Real-time magnetic resonance imaging-guided stenting of aortic coarctation with commercially available catheter devices in Swine

Amish N Raval, James D Telep, Michael A Guttman, Cengizhan Ozturk, Michael Jones, Richard B Thompson, Victor J Wright, William H Schenke, Ranil DeSilva, Ronnier J Aviles, Venkatesh K Raman, Michael C Slack, Robert J Lederman, Amish N Raval, James D Telep, Michael A Guttman, Cengizhan Ozturk, Michael Jones, Richard B Thompson, Victor J Wright, William H Schenke, Ranil DeSilva, Ronnier J Aviles, Venkatesh K Raman, Michael C Slack, Robert J Lederman

Abstract

Background: Real-time MR imaging (rtMRI) is now technically capable of guiding catheter-based cardiovascular interventions. Compared with x-ray, rtMRI offers superior tissue imaging in any orientation without ionizing radiation. Translation to clinical trials has awaited the availability of clinical-grade catheter devices that are both MRI visible and safe. We report a preclinical safety and feasibility study of rtMRI-guided stenting in a porcine model of aortic coarctation using only commercially available catheter devices.

Method and results: Coarctation stenting was performed wholly under rtMRI guidance in 13 swine. rtMRI permitted procedure planning, device tracking, and accurate stent deployment. "Active" guidewires, incorporating MRI antennas, improved device visualization compared with unmodified "passive" nitinol guidewires and shortened procedure time (26+/-11 versus 106+/-42 minutes; P=0.008). Follow-up catheterization and necropsy showed accurate stent deployment, durable gradient reduction, and appropriate neointimal formation. MRI immediately identified aortic rupture when oversized devices were tested.

Conclusions: This experience demonstrates preclinical safety and feasibility of rtMRI-guided aortic coarctation stenting using commercially available catheter devices. Patients may benefit from rtMRI in the future because of combined device and tissue imaging, freedom from ionizing radiation, and the ability to identify serious complications promptly.

Figures

Figure 1
Figure 1
Surgically created aortic coarctation in swine before (A–C) and after (D, E) stenting. A, D, Black-blood DIR-FSE MRI. B, E, Contrast-enhanced MRA. C, F, Radiocontrast angiography. Stent-related shielding artifact obliterates contrast in stent lumen (E). Scalloped appearance of aortic wall from apposed stent (D).
Figure 2
Figure 2
rtMRI multislice axial (top left), sagittal (bottom left), and 3D-rendered (right) display of stent delivery across aortic coarctation. Passive (A) and active (B) guidewire tracking are compared. Dumbbell-shaped balloon carrying centrally crimped stent (white arrows) can be seen in both approaches; however, wire is much more clearly visible with active guidewire. S indicates superior; I, inferior; A, anterior; and P, posterior.
Figure 3
Figure 3
Stent delivery sequence. A, Advancing balloon/stent over active guidewire. Also see Data Supplement Movie I. B, Positioning into coarctation. Also see Data Supplement Movie II. C, Balloon inflation. D, Shielding artifact darkens aorta after balloon and guidewire are withdrawn.
Figure 4
Figure 4
Velocity map before and after stenting through plane indicated by blue line. Aortic cross sections are displayed as color maps during systole. Sample 1 shows jet acceleration from coarctation lesion; sample 2, flow reversal from turbulent blood flow; and sample 3, restoration of laminar flow with minimal blood acceleration after stenting.
Figure 5
Figure 5
Early detection of aortic perforation. A, DIR-FSE image of coarctation at before stenting. B, Contrast exit at superior edge of stent (arrow). C, DIR-FSE shows apposed oversized stent with blood accumulation along aortic wall (arrow).
Figure 6
Figure 6
Intentional aortic rupture. A, Simultaneous coronal (top left), axial (bottom left), and sagittal slices displayed and 3D rendered (right). Active wire is across coarctation (green), and oversized balloon is inflated. B, During continuous rtMRI, periaortic hematoma is evident immediately (arrows). DIR-FSE images before (C, D) and after (E, G) intentional rupture. Extravasation appears white. Abbreviations as in Figure 2. Also see Data Supplement Movie III.
Figure 7
Figure 7
Representative necropsy specimen 66 days after stent deployment. Sutures used to create coarctation are visible (arrows). Stent struts are fully apposed to wall and fully endothelialized.

References

    1. Hamdan MA, Maheshwari S, Fahey JT, Hellenbrand WE. Endovascular stents for coarctation of the aorta: initial results and intermediate-term follow-up. J Am Coll Cardiol. 2001;38:1518–1523.
    1. Slack MC. The role of stenting in coarctation of the aorta. Prog Pediatr Cardiol. 2001;14:45–57.
    1. Cheatham JP. Stenting of coarctation of the aorta. Catheter Cardiovasc Interv. 2001;54:112–125.
    1. Rickers C, Jerosch-Herold M, Hu X, Murthy N, Wang X, Kong H, Seethamraju RT, Weil J, Wilke NM. Magnetic resonance image–guided transcatheter closure of atrial septal defects. Circulation. 2003;107:132–138.
    1. Buecker A, Spuentrup E, Grabitz R, Freudenthal F, Muehler EG, Schaeffter T, van Vaals JJ, Gunther RW. Magnetic resonance–guided placement of atrial septal closure device in animal model of patent foramen ovale. Circulation. 2002;106:511–515.
    1. Schalla S, Saeed M, Higgins CB, Martin A, Weber O, Moore P. Magnetic resonance–guided cardiac catheterization in a swine model of atrial septal defect. Circulation. 2003;108:1865–1870.
    1. Kuehne T, Saeed M, Higgins CB, Gleason K, Krombach GA, Weber OM, Martin AJ, Turner D, Teitel D, Moore P. Endovascular stents in pulmonary valve and artery in swine: feasibility study of MR imaging-guided deployment and postinterventional assessment. Radiology. 2003;226:475–481.
    1. Serfaty JM, Yang X, Foo TK, Kumar A, Derbyshire A, Atalar E. MRI-guided coronary catheterization and PTCA: a feasibility study on a dog model. Magn Reson Med. 2003;49:258–263.
    1. Dick AJ, Guttman MA, Raman VK, Peters DC, Pessanha BS, Hill JM, Smith S, Scott G, McVeigh ER, Lederman RJ. Magnetic resonance fluoroscopy allows targeted delivery of mesenchymal stem cells to infarct borders in swine. Circulation. 2003;108:2899–2904.
    1. Raman VK, Karmarkar PV, Guttman MA, Dick AJ, Peters DC, Ozturk C, Pessanha BS, Thompson RB, Raval AN, Desilva R, Aviles RJ, Atalar E, McVeigh ER, Lederman RJ. Real-time magnetic resonance-guided endovascular repair of experimental abdominal aortic aneurysm in swine. J Am Coll Cardiol. 2005;45:2069–2077.
    1. Spuentrup E, Ruebben A, Schaeffter T, Manning WJ, Gunther RW, Buecker A. Magnetic resonance–guided coronary artery stent placement in a swine model. Circulation. 2002;105:874–879.
    1. Strother CM, Unal O, Frayne R, Turk A, Omary R, Korosec FR, Mistretta CA. Endovascular treatment of experimental canine aneurysms: feasibility with MR imaging guidance. Radiology. 2000;215:516–519.
    1. Manke C, Nitz WR, Djavidani B, Strotzer M, Lenhart M, Volk M, Feuerbach S, Link J. MR imaging–guided stent placement in iliac arterial stenoses: a feasibility study. Radiology. 2001;219:527–534.
    1. Paetzel C, Zorger N, Bachthaler M, Volk M, Seitz J, Herold T, Feuerbach S, Lenhart M, Nitz WR. Feasibility of MR-guided angioplasty of femoral artery stenoses using real-time imaging and intraarterial contrast-enhanced MR angiography. Rofo. 2004;176:1232–1236.
    1. Bucker A, Neuerburg JM, Adam GB, Glowinski A, Schaeffter T, Rasche V, van Vaals JJ, Gunther RW. Real-time MR guidance for inferior vena cava filter placement in an animal model. J Vasc Interv Radiol. 2001;12:753–756.
    1. Buecker A, Adam GB, Neuerburg JM, Kinzel S, Glowinski A, Schaeffter T, Rasche V, van Vaals JJ, Guenther RW. Simultaneous real-time visualization of the catheter tip and vascular anatomy for MR-guided PTA of iliac arteries in an animal model. J Magn Reson Imaging. 2002;16:201–208.
    1. Omary RA, Frayne R, Unal O, Warner T, Korosec FR, Mistretta CA, Strother CM, Grist TM. MR-guided angioplasty of renal artery stenosis in a pig model: a feasibility study. J Vasc Interv Radiol. 2000;11:373–381.
    1. Bartels LW, Bos C, van Der Weide R, Smits HF, Bakker CJ, Viergever MA. Placement of an inferior vena cava filter in a pig guided by high-resolution MR fluoroscopy at 1.5 T. J Magn Reson Imaging. 2000;12:599–605.
    1. Kee ST, Rhee JS, Butts K, Daniel B, Pauly J, Kerr A, O’Sullivan GJ, Sze DY, Razavi MK, Semba CP, Herfkens RJ, Dake MD. 1999 Gary J. Becker Young Investigator Award: MR-guided transjugular porto-systemic shunt placement in a swine model. J Vasc Interv Radiol. 1999;10:529–535.
    1. Kuehne T, Yilmaz S, Meinus C, Moore P, Saeed M, Weber O, Higgins CB, Blank T, Elsaesser E, Schnackenburg B, Ewert P, Lange PE, Nagel E. Magnetic resonance imaging–guided transcatheter implantation of a prosthetic valve in aortic valve position: feasibility study in swine. J Am Coll Cardiol. 2004;44:2247–2249.
    1. Razavi R, Hill DL, Keevil SF, Miquel ME, Muthurangu V, Hegde S, Rhode K, Barnett M, van Vaals J, Hawkes DJ, Baker E. Cardiac catheterisation guided by MRI in children and adults with congenital heart disease. Lancet. 2003;362:1877–1882.
    1. Lock JE, Niemi T, Burke BA, Einzig S, Castaneda-Zuniga WR. Transcutaneous angioplasty of experimental aortic coarctation. Circulation. 1982;66:1280–1286.
    1. Dick AJ, Raman VK, Raval AN, Guttman MA, Thompson RB, Ozturk C, Peters DC, Stine AM, Wright VJ, Schenke WH, Lederman RJ. Invasive human magnetic resonance imaging during angioplasty: feasibility in a combined XMR suite. Catheter Cardiovasc Interv. 2005;64:265–274.
    1. Guttman MA, McVeigh ER. Techniques for fast stereoscopic MRI. Magn Reson Med. 2001;46:317–323.
    1. Guttman MA, Lederman RJ, Sorger JM, McVeigh ER. Real-time volume rendered MRI for interventional guidance. J Cardiovasc Magn Reson. 2002;4:431–442.
    1. Konings MK, Bartels LW, Smits HF, Bakker CJ. Heating around intravascular guidewires by resonating RF waves. J Magn Reson Imaging. 2000;12:79–85.
    1. Wang Y, Truong TN, Yen C, Bilecen D, Watts R, Trost DW, Prince MR. Quantitative evaluation of susceptibility and shielding effects of nitinol, platinum, cobalt-alloy, and stainless steel stents. Magn Reson Med. 2003;49:972–976.
    1. Heart Disease and Stroke Statistics: 2004 Update Dallas, Tex: American Heart Association; 2003.
    1. Brewer LA, 3rd, Fosburg RG, Mulder GA, Verska JJ. Spinal cord complications following surgery for coarctation of the aorta: a study of 66 cases. J Thorac Cardiovasc Surg. 1972;64:368–381.
    1. Kappetein AP, Zwinderman AH, Bogers AJ, Rohmer J, Huysmans HA. More than thirty-five years of coarctation repair: an unexpected high relapse rate. J Thorac Cardiovasc Surg. 1994;107:87–95.
    1. Allen HD, Beekman RH, 3rd, Garson A, Jr, Hijazi ZM, Mullins C, O’Laughlin MP, Taubert KA. Pediatric therapeutic cardiac catheterization: a statement for healthcare professionals from the Council on Cardiovascular Disease in the Young, American Heart Association. Circulation. 1998;97:609–625.
    1. Suarez de Lezo J, Pan M, Romero M, Medina A, Segura J, Lafuente M, Pavlovic D, Hernandez E, Melian F, Espada J. Immediate and follow-up findings after stent treatment for severe coarctation of aorta. Am J Cardiol. 1999;83:400–406.
    1. Ebeid MR, Prieto LR, Latson LA. Use of balloon-expandable stents for coarctation of the aorta: initial results and intermediate-term follow-up. J Am Coll Cardiol. 1997;30:1847–1852.
    1. Korkola SJ, Tchervenkov CI, Shum-Tim D, Roy N. Aortic rupture after stenting of a native coarctation in an adult. Ann Thorac Surg. 2002;74:936.
    1. Ward KE. Complications of balloon coarctation angioplasty. Prog Pediatr Cardiol. 2001;14:59–71.
    1. Risk of ionizing radiation exposure to children: a subject review. American Academy of Pediatrics, Committee on Environmental Health. Pediatrics. 1998;101:717–719.
    1. Raissaki M. Pediatric radiation protection. Eur Radiol Suppl. 2004;14:74–83.
    1. Committee on the Biological Effects of Ionizing Radiation BoRER, Commission on Life Sciences, National Research Council. Health Effects of Exposure to Low Levels of Ionizing Radiation, BEIR V Washington, DC: National Academy Press; 1990.
    1. Brenner D, Elliston C, Hall E, Berdon W. Estimated risks of radiation-induced fatal cancer from pediatric CT. AJR Am J Roentgenol. 2001;176:289–296.
    1. Infante-Rivarde C, Mathonnet G, Sinnet D. Risk of childhood leukemia associated with diagnostic irradiation and polymorphisms in DNA repair genes. Environ Health Perspect. 2000;108:495–498.
    1. McLaughlin JR, Kreiger N, Sloan MP, Benson LN, Hilditch S, Clarke EA. An historical cohort study of cardiac catheterization during childhood and the risk of cancer. Int J Epidemiol. 1993;22:584–591.
    1. Spengler RF, Cook DH, Clarke EA, Olley PM, Newman AM. Cancer mortality following cardiac catheterization: a preliminary follow-up study on 4,891 irradiated children. Pediatrics. 1983;71:235–239.
    1. Modan B, Keinan L, Blumstein T, Sadetzki S. Cancer following cardiac catheterization in childhood. Int J Epidemiol. 2000;29:424–428.
    1. Martin EC, Olson AP, Steeg CN, Casarella WJ. Radiation exposure to the pediatric patient during cardiac catheterization and angiocardiography: emphasis on the thyroid gland. Circulation. 1981;64:153–158.
    1. Wu JR, Huang TY, Wu DK, Hsu PC, Weng PS. Radiation exposure of pediatric patients and physicians during cardiac catheterization and balloon pulmonary valvuloplasty. Am J Cardiol. 1991;68:221–225.
    1. Berrington de Gonzalez A, Darby S. Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries. Lancet. 2004;363:345–351.
    1. Waldman JD, Rummerfield PS, Gilpin EA, Kirkpatrick SE. Radiation exposure to the child during cardiac catheterization. Circulation. 1981;64:158–163.
    1. Moore JD, Shim D, Sweet J, Arheart KL, Beekman RH., 3rd Radiation exposure to children during coil occlusion of the patent ductus arteriosus. Catheter Cardiovasc Interv. 1999;47:449–454.
    1. Boothroyd A, McDonald E, Moores BM, Sluming V, Carty H. Radiation exposure to children during cardiac catheterization. Br J Radiol. 1997;70:180–185.
    1. Rees S, Somerville J, Ward C, Martinez J, Mohiaddin RH, Underwood R, Longmore DB. Coarctation of the aorta: MR imaging in late postoperative assessment. Radiology. 1989;173:499–502.
    1. Nielsen JC, Powell AJ, Gauvreau K, Marcus EN, Prakash A, Geva T. Magnetic resonance imaging predictors of coarctation severity. Circulation. 2005;111:622–628.
    1. Oshinski JN, Parks WJ, Markou CP, Bergman HL, Larson BE, Ku DN, Mukundan S, Jr, Pettigrew RI. Improved measurement of pressure gradients in aortic coarctation by magnetic resonance imaging. J Am Coll Cardiol. 1996;28:1818–1826.
    1. Liu CY, Farahani K, Lu DS, Duckwiler G, Oppelt A. Safety of MRI-guided endovascular guidewire applications. J Magn Reson Imaging. 2000;12:75–78.
    1. Shellock FG, Shellock VJ. Metallic stents: evaluation of MR imaging safety. AJR Am J Roentgenol. 1999;173:543–547.

Source: PubMed

3
Subscribe