Evaluation of a single-channel portable monitor for the diagnosis of obstructive sleep apnea

Burcu Oktay, Thomas B Rice, Charles W Atwood Jr, Michael Passero Jr, Neeraj Gupta, Rachel Givelber, Oliver J Drumheller, Patricia Houck, Nancy Gordon, Patrick J Strollo Jr, Burcu Oktay, Thomas B Rice, Charles W Atwood Jr, Michael Passero Jr, Neeraj Gupta, Rachel Givelber, Oliver J Drumheller, Patricia Houck, Nancy Gordon, Patrick J Strollo Jr

Abstract

Study objective: To validate the ApneaLINK (AL) as an accurate tool for determining the presence of obstructive sleep apnea (OSA) in an at-risk sleep clinic population in a home test environment.

Methods: Consecutive participants referred with the suspicion of OSA were evaluated in the home with the AL portable monitor (AL Home), followed by simultaneous data collection with diagnostic polysomnography (PSG) and AL in the sleep laboratory (AL Lab). Prevalence, sensitivity, specificity, and ROC curves were calculated for PSG vs. AL Lab, PSG vs. AL Home, and AL Lab vs. AL Home test. Pearson correlations and Bland-Altman plots were constructed.

Results: Fifty-three (55% female) participants completed the entire study. The mean age of the population was 45.1 ± 11.3 years, and body mass index was 35.9 ± 9.1 kg/m(2). The prevalence of an apnea hypopnea index (AHI) ≥ 15 in the cohort was 35.9%. The results demonstrated a high sensitivity and specificity of the AL respiratory disturbance index (RDI-AL) compared with the AHI from the PSG. The AL Lab had the highest sensitivity and specificity at RDI-AL values ≥ 20 events/h (sensitivity 100%, specificity 92.5%). The AL Home was most sensitive and specific at an RDI-AL ≥ 20 events/h (sensitivity 76.9%, specificity 92.5%). The Pearson correlations for PSG vs. AL Lab and PSG vs. AL Home were ρ = 0.88 and ρ = 0.82, respectively. The Bland-Altman Plots demonstrated good agreement between the methodologies.

Conclusion: The AL home test is an accurate alternative to PSG in sleep clinic populations at risk for moderate and severe OSA.

Trial registration: clinicaltrials.gov ID: NCT00354614.

Keywords: Obstructive sleep apnea; diagnosis; portable monitoring.

Figures

Figure 1
Figure 1
Consort diagram *Evaluation time on AL home test

Figure 2

Picture of portable monitoring device

Figure 2

Picture of portable monitoring device

Figure 2
Picture of portable monitoring device

Figure 3

(A) Receiver-operator characteristic curves AL…

Figure 3

(A) Receiver-operator characteristic curves AL Lab RDI-AL and polysomnography AHI. (B) Bland-Altman Plot…
Figure 3
(A)Receiver-operator characteristic curves AL Lab RDI-AL and polysomnography AHI. (B) Bland-Altman Plot of AL Lab RDI-AL and Polysomnography AHI during simsltaneous testing. PSG, polysomnography; AUC, area under the curve.

Figure 4

Receiver-operator characteristic curves (A) RDI-AL…

Figure 4

Receiver-operator characteristic curves (A) RDI-AL Home versus RDI-AL from Lab study. (B) RDI-AL…

Figure 4
Receiver-operator characteristic curves (A) RDI-AL Home versus RDI-AL from Lab study. (B) RDI-AL Home versus AHI from PSG. PSG, polysomnography; AUC, area under the curve.

Figure 5

Bland-Altman Plots (A) RDI-AL Home…

Figure 5

Bland-Altman Plots (A) RDI-AL Home versus RDI-AL Lab. (B) RDI-AL Home versus polysomnography…

Figure 5
Bland-Altman Plots (A) RDI-AL Home versus RDI-AL Lab. (B) RDI-AL Home versus polysomnography AHI. PSG, polysomnography.
Figure 2
Figure 2
Picture of portable monitoring device
Figure 3
Figure 3
(A)Receiver-operator characteristic curves AL Lab RDI-AL and polysomnography AHI. (B) Bland-Altman Plot of AL Lab RDI-AL and Polysomnography AHI during simsltaneous testing. PSG, polysomnography; AUC, area under the curve.
Figure 4
Figure 4
Receiver-operator characteristic curves (A) RDI-AL Home versus RDI-AL from Lab study. (B) RDI-AL Home versus AHI from PSG. PSG, polysomnography; AUC, area under the curve.
Figure 5
Figure 5
Bland-Altman Plots (A) RDI-AL Home versus RDI-AL Lab. (B) RDI-AL Home versus polysomnography AHI. PSG, polysomnography.