slovinsko, 1000
- Institute of Oncology Ljubljana
-
-
Kritéria účasti
Kritéria způsobilosti
Věk způsobilý ke studiu
- Dospělý
- Starší dospělý
Přijímá zdravé dobrovolníky
Popis
Inclusion Criteria:
- Adults aged 18 years or older
- Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
- Treatment performed using the TrueBeam STx linear accelerator
- Able and willing to provide written informed consent.
Exclusion Criteria:
- Patients younger than 18 years
- Patients receiving fractionated stereotactic radiotherapy
- Treatment performed using a linear accelerator other than TrueBeam STx.
Studijní plán
Jak je studie koncipována?
Detaily designu
- Primární účel: Jiný
- Přidělení: N/A
- Intervenční model: Přiřazení jedné skupiny
- Maskování: Žádné (otevřený štítek)
Zbraně a zásahy
Skupina účastníků / Arm |
Intervence / Léčba |
|---|---|
|
Experimentální: Manual and AlignRT-Guided Positioning
Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery.
A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.
|
Participants were manually positioned from the initial setup position to the treatment isocenter using the delta couch shift function.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
Participants were automatically positioned at the treatment isocenter using the AlignRT surface-guided radiotherapy system.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
|
Co je měření studie?
Primární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Residual Translational Setup Deviations After Manual and AlignRT-Guided Positioning
Časové okno: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography in the vertical, longitudinal, and lateral directions, expressed in centimeters.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Residual Rotational Setup Deviations After Manual and AlignRT-Guided Positioning
Časové okno: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography for rotation, pitch, and roll, expressed in degrees.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Sekundární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Difference in Absolute Translational Setup Deviations Between AlignRT-Guided and Manual Positioning
Časové okno: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured in the vertical, longitudinal, and lateral directions after AlignRT-guided positioning and manual positioning.
Deviations were expressed in centimeters.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Difference in Absolute Rotational Setup Deviations Between AlignRT-Guided and Manual Positioning
Časové okno: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured for rotation, pitch, and roll after AlignRT-guided positioning and manual positioning.
Deviations were expressed in degrees.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Spolupracovníci a vyšetřovatelé
Sponzor
Vyšetřovatelé
- Vrchní vyšetřovatel: Klemen Salmič, MSc, Institute of Oncology Ljubljana
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Aktuální)
Primární dokončení (Aktuální)
Dokončení studie (Aktuální)
Termíny zápisu do studia
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Více informací
Termíny související s touto studií
Další relevantní podmínky MeSH
Další identifikační čísla studie
- ERIDNPVO-0017/2025
Plán pro data jednotlivých účastníků (IPD)
Plánujete sdílet data jednotlivých účastníků (IPD)?
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .
Klinické studie na Metastázy v mozku
-
Neurolutions, Inc.Zatím nenabírámeMrtvice | Hemiparéza po mrtvici | Rozhraní Brain Computer
-
The Hong Kong Polytechnic UniversityDokončenoMrtvice | Rozhraní Brain ComputerHongkong
-
Uludag UniversityDokončenoPorucha pozornosti s hyperaktivitou (ADHD) | Rozhraní Brain ComputerTurecko (Türkiye)
-
Wright State UniversityAktivní, ne náborBolest hlavy | Bolest hlavy vyvolaná chladem | Brain Freeze | Cephalgie vyvolaná chlademSpojené státy
-
Neurolutions, Inc.Eunice Kennedy Shriver National Institute of Child Health and Human Development...DokončenoMrtvice | Hemiparéza | Spasticita jako pokračování mrtvice | Rozhraní Brain ComputerSpojené státy
Klinické studie na Manual Patient Positioning
-
New York UniversityNew York State Psychiatric InstituteDokončeno
-
Bezmialem Vakif UniversityZápis na pozvánkuDentofaciální deformity | Dentofaciální anomálie [včetně malocclusion] a další poruchy čelisti (M26-M27)Krocan
-
Clinical Centre of SerbiaCenterline Biomedical, Inc.Zatím nenabírámeAneuryzma břišní aorty | Vystavení záření | Aneuryzma břišní aorty bez rupturySrbsko
-
Chang Gung Memorial HospitalNáborMrtvice | Kardiovaskulární onemocnění | Diabetes | HyperlipidémieTchaj-wan
-
University of La LagunaAktivní, ne nábor
-
Peshawar Medical CollegePakistan Association of Cognitive TherapistsDokončenoDeprese | Infarkt myokarduPákistán
-
GE HealthcareUkončenoKrevní tlak (nízký, normální, vysoký)Spojené státy
-
Universitaire Ziekenhuizen KU LeuvenOslo University Hospital; University of Bordeaux; KU Leuven; University College... a další spolupracovníciStaženoGeriatrie | Lékařská onkologie | Kritické cesty | Digitální technologieBelgie
-
National Taipei University of Nursing and Health...DokončenoVylepšená upravená jídla ke zlepšení kvality života předkřehkých poruch žvýkání u starších dospělýchKvalita života | Spokojenost | Výběr potravin | Žvýkací nemocTchaj-wan
-
Peshawar Medical CollegePakistan Association of Cognitive TherapistsDokončenoPostnatální depresePákistán