Use of AlignRT System During Stereotactic Radiosurgery

15 juli 2026 bijgewerkt door: Institute of Oncology Ljubljana
This prospective study evaluated whether the AlignRT surface-guided radiotherapy system improves patient positioning accuracy during stereotactic radiosurgery (SRS) for brain metastases compared with standard manual positioning. Twenty adults scheduled for single-fraction SRS were positioned twice before treatment. Participants were first positioned using the standard manual method with delta couch shift and were then positioned automatically using the AlignRT system. After each positioning method, a partial cone-beam computed tomography (CBCT) scan was performed to measure residual setup deviations in three translational and three rotational directions. The primary objective was to compare the average setup deviations between the two positioning methods and determine whether AlignRT provided more accurate patient positioning. The additional positioning and imaging procedure prolonged the treatment visit by up to 15 minutes and involved a low additional imaging dose.

Studie Overzicht

Gedetailleerde beschrijving

This was a prospective, single-group clinical study conducted at the Institute of Oncology Ljubljana. The study included adult patients with brain metastases who were scheduled to receive single-fraction stereotactic radiosurgery using the TrueBeam STx linear accelerator.

Each participant underwent two positioning procedures before treatment. Standard manual positioning was performed using the delta couch shift function, followed by partial CBCT verification. The participant was then positioned automatically using the AlignRT surface-guided radiotherapy system, followed by a second partial CBCT verification.

Residual positioning deviations were recorded in three translational directions (vertical, longitudinal, and lateral) and three rotational directions (rotation, pitch, and roll). The average deviations obtained after manual positioning were compared with those obtained after AlignRT-guided positioning. Additional analyses considered participant age and sex. Positioning data were obtained from the ARIA information system and analyzed using Microsoft Excel and IBM SPSS.

A total of 20 participants completed the study. No additional follow-up was required after completion of the positioning measurements.

Studietype

Ingrijpend

Inschrijving (Werkelijk)

20

Fase

  • Niet toepasbaar

Contacten en locaties

In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.

Studie Locaties

      • Ljubljana, Slovenië, 1000
        • Institute of Oncology Ljubljana

Deelname Criteria

Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.

Geschiktheidscriteria

Leeftijden die in aanmerking komen voor studie

  • Volwassen
  • Oudere volwassene

Accepteert gezonde vrijwilligers

Nee

Beschrijving

Inclusion Criteria:

  • Adults aged 18 years or older
  • Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
  • Treatment performed using the TrueBeam STx linear accelerator
  • Able and willing to provide written informed consent.

Exclusion Criteria:

  • Patients younger than 18 years
  • Patients receiving fractionated stereotactic radiotherapy
  • Treatment performed using a linear accelerator other than TrueBeam STx.

Studie plan

Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.

Hoe is de studie opgezet?

Ontwerpdetails

  • Primair doel: Ander
  • Toewijzing: NVT
  • Interventioneel model: Opdracht voor een enkele groep
  • Masker: Geen (open label)

Wapens en interventies

Deelnemersgroep / Arm
Interventie / Behandeling
Experimenteel: Manual and AlignRT-Guided Positioning
Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery. A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.
Participants were manually positioned from the initial setup position to the treatment isocenter using the delta couch shift function. A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
Participants were automatically positioned at the treatment isocenter using the AlignRT surface-guided radiotherapy system. A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.

Wat meet het onderzoek?

Primaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Residual Translational Setup Deviations After Manual and AlignRT-Guided Positioning
Tijdsspanne: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Residual setup deviations measured by partial cone-beam computed tomography in the vertical, longitudinal, and lateral directions, expressed in centimeters. Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Residual Rotational Setup Deviations After Manual and AlignRT-Guided Positioning
Tijdsspanne: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Residual setup deviations measured by partial cone-beam computed tomography for rotation, pitch, and roll, expressed in degrees. Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.

Secundaire uitkomstmaten

Uitkomstmaat
Maatregel Beschrijving
Tijdsspanne
Difference in Absolute Translational Setup Deviations Between AlignRT-Guided and Manual Positioning
Tijdsspanne: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Within-participant comparison of the absolute residual setup deviations measured in the vertical, longitudinal, and lateral directions after AlignRT-guided positioning and manual positioning. Deviations were expressed in centimeters. Smaller absolute deviations indicate greater positioning accuracy.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Difference in Absolute Rotational Setup Deviations Between AlignRT-Guided and Manual Positioning
Tijdsspanne: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Within-participant comparison of the absolute residual setup deviations measured for rotation, pitch, and roll after AlignRT-guided positioning and manual positioning. Deviations were expressed in degrees. Smaller absolute deviations indicate greater positioning accuracy.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.

Medewerkers en onderzoekers

Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.

Onderzoekers

  • Hoofdonderzoeker: Klemen Salmič, MSc, Institute of Oncology Ljubljana

Studie record data

Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.

Bestudeer belangrijke data

Studie start (Werkelijk)

1 april 2025

Primaire voltooiing (Werkelijk)

31 augustus 2025

Studie voltooiing (Werkelijk)

31 augustus 2025

Studieregistratiedata

Eerst ingediend

15 juli 2026

Eerst ingediend dat voldeed aan de QC-criteria

15 juli 2026

Eerst geplaatst (Werkelijk)

21 juli 2026

Updates van studierecords

Laatste update geplaatst (Werkelijk)

21 juli 2026

Laatste update ingediend die voldeed aan QC-criteria

15 juli 2026

Laatst geverifieerd

1 juli 2026

Meer informatie

Termen gerelateerd aan deze studie

Plan Individuele Deelnemersgegevens (IPD)

Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?

NEE

Informatie over medicijnen en apparaten, studiedocumenten

Bestudeert een door de Amerikaanse FDA gereguleerd geneesmiddel

Nee

Bestudeert een door de Amerikaanse FDA gereguleerd apparaatproduct

Nee

Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .

Abonneren