- ICH GCP
- US Clinical Trials Registry
- Clinical Trial NCT05445492
Respiration Validation for CPM Device
May 14, 2024 updated by: Analog Device, Inc.
Clinical Validation Study for Noninvasive Cardiopulmonary Management Device
This study is designed to be a clinical validation study to ready the CPM System for FDA 510(k) submission.
This study will be conducted as a prospective non-randomized.
The study is non-significant risk since the CPM Device is noninterventional and noninvasive.
The study is primarily designed to validate the accuracy of the respiration rate and the changes in relative tidal volume.
All participants will be fitted with both the CPM Device as well as reference devices (capnography and ECG).
Participants will be randomized between 4 procedures (A-D) which will determine the order of 10 exercises.
Each exercise has about 2 minutes of device recording and two minutes of rest.
Each exercise contains two positions: sitting up for one minute and lying down for one minute.
For some exercises, the participant will be instructed to breathe at a certain rate (assisted by a metronome).
For other exercises, the participants will breathe at a normal rate but change how deep the are breathing.
For the last, all capnography reference devices will be removed, and an ECG reference device will be placed.
After exercise 18, the participant will have all devices removed and the study visit will terminate.
The participant will receive a safety follow-up call about 1 week after their visit.
Study Overview
Study Type
Interventional
Enrollment (Actual)
41
Phase
- Not Applicable
Contacts and Locations
This section provides the contact details for those conducting the study, and information on where this study is being conducted.
Study Locations
-
-
Missouri
-
Saint Louis, Missouri, United States, 63123
- Clayton Sleep Institute
-
-
Participation Criteria
Researchers look for people who fit a certain description, called eligibility criteria. Some examples of these criteria are a person's general health condition or prior treatments.
Eligibility Criteria
Ages Eligible for Study
18 years and older (Adult, Older Adult)
Accepts Healthy Volunteers
Yes
Study Population
The target population for this cohort is healthy adult volunteers who do not have preexisting heart or lung conditions or illness.
The goal is to recruit a participant population with a range of body types and BMIs and an approximately even split of genders.
Description
Inclusion Criteria:
- Adults over the age of 18 and who are willing and able to give informed consent
- Willing and able to participate in all activities related to this study, including trimming chest hair and wearing a reference device and the CPM wearable device
- Volunteers of any race, any gender
- Range of physiques
- Healthy
Exclusion Criteria:
- Injury or skin disturbance in the area of the test device
- Allergies or sensitivities to silicone/acrylic-based adhesive
- Pregnant
- Currently smokes cigarettes
Has known respiratory conditions that might prevent them from following the study procedure such as:
- Flu
- Pneumonia/bronchitis
- Shortness of breath/respiratory distress
- Respiratory or lung surgery
- Emphysema, COPD, lung disease
- Has self-reported heart or cardiovascular conditions such as chest pain, AFib, CHF, cardiomyopathy, or other conditions that could interfere with cardiopulmonary function
- Has other self-reported health conditions that could interfere with the breathing patterns and exercises detailed in the protocol (including wearing a capnography mask)
Study Plan
This section provides details of the study plan, including how the study is designed and what the study is measuring.
How is the study designed?
Design Details
- Primary Purpose: Diagnostic
- Allocation: N/A
- Interventional Model: Single Group Assignment
- Masking: None (Open Label)
Arms and Interventions
Participant Group / Arm |
Intervention / Treatment |
|---|---|
|
Experimental: Healthy Cohort
The target population for this cohort is healthy adult volunteers who do not have preexisting heart or lung conditions or illness.
The goal is to recruit a participant population with a range of body types and BMIs and an approximately even split of genders and compare the reference device measurements to the test device.
|
The CPM device and reference device are used simultaneous to validate the CPM device.
|
What is the study measuring?
Primary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Accuracy of CPM System Calculated RR Versus Reference Device Via Percent of Error Accuracy of Respiration Rate
Time Frame: 90 minutes
|
Percentage of respiration rate error within ± 2bpm.
Acceptance criteria for clinical utility is >= 85%.
The percentage of respiration rate within ± 2brpm was calculated by data points, regardless of its rate and breathing condition, which the CPM System did not screen, therefore reporting the numeric values.
Each data point will produce binary result (True or False) if the error is within ± 2 brpm.
Assuming this binary result follows a binominal distribution, its confidence interval is also calculated.
|
90 minutes
|
|
Root Mean Squared Error (RMSE) of Respiration Rates From the CPM System vs Reference Device for Metronome Guided Breathing and Non-metronome Guided Breathing.
Time Frame: 90 minutes
|
First, the subject-specific linear fitting is performed after taking the natural log of both CPM's rTV and reference TV.
Then, the residual errors in rTV after the fitting were calculated.
Finally, all subject's residual errors are pooled together to calculate the root mean squared (RMS) residual relative errors.
The confidence interval was calculated assuming the square of this error follows Chi-square distribution.
This error characterized in a log-log domain is converted back to characterize a relative accuracy by taking an exponential of natural constant followed by subtraction of one.
The RMS of residual error in % is characterized in this way because the linear fitting in the log-log domain and its residual errors represents faithful relative residual error regardless of an absolute value of TV.
|
90 minutes
|
Secondary Outcome Measures
Outcome Measure |
Measure Description |
Time Frame |
|---|---|---|
|
Performance of CPM System's ECG by % of CPM ECG Strip Equivalent to Lead II Strip
Time Frame: device measurement (3 minutes)
|
Performance of CPM System's ECG is measurement by comparing CPM System ECG metrics to the reference device Lead II ECG strip.
Clinicians reviewed the criteria below and decided if the CPM strip had equivalence to the Lead II strip.
|
device measurement (3 minutes)
|
|
ECG Accuracy Confirmation Comparing CPM Systems Generated ECG Metrics and Reference Lead II Generated ECG Metrics
Time Frame: device measurement (3 minutes)
|
Determine the accuracy of CPM System's ECG metrics to the reference device.
Results are displayed as mean error with standard deviation of error.
|
device measurement (3 minutes)
|
Collaborators and Investigators
This is where you will find people and organizations involved with this study.
Sponsor
Collaborators
Study record dates
These dates track the progress of study record and summary results submissions to ClinicalTrials.gov. Study records and reported results are reviewed by the National Library of Medicine (NLM) to make sure they meet specific quality control standards before being posted on the public website.
Study Major Dates
Study Start (Actual)
June 22, 2022
Primary Completion (Actual)
August 4, 2022
Study Completion (Actual)
August 11, 2022
Study Registration Dates
First Submitted
June 30, 2022
First Submitted That Met QC Criteria
June 30, 2022
First Posted (Actual)
July 6, 2022
Study Record Updates
Last Update Posted (Actual)
June 12, 2024
Last Update Submitted That Met QC Criteria
May 14, 2024
Last Verified
May 1, 2024
More Information
Terms related to this study
Additional Relevant MeSH Terms
Other Study ID Numbers
- Resp CPM Validation Clayton
Plan for Individual participant data (IPD)
Plan to Share Individual Participant Data (IPD)?
NO
Drug and device information, study documents
Studies a U.S. FDA-regulated drug product
No
Studies a U.S. FDA-regulated device product
Yes
product manufactured in and exported from the U.S.
Yes
This information was retrieved directly from the website clinicaltrials.gov without any changes. If you have any requests to change, remove or update your study details, please contact register@clinicaltrials.gov. As soon as a change is implemented on clinicaltrials.gov, this will be updated automatically on our website as well.
Clinical Trials on Cardiopulmonary Disease
-
Peking UniversityCompleted
-
Nanjing Medical UniversityWest China Hospital; The Affiliated Hospital of Qingdao University; Beijing Anzhen... and other collaboratorsNot yet recruitingCardiopulmonary Bypass | Cardio Vascular DiseaseChina
-
Nanjing Medical UniversityWest China Hospital; The First Affiliated Hospital of Guangzhou Medical University and other collaboratorsEnrolling by invitationCardiovascular Disease | Cardiopulmonary BypassChina
-
Baxter Healthcare CorporationCompletedCardiopulmonary Bypass | Sternotomy | Cardiac Surgery Requiring Cardiopulmonary Bypass and Median SternotomyUnited States
-
Riphah International UniversityRecruiting
-
Lina ZhangZhejiang Chinese Medical UniversityCompleted
-
King's College Hospital NHS TrustKing's College London; Guy's and St Thomas' NHS Foundation Trust; National Institute... and other collaboratorsCompletedAdults With Cardiopulmonary Disease at Risk of Prolonged Weaning From Mechanical VentilationUnited Kingdom
-
Analog Device, Inc.MCRACompletedCardiopulmonary DiseaseUnited States
-
Mayo ClinicCompletedCardiopulmonary DiseaseUnited States
Clinical Trials on CPM Device
-
Analog Device, Inc.Recruiting
-
Analog Device, Inc.Columbia UniversityRecruiting
-
Analog Device, Inc.Withdrawn
-
Chang Gung UniversityRecruiting
-
Nemours Children's ClinicNot yet recruiting
-
Analog Device, Inc.Hope Research InstituteActive, not recruiting
-
Analog Device, Inc.Mayo ClinicCompleted
-
Brigham and Women's HospitalThe New England Baptist HospitalCompletedKnee Arthroplasty
-
Xeras Medical TechnologiesCompletedTotal Knee ReplacementUnited States
-
ROM Technologies, INCCompletedJoint Pain | Knee Osteoarthritis | Osteoarthritis, Knee | Osteo Arthritis Knee | Knee Arthritis | Knee Pain Chronic | Knee DiseaseUnited States