Focused Ultrasound Neuromodulation for Treatment of Temporal Lobe Epilepsy (LIFUP)

March 14, 2023 updated by: Darin Dougherty, MD, Massachusetts General Hospital
The aim of the proposed study is to assess the feasibility of using the BrainSonix BX Pulsar 1002 low-frequency and Low-Intensity Focused Ultrasound Pulsations (LIFUP) in human subjects suffering from intractable temporal lobe epilepsy. The patients selected will already be scheduled to undergo surgery for resection of the temporal lobe, and the investigational therapy will be applied to the temporal lobe at least one day prior to its scheduled removal. The study is intended to provide preliminary evidence of safety, and establish the feasibility of LIFUP treatment as evidenced by a modulation of the Blood-oxygenation level dependent (BOLD) signal in functional MRI (fMRI), and normal findings from histological examination of the resected brain tissue.

Study Overview

Status

Withdrawn

Intervention / Treatment

Detailed Description

Non-invasive and controllable suppression of regional brain activity may open new avenues for the treatment and management of epilepsy. There is good evidence that direct stimulation of epileptogenic sources can powerfully suppress seizure rates, but so far minimally invasive attempts have shown limited success. Currently, there are no approved non-invasive procedures for seizure suppression.

Focused ultrasound (FUS) is a promising new potential treatment, because it can target any region in the brain, regardless of depth, with high spatial specificity. FUS allows for the non-invasive delivery of acoustic energy to a well-localized and circumscribed brain region of a few millimeters in diameter, depositing mechanical or thermal energy. The possibility of transcranial application of FUS into deep brain areas has already been demonstrated by the strategic placement and operation of an array of multiple ultrasound transducers.

The ultrasound based BrainSonix device is a highly novel non-invasive alternative to direct stimulation. Capitalizing on non-invasive high-intensity focused ultrasound (HIFU), BrainSonix, using an animal model, found early evidence of selective suppression of regional neural activity in the brain. Importantly these effects were non-thermal and reversible. Further study is required to validate and characterize the degree of suppressive effects on epilepsy to ensure safe application in human subjects. Low Intensity Focused Ultrasound Pulsation (LIFUP) differs from HIFU because the intensities of ultrasound used are an order of magnitude lower than HIFU. Also, whereas ablative HIFU is administered continuously, LIFUP is administered in short pulses, which reduces energy deposition. The investigators' and others data have demonstrated that administration of ultrasound, as LIFUP, causes a significant, but reversible, neuromodulatory effect, while preventing the tissue from being thermally damaged.

Study Procedures

Visit 1 Assessment The initial assessment will be performed shortly after signing the consent form. The patient will meet with the clinician and baseline demographic information and general physical information will be collected. The patient will also undergo a psychiatric/neurological evaluation, including a neurological exam, and several neuropsychiatric questionnaires using the Brief Symptom Inventory (BSI) self-report, the Beck Depression Inventory (BDI) self-report, a neuropsychiatric battery of tests described in section 9.2.4 below, the Mini-International Neuropsychiatric Interview (MINI) and the Mini-Mental Status Assessment (MMSE). They will also be required to complete an Magnetic Resonance Imaging (MRI) screening questionnaire to detect the presence of any devices that might present a safety risk.

Visit 2 Treatment and Assessment At visit 2, the location of the epileptic focus will be determined based on previously performed studies as part of the individual's routine epilepsy evaluation. The location of the epileptic focus will be determined based on previously performed studies as part of the individual's routine epilepsy evaluation. Brain MRI, interictal electroencephalogram (EEG), ictal EEG, interictal positron emission tomography (PET), ictal PET, interictal single-photon emission computerized tomography (SPECT), ictal SPECT, interictal magnetoencephalography (MEG), ictal MEG, and magnetic resonance (MR) spectroscopy data may be available and utilized to determine each participant's epileptogenic zone.

LIFUP will be administered to the temporal region on the side to undergo surgery, within the focus producing seizures (as determined by the comprehensive epilepsy evaluation). LIFUP session will take place during the 7 days prior to the operation. See Section 8.2.1 for more details regarding the LIFUP Sonication procedure.

The subject's heart rate and blood pressure will be monitored during the procedure. Routine electrode electroencephalography (EEG) will be recorded immediately before and after the LIFUP sonication session. Recording will last approximately 60 minutes.

The same neurological exam and neuropsychiatric questionnaires using the Brief Symptom Inventory (BSI) self-report, the Beck Depression Inventory (BDI) self-report, and the neuropsychiatric battery of tests described above will be performed at Visit 2, as well. Testing will last approximately 2 hours.

LIFUP Sonication Procedure

Low-Intensity Focused Ultrasound Pulsation (LIFUP) sonications will be conducted using the LIFUP experimental device BX Pulsar 1002 produced by the Brainsonix Corporation. LIFUP will be administered to the temporal region on the side to undergo surgery, within the focus producing seizures (as determined by the comprehensive epilepsy evaluation). In addition the functional magnet resonance imaging (fMRI) recording will be done during the LIFUP to determine the target area of activity and to navigate the focus of the transducer to that area. The ultrasound will be focused on the highest activity area within the temporal lobe. Functional MRI of the brain will be obtained throughout the LIFUP session. Initially LIFUP will be administered below the minimum effective dose determined in previous animal experiments. Up to 8 sonications will be administered during the LIFUP session. The total dose of LIFUP is not to exceed 2 min. The total time within the MRI machine will be up to 90 minutes.

After the participant is fitted with the transducer they will be placed in the MRI machine and undergo a series of scans. After placing the transducer over the anterior tip of the temporal lobe the participant will undergo fMRI BOLD scans with LIFUP excitation: Tone Burst Duration = 2ms and Pulse Repetition Frequency = 250Hz) at four levels of intensity (720mW/cm2, 1440mW/cm2, 2880mW/cm2 and 5760mW/cm2). Before increasing the intensity, the transducer will be moved approximately 5mm posterior. The excitation scans will be analyzed immediately for evidence of BOLD activation. After the last excitation scan the LIFUP parameters will be changed to inhibitory parameters: Tone Burst Duration = 0.5ms and Pulse Repetition Frequency = 100Hz).

Surgical Resection and Histology After surgical resection of the anterior temporal lobe, histology will be performed on the resected tissue to look for evidence of tissue injury not characteristic of temporal lobe epilepsy or previously identified on the routine brain MRI.

Three Month Follow-up At three months the epilepsy surgeon will assess the healing of the surgical wounds and will conduct a Structural MR. The results will be reported back to the investigator.

Six Month Follow-up At six months the epilepsy surgeon will carry out a neurogical exam, and several neuropsychiatric questionnaires using the Brief Symptom Inventory (BSI) self-report, the Beck Depression Inventory (BDI) self-report, a neuropsychiatric battery of tests described in section 9.2.4 below, the MINI, and the Mini Mental Status Assessment.

Neuropsychiatric Battery of Tests The following battery of neuropsychiatric tests will be performed on the study subjects at one of the participant's standard clinical visits, as part of their standard clinical care and at visit 2 following the LIFUP sonication procedure. This battery of tests has been designed to not interfere with their standard clinical care. The Visit 2 scores will be compared to the Visit 1 (Baseline) scores to assess neuropsychiatric changes as a result of the LIFUP procedure.

Seizure Account Participants will be instructed to maintain a daily log of all seizures from the date of consenting until the date of the operation.

Heart Rate and Blood Pressure Monitoring The patient will be affixed with a heart rate monitor. A periodic blood pressure check will be performed if the heart rate increases more than 30% during or immediately after sonication. The investigators' lab has the capacity to measure those parameters in MRI. If the patient's blood pressure has also increased more than 10% then they will be withdrawn from the study.

EEG Monitoring Routine electrode electroencephalography (EEG) recorded immediately before and after the LIFUP sonication session. Recording will last approximately 60 minutes.

A 30-minute awake EEG using scalp electrodes placed according to the International 10-20 System of electrode placement will be performed immediately before and immediately after the LIFUP treatment.

Interim Safety Analysis The data will be analyzed after the first subject in order to determine subsequent treatment paradigm in the next 2 subjects. After the first participant, histology data will also be analyzed by the investigator. If the participant shows histological damage in the a category (i.e. apoptosis, irreversible blood brain barrier (BBB) disruption, ischemia) at a given intensity then the investigators will continue with the next cohort of 2 participants, but the investigators will no longer use that intensity, or higher intensities. If more than one participant shows histological damage in the same category at the lowest intensity (720mW/cm2), then the study will be stopped.

Biostatistical Analysis

Statistical analyses will be performed using SAS® v9.3 or higher (SAS Institute, Cary NC, USA).

All statistical analyses of safety and performance measures will be descriptive in nature. Continuous variables will be summarized by a mean, standard deviation, minimum, median and maximum, and categorical variables by a count and percentage. Confidence intervals will be provided where relevant.

If multiple evaluations are taken in a single patient, statistics described below will be appropriately modified to accommodate the within patient correlation.

Demographic, medical and clinical history variables will be tabulated. Continuous variables will be summarized by a mean, standard deviation, minimum, median and maximum, and categorical variables by a count and percentage.

The numbers of patients who were enrolled and completed each visit of the study will be provided, as well as the reasons for all enrollment discontinuations, grouped by major reason (e.g., lost to follow-up, adverse event). A list of discontinued patients, protocol deviations, and patients excluded from the efficacy analysis will be provided as well.

Histological results in the resected tissue of the anterior temporal lobe will be presented.

Comparison of all neurological assessment scores at the pre-treatment visit and at the post-treatment visit will be presented.

Comparison of the number of epileptiform discharges and Epileptiform Discharge frequency (DF) during a 30-minute period before and a 30-minute period after LIFUP will be presented.

Change in BOLD signal during or after LIFUP sonication will be presented.

Subjects who do not have valid data for the efficacy measures will be treated as missing values. No imputation is planned.

Study Type

Interventional

Phase

  • Not Applicable

Contacts and Locations

This section provides the contact details for those conducting the study, and information on where this study is being conducted.

Study Contact

Participation Criteria

Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.

Eligibility Criteria

Ages Eligible for Study

18 years to 60 years (Adult)

Accepts Healthy Volunteers

No

Genders Eligible for Study

All

Description

Inclusion Criteria:

  • Male and female subjects, 18 to 60 years old.
  • Subjects with clinical evidence from their diagnostic evaluations of unilateral hippocampal - dysfunction and epileptogenicity, confirmed via intracarotid amobarbital procedure (IAP) and neuropsychological testing.
  • Subjects with seizures that have been refractory to treatment with at least three currently marketed antiepileptic drugs.
  • Subjects currently taking antiepileptic medications.
  • Subjects with at least 3 seizures/month based on seizure diary.
  • Subjects with epilepsy who would clearly benefit from surgical intervention.
  • Subjects who have been offered an anterior-mesial temporal lobe resection, performed en-bloc, as treatment for medication refractory epilepsy. This includes both dominant or non-dominant mesial temporal lobe focal epilepsy.

Exclusion Criteria:

  • Subjects with a cognitive or psychiatric disorder that limits the ability to give informed consent or are unable to cooperate with the testing.
  • Subjects with dementia, delirium and psychotic symptoms.
  • Subjects with ferromagnetic materials in the head.
  • Subjects with severe cardiac disease, increased intracranial pressure, or a transcutaneous electrical nerve stimulation (TENS) unit.
  • Subjects who exhibit primary generalized seizures or pseudoseizures.
  • Subjects who have seizures secondary to drugs, alcohol, metabolic illness or progressive degenerative disease.
  • Subjects who have experienced status epilepticus during the 3-week period prior to the LIFUP procedure.
  • Subjects (females) who are pregnant.
  • Patients who have had electrodes implanted in the brain that were not explanted at least 6 weeks prior to the LIFUP procedure.
  • Patients who have had electrodes implanted into the focal point of the temporal lobe that is proposed for the ultrasonic stimulation using LIFUP.

Study Plan

This section provides details of the study plan, including how the study is designed and what the study is measuring.

How is the study designed?

Design Details

  • Primary Purpose: Device Feasibility
  • Allocation: N/A
  • Interventional Model: Single Group Assignment
  • Masking: None (Open Label)

Arms and Interventions

Participant Group / Arm
Intervention / Treatment
Experimental: BX Pulsar 1002
Low-Intensity Focused Ultrasound Pulsation (LIFUP) sonications will be conducted using the LIFUP experimental device BX Pulsar 1002 produced by the Brainsonix Corporation. For the purposes of safety LIFUP sonications will be initiated at the FDA limit for diagnostic ultrasound. However, the minimally effective dose in humans applications, according to (Lee et al., 2015), when derated is approximately 1125mW/cm2.

LIFUP will be administered to the temporal region on the side to undergo surgery, within the focus producing seizures. The ultrasound will be focused on the highest activity area within the temporal lobe. Functional MRI of the brain will be obtained throughout the LIFUP session.

After the participant is fitted with the transducer they will be placed in the MRI machine and undergo a series of scans. After placing the transducer over the anterior tip of the temporal lobe the participant will undergo fMRI BOLD scans with LIFUP excitation (from (Lee et al., 2015): Tone Burst Duration = 2ms and Pulse Repetition Frequency = 250Hz) at four levels of intensity (720mW/cm2, 1440mW/cm2, 2880mW/cm2 and 5760mW/cm2). Before increasing the intensity, the transducer will be moved approximately 5mm posterior. After the last excitation scan the LIFUP parameters will be changed to inhibitory parameters (from (Yoo et al., 2011): Tone Burst Duration = 0.5ms and Pulse Repetition Frequency = 100Hz).

Other Names:
  • LIFUP

What is the study measuring?

Primary Outcome Measures

Outcome Measure
Measure Description
Time Frame
FMRI - (Change in BOLD signal)
Time Frame: 6 Months
Change in BOLD signal during or after LIFUP sonication
6 Months

Collaborators and Investigators

This is where you will find people and organizations involved with this study.

Publications and helpful links

The person responsible for entering information about the study voluntarily provides these publications. These may be about anything related to the study.

General Publications

Study record dates

These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.

Study Major Dates

Study Start (Anticipated)

January 1, 2022

Primary Completion (Actual)

March 1, 2023

Study Completion (Actual)

March 1, 2023

Study Registration Dates

First Submitted

August 30, 2018

First Submitted That Met QC Criteria

August 31, 2018

First Posted (Actual)

September 4, 2018

Study Record Updates

Last Update Posted (Actual)

March 16, 2023

Last Update Submitted That Met QC Criteria

March 14, 2023

Last Verified

March 1, 2023

More Information

Terms related to this study

Plan for Individual participant data (IPD)

Plan to Share Individual Participant Data (IPD)?

NO

Drug and device information, study documents

Studies a U.S. FDA-regulated drug product

No

Studies a U.S. FDA-regulated device product

Yes

This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.

Clinical Trials on Low Intensity Ultrasound Pulsarions

Clinical Trials on BX Pulsar 1002

3
Subscribe