Comparison of Nasal and Forehead Oximetry in Critically Ill Patients at Risk for Decreased Peripheral Perfusion
8 mei 2017 bijgewerkt door: Lynn Schallom, Washington University School of Medicine
Continuous pulse oximetry monitoring is the standard of care in critically ill patients in emergency departments, operating rooms and intensive care units.
In patients with poor peripheral perfusion (low blood flow) due to peripheral vascular disease, low body temperature, or shock and the use of medications to raise the blood pressure, clinicians have difficulty obtaining an accurate measurement.
This study compares the accuracy of forehead oximetry sensors to nasal alar sensors to lab oximetry measures and on the rate of device related pressure ulcers with both.
Studie Overzicht
Toestand
Voltooid
Conditie
Interventie / Behandeling
Gedetailleerde beschrijving
Continuous pulse oximetry monitoring is the standard of care in critically ill patients in emergency departments, operating rooms and intensive care units.
In patients with poor peripheral perfusion (low blood flow) due to peripheral vascular disease, low body temperature, or shock and the use of medications to raise the blood pressure, clinicians have difficulty obtaining an accurate measurement.
Several studies have demonstrated the utility of forehead sensor measurements under these clinical conditions.
Forehead sensor measurement is considered to be a more central measurement than with digit or ear sensors.
However the sensor requires a head band for accurate measurement.
On a regular basis, the forehead sensor has led to pressure ulcer development at Barnes-Jewish Hospital despite following vendor recommendations for alternating placement every 8 hours from one side of the forehead to the other.
An alternate sensor placed on the nose has recently demonstrated rapid detection of induced drops in oxygen saturation.
It has also demonstrated correlation with arterial oxygen saturation measured in a clinical laboratory.
The oxygen saturation measurement from the nose is also considered a central measurement.
However, studies of the nose sensor were conducted in healthy subjects or during anesthesia care over several hours.
Therefore, research is needed to examine the accuracy of the nose sensor in the ICU patient population.
In addition, due to concerns for sensor related pressure ulcers in patients with decreased perfusion, the sensor needs to be evaluated for extended periods of time.
Studietype
Ingrijpend
Inschrijving (Werkelijk)
43
Fase
- Niet toepasbaar
Contacten en locaties
In dit gedeelte vindt u de contactgegevens van degenen die het onderzoek uitvoeren en informatie over waar dit onderzoek wordt uitgevoerd.
Studie Locaties
-
-
Missouri
-
Saint Louis, Missouri, Verenigde Staten, 63110
- Barnes-Jewish Hospital at Washington University
-
-
Deelname Criteria
Onderzoekers zoeken naar mensen die aan een bepaalde beschrijving voldoen, de zogenaamde geschiktheidscriteria. Enkele voorbeelden van deze criteria zijn iemands algemene gezondheidstoestand of eerdere behandelingen.
Geschiktheidscriteria
Leeftijden die in aanmerking komen voor studie
18 jaar tot 90 jaar (Volwassen, Oudere volwassene)
Accepteert gezonde vrijwilligers
Nee
Geslachten die in aanmerking komen voor studie
Allemaal
Beschrijving
Inclusion Criteria:
1. Presence of an existing arterial catheter for blood sampling and one of the following:
- Difficulty obtaining a consistent signal from a digit or ear sensor or
- On pressors of at least 0 .10 mcg/kg/min of norepinephrine or
- Core temperature < or equal to 35 degrees C
Exclusion Criteria:
- Anatomic impediments (burns, wounds, dressings, etc.) to placement of the sensor on the forehead or nasal alar
- History of known dyshemoglobinemias evidenced by carboxyhemoglobin levels > 10% or methemoglobin level > 2%
- Severe anemia with hemoglobin < 5 g/dL
- Pregnant women
Studie plan
Dit gedeelte bevat details van het studieplan, inclusief hoe de studie is opgezet en wat de studie meet.
Hoe is de studie opgezet?
Ontwerpdetails
- Primair doel: Diagnostisch
- Toewijzing: NVT
- Interventioneel model: Opdracht voor een enkele groep
- Masker: Geen (open label)
Wapens en interventies
Deelnemersgroep / Arm |
Interventie / Behandeling |
|---|---|
|
Ander: Nasal alar oxygen sensor
Application of a nasal alar oxygen sensor
|
Application of a nasal alar oxygen sensor
|
Wat meet het onderzoek?
Primaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Accuracy as Indicated by Co-oximetry Measure of Arterial Oxygen Saturation
Tijdsspanne: 24 hours
|
accuracy of sensor measure was defined as sensor measurements within 3% of co-oximetry measures
|
24 hours
|
Secundaire uitkomstmaten
Uitkomstmaat |
Maatregel Beschrijving |
Tijdsspanne |
|---|---|---|
|
Device Related Pressure Ulcer
Tijdsspanne: 5 days
|
assessment for development of pressure ulcer at forehead sensor, "OxiMaxTM, Nellcor,Covidien" and nasal alar sensor, "Alar One-SenseTM, Xhale Assurance"
|
5 days
|
Medewerkers en onderzoekers
Hier vindt u mensen en organisaties die betrokken zijn bij dit onderzoek.
Onderzoekers
- Hoofdonderzoeker: Marilyn E Schallom, RN, PhD, Barnes-Jewish Hospital at Washington University
Publicaties en nuttige links
De persoon die verantwoordelijk is voor het invoeren van informatie over het onderzoek stelt deze publicaties vrijwillig ter beschikking. Dit kan gaan over alles wat met het onderzoek te maken heeft.
Studie record data
Deze datums volgen de voortgang van het onderzoeksdossier en de samenvatting van de ingediende resultaten bij ClinicalTrials.gov. Studieverslagen en gerapporteerde resultaten worden beoordeeld door de National Library of Medicine (NLM) om er zeker van te zijn dat ze voldoen aan specifieke kwaliteitscontrolenormen voordat ze op de openbare website worden geplaatst.
Bestudeer belangrijke data
Studie start
1 oktober 2014
Primaire voltooiing (Werkelijk)
1 juli 2016
Studie voltooiing (Werkelijk)
1 september 2016
Studieregistratiedata
Eerst ingediend
10 februari 2015
Eerst ingediend dat voldeed aan de QC-criteria
5 maart 2015
Eerst geplaatst (Schatting)
6 maart 2015
Updates van studierecords
Laatste update geplaatst (Werkelijk)
7 juni 2017
Laatste update ingediend die voldeed aan QC-criteria
8 mei 2017
Laatst geverifieerd
1 mei 2017
Meer informatie
Termen gerelateerd aan deze studie
Aanvullende relevante MeSH-voorwaarden
Andere studie-ID-nummers
- 201408123
Plan Individuele Deelnemersgegevens (IPD)
Bent u van plan om gegevens van individuele deelnemers (IPD) te delen?
NEE
Deze informatie is zonder wijzigingen rechtstreeks van de website clinicaltrials.gov gehaald. Als u verzoeken heeft om uw onderzoeksgegevens te wijzigen, te verwijderen of bij te werken, neem dan contact op met register@clinicaltrials.gov. Zodra er een wijziging wordt doorgevoerd op clinicaltrials.gov, wordt deze ook automatisch bijgewerkt op onze website .