How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough?

Sarah E Stahl, Hyun-Sung An, Danae M Dinkel, John M Noble, Jung-Min Lee, Sarah E Stahl, Hyun-Sung An, Danae M Dinkel, John M Noble, Jung-Min Lee

Abstract

Background: Heart rate (HR) monitors are valuable devices for fitness-orientated individuals. There has been a vast influx of optical sensing blood flow monitors claiming to provide accurate HR during physical activities. These monitors are worn on the arm and wrist to detect HR with photoplethysmography (PPG) techniques. Little is known about the validity of these wearable activity trackers.

Aim: Validate the Scosche Rhythm (SR), Mio Alpha (MA), Fitbit Charge HR (FH), Basis Peak (BP), Microsoft Band (MB), and TomTom Runner Cardio (TT) wireless HR monitors.

Methods: 50 volunteers (males: n=32, age 19-43 years; females: n=18, age 19-38 years) participated. All monitors were worn simultaneously in a randomised configuration. The Polar RS400 HR chest strap was the criterion measure. A treadmill protocol of one 30 min bout of continuous walking and running at 3.2, 4.8, 6.4, 8.0, and 9.6 km/h (5 min at each protocol speed) with HR manually recorded every minute was completed.

Results: For group comparisons, the mean absolute percentage error values were: 3.3%, 3.6%, 4.0%, 4.6%, 4.8% and 6.2% for TT, BP, RH, MA, MB and FH, respectively. Pearson product-moment correlation coefficient (r) was observed: r=0.959 (TT), r=0.956 (MB), r=0.954 (BP), r=0.933 (FH), r=0.930 (RH) and r=0.929 (MA). Results from 95% equivalency testing showed monitors were found to be equivalent to those of the criterion HR (±10% equivalence zone: 98.15-119.96).

Conclusions: The results demonstrate that the wearable activity trackers provide an accurate measurement of HR during walking and running activities.

Keywords: Assessing physiological demands of physical activity; Assessing validity and reliability of test of physiological parameters; Health promotion; Physical activity; Validity.

Figures

Figure 1
Figure 1
Bland-Altman plots for all monitors. HR, heart rate.

References

    1. Laukkanen RM, Virtanen PK. Heart rate monitors: state of the art. J Sport Sci 1998;16(Suppl 1):3–7. 10.1080/026404198366920
    1. Terbizan DJ, Dolezal BA, Albano C. Validity of seven commercially available heart rate monitors. Meas in Phys Educ Exerc Sci 2002;6:243–7. 10.1207/S15327841MPEE0604_3
    1. Lee JM, Kim Y, Welk GJ. Validity of consumer-based physical activity monitors. Med Sci Sports Exerc 2014;46:1840–8. 10.1249/MSS.0000000000000287
    1. Brage S, Brage N, Ekelund U et al. . Effect of combined movement and heart rate monitor placement on physical activity estimates during treadmill locomotion and free-living. Eur J Appl Physiol 2006;96:517–24. 10.1007/s00421-005-0112-6
    1. Maddison R, Ni Mhurchu C. Global positioning system: a new opportunity in physical activity measurement. Int J Behav Nutr Phys Act 2009;6:73 10.1186/1479-5868-6-73
    1. Montgomery PG, Green DJ, Etxebarria N et al. . Validation of heart rate monitor-based predictions of oxygen uptake and energy expenditure. J Strength Cond Res 2009;23:1489–95. 10.1519/JSC.0b013e3181a39277
    1. Spierer DK, Rosen Z, Litman LL et al. . Validation of photoplethysmography as a method to detect heart rate during rest and exercise. J Med Eng Technol 2015;39:264–71. 10.3109/03091902.2015.1047536
    1. Engström E, Ottosson E, Wohlfart B et al. . Comparison of heart rate measured by Polar RS400 and ECG, validity and repeatability. Adv Physiother 2012;14:115–22. 10.3109/14038196.2012.694118
    1. Parak J, Korhonen I. Evaluation of wearable consumer heart rate monitors based on photoplethysmography. Engineering in Medicine and Biology Society (EMBC), 36th Annual International Conference of the IEEE; IEEE, 2014.
    1. Bland JM, Altman D. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986;327:307–10. 10.1016/S0140-6736(86)90837-8
    1. de Rezende Barbosa MP, Silva NT, de Azevedo FM et al. . Comparison of Polar® RS800G3™ heart rate monitor with Polar® S810i™ and electrocardiogram to obtain the series of RR intervals and analysis of heart rate variability at rest. Clin Physiol Funct Imaging 2016;36:112–17. 10.1111/cpf.12203
    1. Lee CM, Gorelick M. Validity of the Smarthealth watch to measure heart rate during rest and exercise. Meas Phys Educ Exerc Sci 2011;15:18–25. 10.1080/1091367X.2011.539089

Source: PubMed

3
Se inscrever