Ultrasonographic inferior vena cava diameter response to trauma resuscitation after 1 hour predicts 24-hour fluid requirement

Jay Joseph Doucet, Paula Ferrada, Sarah Murthi, Ram Nirula, Sara Edwards, Emily Cantrell, Jinfeng Han, Daniel Haase, Andrew Singleton, Yekaterina Birkas, Giovanna Casola, Raul Coimbra, AAST Multi-Institutional Trials Committee, Jay Joseph Doucet, Paula Ferrada, Sarah Murthi, Ram Nirula, Sara Edwards, Emily Cantrell, Jinfeng Han, Daniel Haase, Andrew Singleton, Yekaterina Birkas, Giovanna Casola, Raul Coimbra, AAST Multi-Institutional Trials Committee

Abstract

Background: Identification of occult hypovolemia in trauma patients is difficult. We hypothesized that in acute trauma patients, the response of ultrasound-measured minimum inferior vena cava diameter (IVCDMIN), IVC Collapsibility Index (IVCCI), minimum internal jugular diameter (IJVDMIN) or IJV Collapsibility Index (IJVCI) after up to 1 hour of fluid resuscitation would predict 24-hour resuscitation intravenous fluid requirements (24FR).

Methods: An NTI-funded, American Association for the Surgery of Trauma Multi-Institutional Trials Committee prospective, cohort trial was conducted at four Level I Trauma Centers. Major trauma patients were screened for an IVCD of 12 mm or less or IVCCI of 50% or less on initial focused assessment sonographic evaluations for trauma. A second IVCD was obtained 40 minutes to 60 minutes later, after standard-of-care fluid resuscitation. Patients whose second measured IVCD was less than 10 mm were deemed nonrepleted (NONREPLETED), those 10 mm or greater were repleted (REPLETED). Prehospital and initial resuscitation fluids and 24FR were recorded. Demographics, Injury Severity Score, arterial blood gasses, length of stay, interventions, and complications were recorded. Means were compared by ANOVA and categorical variables were compared via χ. Receiver operating characteristic curves analysis was used to compare the measures as 24FR predictors.

Results: There were 4,798 patients screened, 196 were identified with admission IVCD of 12 mm or IVCCI of 50% or less, 144 were enrolled. There were 86 REPLETED and 58 NONREPLETED. Demographics, initial hemodynamics, or laboratory measures were not significantly different. NONREPLETED had smaller IVCD (6.0 ± 3.7 mm vs. 14.2 ± 4.3 mm, p < 0.001) and higher IVCCI (41.7% ± 30.0% vs. 13.2% ± 12.7%, p < 0.001) but no significant difference in IJVD or IJVCCI. REPLETED had greater 24FR than NONREPLETED (2503 ± 1751 mL vs. 1,243 ± 1,130 mL, p = 0.003). Receiver operating characteristic analysis indicates IVCDMIN predicted 24FR (area under the curve [AUC], 0.74; 95% confidence interval [CI], 0.64-0.84; p < 0.001) as did IVCCI (AUC, 0.75; 95% CI, 0.65-0.85; p < 0.001) but not IJVDMIN (AUC, 0.48; 95% CI, 0.24-0.60; p = 0.747) or IJVCI (AUC, 0.54; 95% CI, 0.42-0.67; p = 0.591).

Conclusion: Ultrasound assessed IVCDMIN and IVCCI response initial resuscitation predicts 24-hour fluid resuscitation requirements.

Level of evidence: Diagnostic tests or criteria, level II.

Trial registration: ClinicalTrials.gov NCT01989273.

References

    1. Marik PE, Baram M, Vahid B. Does central venous pressure predict fluid responsiveness? A systematic review of the literature and the tale of seven mares. Chest. 2008;134(1):172–178.
    1. Anazodo AN, Murthi SB, Frank MK, et al. Assessing trauma care provider judgement in the prediction of need for life-saving interventions. Injury. 2015;46(5):791–797.
    1. ATLS Subcommittee; American College of Surgeons' Committee on Trauma; International ATLS Working Group. Advanced Trauma Life Support (ATLS®). the ninth edition. J Trauma Acute Care Surg. 2013;74(5):1363–1366.
    1. Saul T, Lewiss RE, Langsfeld A, Radeos MS, Del Rios M. Inter-rater reliability of sonographic measurements of the inferior vena cava. J Emerg Med. 2012;42(5):600–605.
    1. Guarracino F, Ferro B, Forfori F, Bertini P, Magliacano L, Pinsky MR. Jugular vein distensibility predicts fluid responsiveness in septic patients. Crit Care. 2014;18(6):647.
    1. Thudium M, Klaschik S, Ellerkmann RK, Putensen C, Hilbert T. Is internal jugular vein extensibility associated with indices of fluid responsiveness in ventilated patients? Acta Anaesthesiol Scand. 2016;60(6):723–733.
    1. Akilli NB, Cander B, Dundar ZD, Koylu R. A new parameter for the diagnosis of hemorrhagic shock: jugular index. J Crit Care. 2012;27(5):530.e13–530.e18.
    1. Unluer EE, Kara PH. Ultrasonography of jugular vein as a marker of hypovolemia in healthy volunteers. Am J Emerg Med. 2013;31(1):173–177.
    1. McGraw K, Wong SP. Forming Inferences About Some Intraclass Correlation Coefficients. 1996. 30-46 p.
    1. Coste J, Jourdain P, Pouchot J. A gray zone assigned to inconclusive results of quantitative diagnostic tests: application to the use of brain natriuretic peptide for diagnosis of heart failure in acute dyspneic patients. Clin Chem. 2006;52(12):2229–2235.
    1. Duwat A, Zogheib E, Guinot P, Levy F, Trojette F, Diouf M, Slama M, Dupont H. The gray zone of the qualitative assessment of respiratory changes in inferior vena cava diameter in ICU patients. Crit Care. 2014;18(1):R14.
    1. Rozycki GS, Ochsner MG, Schmidt JA, Frankel HL, Davis TP, Wang D, Champion HR. A prospective study of surgeon-performed ultrasound as the primary adjuvant modality for injured patient assessment. J Trauma. 1995;39(3):492–498; discussion 498-500.
    1. Brown MA, Sirlin CB, Hoyt DB, Casola G. Screening ultrasound in blunt abdominal trauma. J Intensive Care Med. 2003;18(5):253–260.
    1. Dehqanzada ZA, Meisinger Q, Doucet J, Smith A, Casola G, Coimbra R. Complete ultrasonography of trauma in screening blunt abdominal trauma patients is equivalent to computed tomographic scanning while reducing radiation exposure and cost. J Trauma Acute Care Surg. 2015;79(2):199–205.
    1. Dipti A, Soucy Z, Surana A, Chandra S. Role of inferior vena cava diameter in assessment of volume status: a meta-analysis. Am J Emerg Med. 2012;30(8):1414–9.e1.
    1. Monnet X, Marik PE, Teboul JL. Prediction of fluid responsiveness: an update. Ann Intensive Care. 2016;6(1):111.
    1. Rudski LG, Lai WW, Afilalo J, Hua L, Handschumacher MD, Chandrasekaran K, Solomon SD, Louie EK, Schiller NB. Guidelines for the echocardiographic assessment of the right heart in adults: a report from the American Society of Echocardiography endorsed by the European Association of Echocardiography, a registered branch of the European Society of Cardiology, and the Canadian Society of Echocardiography. J Am Soc Echocardiogr. 2010;23(7):685–713; quiz 86-8.
    1. Zengin S, Al B, Genc S, Yildirim C, Ercan S, Dogan M, Altunbas G. Role of inferior vena cava and right ventricular diameter in assessment of volume status: a comparative study: ultrasound and hypovolemia. Am J Emerg Med. 2013;31(5):763–767.
    1. Kent A, Bahner DP, Boulger CT, et al. Sonographic evaluation of intravascular volume status in the surgical intensive care unit: a prospective comparison of subclavian vein and inferior vena cava collapsibility index. J Surg Res. 2013;184(1):561–566.
    1. Stawicki SP, Adkins EJ, Eiferman DS, et al. Prospective evaluation of intravascular volume status in critically ill patients: does inferior vena cava collapsibility correlate with central venous pressure? J Trauma Acute Care Surg. 2014;76(4):956–963; discussion 63-4.
    1. Murthi SB, Fatima S, Menne AR, Glaser JJ, Galvagno SM, Biederman S, Fang R, Chen H, Scalea TM. Ultrasound assessment of volume responsiveness in critically ill surgical patients: two measurements are better than one. J Trauma Acute Care Surg. 2017;82(3):505–511.
    1. Barbier C, Loubieres Y, Schmit C, Hayon J, Ricome JL, Jardin F, Vieillard-Baron A. Respiratory changes in inferior vena cava diameter are helpful in predicting fluid responsiveness in ventilated septic patients. Intensive Care Med. 2004;30(9):1740–1746.
    1. Schefold JC, Storm C, Bercker S, Pschowski R, Oppert M, Kruger A, Hasper D. Inferior vena cava diameter correlates with invasive hemodynamic measures in mechanically ventilated intensive care unit patients with sepsis. J Emerg Med. 2010;38(5):632–637.
    1. Feissel M, Michard F, Faller JP, Teboul JL. The respiratory variation in inferior vena cava diameter as a guide to fluid therapy. Intensive Care Med. 2004;30(9):1834–1837.
    1. Airapetian N, Maizel J, Alyamani O, et al. Does inferior vena cava respiratory variability predict fluid responsiveness in spontaneously breathing patients? Crit Care. 2015;19:400.
    1. Corl KA, George NR, Romanoff J, Levinson AT, Chheng DB, Merchant RC, Levy MM, Napoli AM. Inferior vena cava collapsibility detects fluid responsiveness among spontaneously breathing critically-ill patients. J Crit Care. 2017;41:130–137.
    1. Juhl-Olsen P, Vistisen ST, Christiansen LK, Rasmussen LA, Frederiksen CA, Sloth E. Ultrasound of the inferior vena cava does not predict hemodynamic response to early hemorrhage. J Emerg Med. 2013;45(4):592–597.
    1. Lyon M, Blaivas M, Brannam L. Sonographic measurement of the inferior vena cava as a marker of blood loss. Am J Emerg Med. 2005;23(1):45–50.
    1. Pasquero P, Albani S, Sitia E, Taulaigo AV, Borio L, Berchialla P, Castagno F, Porta M. Inferior vena cava diameters and collapsibility index reveal early volume depletion in a blood donor model. Crit Ultrasound J. 2015;7(1):17.
    1. Resnick J, Cydulka R, Platz E, Jones R. Ultrasound does not detect early blood loss in healthy volunteers donating blood. J Emerg Med. 2011;41(3):270–275.
    1. Shaik Farid AW, Mohd Hashairi F, Nik Hisamuddin NA, Chew KS, Rashidi A. Ultrasonography measurement of inferior vena cava diameter of blood donors as predictors for early blood loss in tertiary hospital northeastern, Malaysia. Med J Malaysia. 2013;68(6):465–468.
    1. Corl K, Napoli AM, Gardiner F. Bedside sonographic measurement of the inferior vena cava caval index is a poor predictor of fluid responsiveness in emergency department patients. Emerg Med Australas. 2012;24(5):534–539.
    1. Çelik ÖF, Akoğlu H, Çelik A, Asadov R, Onur ÖE, Denizbaşi A. Initial inferior vena cava and aorta diameter parameters measured by ultrasonography or computed tomography does not correlate with vital signs, hemorrhage or shock markers in trauma patients. Ulus Travma Acil Cerrahi Derg. 2018;24(4):351–358.
    1. Zhang Z, Xu X, Ye S, Xu L. Ultrasonographic measurement of the respiratory variation in the inferior vena cava diameter is predictive of fluid responsiveness in critically ill patients: systematic review and meta-analysis. Ultrasound Med Biol. 2014;40(5):845–853.
    1. Muller L, Bobbia X, Toumi M, et al. Respiratory variations of inferior vena cava diameter to predict fluid responsiveness in spontaneously breathing patients with acute circulatory failure: need for a cautious use. Crit Care. 2012;16(5):R188.
    1. Seif D, Mailhot T, Perera P, Mandavia D. Caval sonography in shock: a noninvasive method for evaluating intravascular volume in critically ill patients. J Ultrasound Med. 2012;31(12):1885–1890.
    1. Zhao J, Wang G. Inferior vena cava collapsibility index is a valuable and non-invasive index for elevated general heart end-diastolic volume index estimation in septic shock patients. Med Sci Monit. 2016;22:3843–3848.
    1. Vieillard-Baron A, Evrard B, Repesse X, et al. Limited value of end-expiratory inferior vena cava diameter to predict fluid responsiveness impact of intra-abdominal pressure. Intensive Care Med. 2018;44(2):197–203.
    1. Kent A, Patil P, Davila V, et al. Sonographic evaluation of intravascular volume status: can internal jugular or femoral vein collapsibility be used in the absence of IVC visualization? Ann Thorac Med. 2015;10(1):44–49.
    1. Zhang J, Critchley LA. Inferior vena cava ultrasonography before general anesthesia can predict hypotension after induction. Anesthesiology. 2016;124(3):580–589.
    1. Hu H, Zhu X, Wang C, et al. Stretchable ultrasonic transducer arrays for three-dimensional imaging on complex surfaces. Sci Adv. 2018;4(3):eaar3979.
    1. Huang Z, Hao Y, Li Y, et al. Three-dimensional integrated stretchable electronics. Nat Electron. 2018;1(8):473–480.

Source: PubMed

3
Se inscrever