Long-Term Evaluation of Biotronik Linox and Linox(smart) Implantable Cardioverter Defibrillator Leads

Eric D Good, Ivan Cakulev, Michael V Orlov, David Hirsh, John Simeles, Kelly Mohr, Phil Moll, Heather Bloom, Eric D Good, Ivan Cakulev, Michael V Orlov, David Hirsh, John Simeles, Kelly Mohr, Phil Moll, Heather Bloom

Abstract

Introduction: Expert consensus holds that post-market, systematic surveillance of ICD leads is essential to ensure confirmation of adequate lead performance. GALAXY (NCT00836589) and CELESTIAL (NCT00810264) are ongoing multicenter, prospective, non-randomized registries conducted to confirm the long-term safety and reliability of Biotronik leads.

Methods and results: ICD and CRT-D patients are followed for Linox and Linox(smart) ICD lead performance and safety for 5 years post-implant. All procedural and system-related adverse events (AEs) were assessed at each follow-up, along with lead electrical parameters. An independent CEC of EPs adjudicated AEs to determine AE category and lead relatedness. The analysis used categories of lead observations per ISO 5841-2 (Third edition). A total of 3,933 leads were implanted in 3,840 patients (73.0% male, mean age 67.0 ± 12.2 years) at 146 US centers. The estimated cumulative survival probability was 96.3% at 5 years after implant for Linox leads and 96.6% at 4 years after implant for Linox(smart) leads. A comparison of the Linox and Linox(smart) survival functions did not find evidence of a difference (P = 0.2155). The most common AEs were oversensing (23, 0.58%), conductor fracture (14, 0.36%), failure to capture (13, 0.33%), lead dislodgement (12, 0.31%), insulation breach (10, 0.25%), and abnormal pacing impedance (8, 0.20%).

Conclusions: Linox and Linox(smart) ICD leads are safe, reliable and infrequently associated with lead-related AEs. Additionally, estimated cumulative survival probability is clinically acceptable and well within industry standards. Ongoing data collection will confirm the longer-term safety and performance of the Linox family of ICD leads.

Keywords: Biotronik; Linox lead; Performance; adverse event; cardiac resynchronization therapy; implantable cardioverter defibrillator; lead; reliability; survival.

© 2016 Wiley Periodicals, Inc.

Source: PubMed

Подписаться