Galilei Lens Professional vs. Predicate Devices: a Comparison Study
GALILEI Lens Professional vs. Predicate Devices: a Comparison Study
Studie Overzicht
Toestand
Conditie
Interventie / Behandeling
Gedetailleerde beschrijving
Background: An accessory to the commercially available GALILEI Dual-Scheimpflug Corneal Topographer/Keratometer with the name "EBR Accessory" was developed by the company Ziemer Ophthalmic Systems AG for the measurement of intraocular distances. The combination is called "GALILEI Lens Professional". The measurement principle is based on on short coherence interferometry/-reflectometry, which has gained wide-spread application and has been applied clinically, including one of the members of this study group, for the precise measurement of axial, intraocular distances (12.1.). Such EBR Accessory measurements, in combination with measurements by the GALILEI, may be applied to given intra-ocular lens (IOL) types and IOL equations, an permit the calculation of suggested IOLs that are implanted during cataract surgery, in order to achieve the desired vision correction. To register this accessory for sale, the precision and agreement to existing devices with similar applications ("predicate devices") must be demonstrated. One complete measurement consists of a cornea-topography/keratometry scan, followed immediately by three consecutive, axial biometry scans of the anterior segment (cornea and crystalline lens) and - after a short re-alignment - three consecutive, axial biometry scans of the retina. The measurement process is continuous, but divided in two steps. Blinking by the patient is possible in between the two steps, which increases patient comfort and, in certain cases, measurement success. Precision is evaluated by three repeated, complete measurements with the GALILEI Lens Professional and each predicate device. Two standard devices with comparable optical technology that are used in routine cataract surgery were declared as predicate devices. These two devise are typically used in combination with standard keratometers, as which the GALILEI Lens Professional may be employed as well.
Goal: The comparison of the GALILEI Lens Professional to predicate devices with respect to precision in measuring postoperative, calculated target refractions in cataract patients, and versatility of measurable quantities.
Studietype
Inschrijving (Werkelijk)
Fase
- Niet toepasbaar
Contacten en locaties
Studie Locaties
-
-
Aargau
-
Reinach, Aargau, Zwitserland, 5734
- Eye Clinic Orasis
-
-
Deelname Criteria
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
Accepteert gezonde vrijwilligers
Geslachten die in aanmerking komen voor studie
Beschrijving
Inclusion Criteria:
- spherical equivalent between -10 D and +10 D
- best-corrected VA of 20/100 or better.
Exclusion Criteria:
- strabism
- blepharitis
- nystagmus
- amblyopia
- anisometropia (spherical equivalent > 1D)
- angle closure glaucoma
- seizure disorder
- brain damages
- Down syndrome
- trisomy 13 or 18
- cerebral palsy
- other serious disorders of the eye or central nervous systems
Studie plan
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Diagnostisch
- Toewijzing: Niet-gerandomiseerd
- Interventioneel model: Opdracht voor een enkele groep
- Masker: Enkel
Wapens en interventies
Deelnemersgroep / Arm |
Interventie / Behandeling |
|---|---|
|
Experimenteel: Normal
Arm: Normal eyes Interventions: Galilei Lens Professional, IOLMaster, Lenstar
|
Measurement device
Measurement device
Measurement device
|
|
Actieve vergelijker: Mild Cataract
Arm: Eyes with mild cataract Interventions: Galilei Lens Professional, IOLMaster, Lenstar
|
Measurement device
Measurement device
Measurement device
|
|
Experimenteel: Severe cataract
Arm: Eyes with severe cataract Interventions: Galilei Lens Professional, IOLMaster, Lenstar
|
Measurement device
Measurement device
Measurement device
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Agreement With Predicate Devices in Terms of Agreement 1
Tijdsspanne: 3 months
|
Assessed parameters: Axial Length (AL; G6, IOLM, LS) Central Corneal Thickness (CCT; G6, LS)) Anterior Chamber Depth (ACD; G6, IOLM, LS) Lens Thickness (LT; G6, LS) Corneal Curvature (SimK; G6, IOLM, LS) White-to-White (WtW; G6, IOL, LS) |
3 months
|
|
Agreement With Predicate Devices in Terms of Agreement 2
Tijdsspanne: 3 months
|
Assessed Parameters: Simulated Corneal Curvature (SimK; G6, IOLM, LS) |
3 months
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Agreement With Devices of the Same Type in Terms of Repeatability
Tijdsspanne: 3 months
|
Assessed parameters: Axial Length (AL) Central Corneal Thickness (CCT) Anterior Chamber Depth (ACD) Lens Thickness (LT) Corneal Curvature (SimK) White-to-White (WtW) |
3 months
|
Medewerkers en onderzoekers
Sponsor
Onderzoekers
- Hoofdonderzoeker: Bojan Pajic, MD, Eye Clinic Orasis
Studie record data
Bestudeer belangrijke data
Studie start
Primaire voltooiing (Werkelijk)
Studie voltooiing (Werkelijk)
Studieregistratiedata
Eerst ingediend
Eerst ingediend dat voldeed aan de QC-criteria
Eerst geplaatst (Schatting)
Updates van studierecords
Laatste update geplaatst (Werkelijk)
Laatste update ingediend die voldeed aan QC-criteria
Laatst geverifieerd
Meer informatie
Termen gerelateerd aan deze studie
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- ZIOS-13/01
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .