Use of AlignRT System During Stereotactic Radiosurgery

2026年7月15日 更新者:Institute of Oncology Ljubljana
This prospective study evaluated whether the AlignRT surface-guided radiotherapy system improves patient positioning accuracy during stereotactic radiosurgery (SRS) for brain metastases compared with standard manual positioning. Twenty adults scheduled for single-fraction SRS were positioned twice before treatment. Participants were first positioned using the standard manual method with delta couch shift and were then positioned automatically using the AlignRT system. After each positioning method, a partial cone-beam computed tomography (CBCT) scan was performed to measure residual setup deviations in three translational and three rotational directions. The primary objective was to compare the average setup deviations between the two positioning methods and determine whether AlignRT provided more accurate patient positioning. The additional positioning and imaging procedure prolonged the treatment visit by up to 15 minutes and involved a low additional imaging dose.

研究概览

详细说明

This was a prospective, single-group clinical study conducted at the Institute of Oncology Ljubljana. The study included adult patients with brain metastases who were scheduled to receive single-fraction stereotactic radiosurgery using the TrueBeam STx linear accelerator.

Each participant underwent two positioning procedures before treatment. Standard manual positioning was performed using the delta couch shift function, followed by partial CBCT verification. The participant was then positioned automatically using the AlignRT surface-guided radiotherapy system, followed by a second partial CBCT verification.

Residual positioning deviations were recorded in three translational directions (vertical, longitudinal, and lateral) and three rotational directions (rotation, pitch, and roll). The average deviations obtained after manual positioning were compared with those obtained after AlignRT-guided positioning. Additional analyses considered participant age and sex. Positioning data were obtained from the ARIA information system and analyzed using Microsoft Excel and IBM SPSS.

A total of 20 participants completed the study. No additional follow-up was required after completion of the positioning measurements.

研究类型

介入性

注册 (实际的)

20

阶段

  • 不适用

联系人和位置

本节提供了进行研究的人员的详细联系信息,以及有关进行该研究的地点的信息。

学习地点

      • Ljubljana、斯洛文尼亚、1000
        • Institute of Oncology Ljubljana

参与标准

研究人员寻找符合特定描述的人,称为资格标准。这些标准的一些例子是一个人的一般健康状况或先前的治疗。

资格标准

适合学习的年龄

  • 成人
  • 年长者

接受健康志愿者

不

描述

Inclusion Criteria:

  • Adults aged 18 years or older
  • Patients with brain metastases eligible for single-fraction stereotactic radiosurgery
  • Treatment performed using the TrueBeam STx linear accelerator
  • Able and willing to provide written informed consent.

Exclusion Criteria:

  • Patients younger than 18 years
  • Patients receiving fractionated stereotactic radiotherapy
  • Treatment performed using a linear accelerator other than TrueBeam STx.

学习计划

本节提供研究计划的详细信息,包括研究的设计方式和研究的衡量标准。

研究是如何设计的?

设计细节

  • 主要用途:其他
  • 分配:不适用
  • 介入模型:单组作业
  • 屏蔽:无(打开标签)

武器和干预

参与者组/臂
干预/治疗
实验性的:Manual and AlignRT-Guided Positioning
Participants underwent standard manual positioning followed by automatic positioning using the AlignRT system before stereotactic radiosurgery. A partial cone-beam computed tomography scan was performed after each positioning method to measure residual translational and rotational setup deviations.
Participants were manually positioned from the initial setup position to the treatment isocenter using the delta couch shift function. A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.
Participants were automatically positioned at the treatment isocenter using the AlignRT surface-guided radiotherapy system. A partial cone-beam computed tomography scan was then performed to measure residual setup deviations in three translational and three rotational directions.

研究衡量的是什么?

主要结果指标

结果测量
措施说明
大体时间
Residual Translational Setup Deviations After Manual and AlignRT-Guided Positioning
大体时间:Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Residual setup deviations measured by partial cone-beam computed tomography in the vertical, longitudinal, and lateral directions, expressed in centimeters. Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Residual Rotational Setup Deviations After Manual and AlignRT-Guided Positioning
大体时间:Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Residual setup deviations measured by partial cone-beam computed tomography for rotation, pitch, and roll, expressed in degrees. Mean deviations were compared between manual positioning and AlignRT-guided positioning within the same participants.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.

次要结果测量

结果测量
措施说明
大体时间
Difference in Absolute Translational Setup Deviations Between AlignRT-Guided and Manual Positioning
大体时间:Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Within-participant comparison of the absolute residual setup deviations measured in the vertical, longitudinal, and lateral directions after AlignRT-guided positioning and manual positioning. Deviations were expressed in centimeters. Smaller absolute deviations indicate greater positioning accuracy.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Difference in Absolute Rotational Setup Deviations Between AlignRT-Guided and Manual Positioning
大体时间:Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.
Within-participant comparison of the absolute residual setup deviations measured for rotation, pitch, and roll after AlignRT-guided positioning and manual positioning. Deviations were expressed in degrees. Smaller absolute deviations indicate greater positioning accuracy.
Immediately after each positioning procedure and before stereotactic radiosurgery, during the same treatment visit.

合作者和调查者

在这里您可以找到参与这项研究的人员和组织。

调查人员

  • 首席研究员:Klemen Salmič, MSc、Institute of Oncology Ljubljana

研究记录日期

这些日期跟踪向 ClinicalTrials.gov 提交研究记录和摘要结果的进度。研究记录和报告的结果由国家医学图书馆 (NLM) 审查,以确保它们在发布到公共网站之前符合特定的质量控制标准。

研究主要日期

学习开始 (实际的)

2025年4月1日

初级完成 (实际的)

2025年8月31日

研究完成 (实际的)

2025年8月31日

研究注册日期

首次提交

2026年7月15日

首先提交符合 QC 标准的

2026年7月15日

首次发布 (实际的)

2026年7月21日

研究记录更新

最后更新发布 (实际的)

2026年7月21日

上次提交的符合 QC 标准的更新

2026年7月15日

最后验证

2026年7月1日

更多信息

与本研究相关的术语

计划个人参与者数据 (IPD)

计划共享个人参与者数据 (IPD)?

不

药物和器械信息、研究文件

研究美国 FDA 监管的药品

不

研究美国 FDA 监管的设备产品

不

此信息直接从 clinicaltrials.gov 网站检索,没有任何更改。如果您有任何更改、删除或更新研究详细信息的请求,请联系 register@clinicaltrials.gov. clinicaltrials.gov 上实施更改,我们的网站上也会自动更新.

订阅