Clinical Evaluation of a Wireless Monitoring Device to Reduce Falls in the Elderly and Others at High Risk of Falling
NIH-NIA SBIR Phase I/II Study: Clinical Evaluation of a Device to Reduce Patient Falls
Studieöversikt
Status
Betingelser
Intervention / Behandling
Detaljerad beskrivning
OBJECTIVE. No existing devices have proven effective in preventing falls in at-risk individuals. The objective of this study was to determine whether the FallSaver device reduces falls and fall-related injuries in elderly skilled nursing facility (SNF) residents.
DESIGN. This was a randomized, prospective, open-label, cross-over study conducted over a six-month period.
SETTING: The study took place in a 100-bed State Veterans SNF in Washington State.
PARTICIPANTS. Forty-four subjects with anticipated institutional stays of at least 120 days participated in the study. The subjects' mean age was 82.2 + 7.1 years and all were rated high fall risks using the Morse scale.
INTERVENTION Subjects were randomly assigned to the FallSaver device or no device (Observation) for 60 days. Following the end of the 60-day period subjects were crossed over to the opposite treatment. The device, enclosed in an adhesive patch, was applied to the subject's thigh.
MAIN OUTCOME MEASURE The primary outcome measures were falls and fall-related injuries. Secondary outcomes were falls resulting from documented attempts to stand or ambulate without assistance, skin-tolerance to the patch, and compliance.
RESULTS. Total patient-days for the FallSaver and Observation phases were 1,923 and 2,299, respectively. A total of 37 falls occurred in 18 subjects. Thirteen falls occurred during the FallSaver phase and 24 during Observation (p = < 0.05). A total of 7 fall-related injuries occurred; one during the FallSaver phase and 6 during Observation (p = < 0.01). Nineteen falls resulted from documented attempts to stand or ambulate without assistance, six during the FallSaver phase and 13 during Observation (p = < 0.05). The patch was well tolerated and no serious adverse effects were observed.
CONCLUSION. The FallSaver device reduced the frequency of falls by 50% and fall-related injuries by 82% in elderly at-risk SNF residents. The device and associated patch enclosure was well tolerated and devoid of serious problems. Significant cost savings and fewer reductions in quality of life are possible if fall-related injuries can be reduced.
Studietyp
Inskrivning
Fas
- Fas 2
- Fas 1
Kontakter och platser
Studieorter
-
-
Washington
-
Spokane, Washington, Förenta staterna, 99202
- Spokane Veterans Homes
-
-
Deltagandekriterier
Urvalskriterier
Åldrar som är berättigade till studier
Tar emot friska volontärer
Kön som är behöriga för studier
Beskrivning
Inclusion Criteria:
- Residents of a skilled nursing facility. An anticipated institutional stay of at least 120 days. Individuals with a high risk of falling according to the Morse scale (a score of 55+).
Exclusion Criteria:
- Individuals who were allowed to self-ambulate. Individuals with a history of adverse reactions to medical adhesives. Individuals with a history or presence of significant skin breakdown on the legs.
Studieplan
Hur är studien utformad?
Designdetaljer
- Primärt syfte: Förebyggande
- Tilldelning: Randomiserad
- Interventionsmodell: Crossover tilldelning
- Maskning: Ingen (Open Label)
Vad mäter studien?
Primära resultatmått
Resultatmått |
---|
Falls and fall-related injuries.
|
Sekundära resultatmått
Resultatmått |
---|
Efterlevnad.
|
Falls resulting from documented attempts to stand or ambulate without assistance.
|
Skin tolerance to the adhesive.
|
Samarbetspartners och utredare
Sponsor
Utredare
- Studierektor: Kathryn E Kelly, DrPH, NOCwatch International, Inc.
- Huvudutredare: Dennis Clifton, PharmD, Washington State University
Publikationer och användbara länkar
Studieavstämningsdatum
Studera stora datum
Studiestart
Avslutad studie
Studieregistreringsdatum
Först inskickad
Först inskickad som uppfyllde QC-kriterierna
Första postat (Uppskatta)
Uppdateringar av studier
Senaste uppdatering publicerad (Uppskatta)
Senaste inskickade uppdateringen som uppfyllde QC-kriterierna
Senast verifierad
Mer information
Termer relaterade till denna studie
Nyckelord
Andra studie-ID-nummer
- 4R44AG022270-02 (U.S.S. NIH-anslag/kontrakt)
Denna information hämtades direkt från webbplatsen clinicaltrials.gov utan några ändringar. Om du har några önskemål om att ändra, ta bort eller uppdatera dina studieuppgifter, vänligen kontakta register@clinicaltrials.gov. Så snart en ändring har implementerats på clinicaltrials.gov, kommer denna att uppdateras automatiskt även på vår webbplats .