Use of AlignRT System During Stereotactic Radiosurgery
Studieoversikt
Status
Forhold
Intervensjon / Behandling
Detaljert beskrivelse
This was a prospective, single-group clinical study conducted at the Institute of Oncology Ljubljana. The study included adult patients with brain metastases who were scheduled to receive single-fraction stereotactic radiosurgery using the TrueBeam STx linear accelerator.
Each participant underwent two positioning procedures before treatment. Standard manual positioning was performed using the delta couch shift function, followed by partial CBCT verification. The participant was then positioned automatically using the AlignRT surface-guided radiotherapy system, followed by a second partial CBCT verification.
Residual positioning deviations were recorded in three translational directions (vertical, longitudinal, and lateral) and three rotational directions (rotation, pitch, and roll). The average deviations obtained after manual positioning were compared with those obtained after AlignRT-guided positioning. Additional analyses considered participant age and sex. Positioning data were obtained from the ARIA information system and analyzed using Microsoft Excel and IBM SPSS.
A total of 20 participants completed the study. No additional follow-up was required after completion of the positioning measurements.
Studietype
Registrering (Faktiske)
Fase
- Ikke aktuelt
Kontakter og plasseringer
Studiesteder
-
-
-
Ljubljana, Slovenia, 1000
- Institute of Oncology Ljubljana
-
-
Deltakelseskriterier
Kvalifikasjonskriterier
Alder som er kvalifisert for studier
- Voksen
- Eldre voksen
Tar imot friske frivillige
Beskrivelse
Inclusion Criteria:
- Adults aged 18 years or older
- Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
- Treatment performed using the TrueBeam STx linear accelerator
- Able and willing to provide written informed consent.
Exclusion Criteria:
- Patients younger than 18 years
- Patients receiving fractionated stereotactic radiotherapy
- Treatment performed using a linear accelerator other than TrueBeam STx.
Studieplan
Hvordan er studiet utformet?
Designdetaljer
- Primært formål: Annen
- Tildeling: N/A
- Intervensjonsmodell: Enkeltgruppeoppdrag
- Masking: Ingen (Open Label)
Våpen og intervensjoner
Deltakergruppe / Arm |
Intervensjon / Behandling |
|---|---|
|
Eksperimentell: Manual and AlignRT-Guided Positioning
Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery.
A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.
|
Participants were manually positioned from the initial setup position to the treatment isocenter using the delta couch shift function.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
Participants were automatically positioned at the treatment isocenter using the AlignRT surface-guided radiotherapy system.
A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
|
Hva måler studien?
Primære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
Residual Translational Setup Deviations After Manual and AlignRT-Guided Positioning
Tidsramme: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography in the vertical, longitudinal, and lateral directions, expressed in centimeters.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Residual Rotational Setup Deviations After Manual and AlignRT-Guided Positioning
Tidsramme: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Residual setup deviations measured by partial cone-beam computed tomography for rotation, pitch, and roll, expressed in degrees.
Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Sekundære resultatmål
Resultatmål |
Tiltaksbeskrivelse |
Tidsramme |
|---|---|---|
|
Difference in Absolute Translational Setup Deviations Between AlignRT-Guided and Manual Positioning
Tidsramme: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured in the vertical, longitudinal, and lateral directions after AlignRT-guided positioning and manual positioning.
Deviations were expressed in centimeters.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
|
Difference in Absolute Rotational Setup Deviations Between AlignRT-Guided and Manual Positioning
Tidsramme: Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Within-participant comparison of the absolute residual setup deviations measured for rotation, pitch, and roll after AlignRT-guided positioning and manual positioning.
Deviations were expressed in degrees.
Smaller absolute deviations indicate greater positioning accuracy.
|
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
|
Samarbeidspartnere og etterforskere
Sponsor
Etterforskere
- Hovedetterforsker: Klemen Salmič, MSc, Institute of Oncology Ljubljana
Studierekorddatoer
Studer hoveddatoer
Studiestart (Faktiske)
Primær fullføring (Faktiske)
Studiet fullført (Faktiske)
Datoer for studieregistrering
Først innsendt
Først innsendt som oppfylte QC-kriteriene
Først lagt ut (Faktiske)
Oppdateringer av studieposter
Sist oppdatering lagt ut (Faktiske)
Siste oppdatering sendt inn som oppfylte QC-kriteriene
Sist bekreftet
Mer informasjon
Begreper knyttet til denne studien
Ytterligere relevante MeSH-vilkår
Andre studie-ID-numre
- ERIDNPVO-0017/2025
Plan for individuelle deltakerdata (IPD)
Planlegger du å dele individuelle deltakerdata (IPD)?
Legemiddel- og utstyrsinformasjon, studiedokumenter
Studerer et amerikansk FDA-regulert medikamentprodukt
Studerer et amerikansk FDA-regulert enhetsprodukt
Denne informasjonen ble hentet direkte fra nettstedet clinicaltrials.gov uten noen endringer. Hvis du har noen forespørsler om å endre, fjerne eller oppdatere studiedetaljene dine, vennligst kontakt register@clinicaltrials.gov. Så snart en endring er implementert på clinicaltrials.gov, vil denne også bli oppdatert automatisk på nettstedet vårt. .