- ICH GCP
- Registr klinických studií v USA
- Klinická studie NCT07724262
Evaluation of the Argos System to Estimate Cardiac Output
Evaluation of the Argos System to Estimate Cardiac Output in Critically Ill Patients
The goal of this observational study is to evaluate the reliability of the Argos system to track changes in cardiac output in critically ill patients. The main question it aims to answer is:
Does cardiac output changes induced by fluid administration or changes in norepinephrine dosage can be tracked by the Argos system, compared to transpulmonary thermodilution device.
Přehled studie
Detailní popis
In patients with acute circulatory failure, hemodynamic monitoring including cardiac index (CI) assessment is recommended for most severe patients. Transpulmonary thermodilution (TPTD) is one of the methods that are recommended for this purpose, in particular because it is considered as reliable in this type of patients. The PiCCO2 system (Pulsion Medical Systems, Getinge, Feldkirchen, Germany) relies on TPTD method and associate two methods to measure CI. On the one hand, TPTD is considered as one of the gold standards for CI evaluation (CITPTD). On the other hand, pulse contour analysis calculates stroke volume from the waveform of the systolic part of the arterial pressure curve (CIpwa). The way this estimation is performed and it is regularly re-evaluated differs among the devices that are available on the market. This estimation of CIpwa is calibrated when TPTD is performed. Thus, it allows a real-time pulse contour derived CI assessment, which is very precise. However, TPTD system are considered invasive, expensive and are not available in all settings.
The Argos monitor (Retia Medical, Valhalla, USA) uses the so-called multi-beat analysis (MBA) to estimate CI after analyzing the arterial pressure waveform over periods of multiple heartbeats (up to 20) and scaling CI estimations to biometric data. The algorithm estimates the theoretical central arterial blood pressure waveform using MBA, minimizing and avoiding the cofounding effect of the wave reflection phenomena. Based on this central arterial blood pressure waveform and biometric patient data, the system estimates cardiac output. The Argos monitor has a second advantage. It simply connects to the analogue output of the blood pressure signal from a compatible multi-parameter monitor and needs no additional specific pressure sensor. During cardiac surgery, the Argos monitor showed reasonable agreement and trending ability compared to the pulmonary artery catheter. However, in patients after cardiac surgery, including those with low cardiac output and arrhythmia, CI estimated by the device was not interchangeable with the continuous thermodilution method via a PAC.
The primary objective of the study is to evaluate the reliability of the Argos system to track changes of CI induced by changes in norepinephrine or fluid administration, compared to CITPTD in critically ill patients.
Typ studie
Zápis (Odhadovaný)
Kontakty a umístění
Studijní kontakt
- Jméno: Christopher Lai, MD, PhD
- Telefonní číslo: 12633816521
- E-mail: christopher.lai@aphp.fr
Studijní místa
-
-
-
Le Kremlin-Bicêtre, Francie
- Nábor
- Bicêtre Hospital
-
Kontakt:
- Christopher Lai
- Telefonní číslo: 2633816521
- E-mail: christopher.lai@aphp.fr
-
-
Kritéria účasti
Kritéria způsobilosti
Věk způsobilý ke studiu
- Dospělý
- Starší dospělý
Přijímá zdravé dobrovolníky
Metoda odběru vzorků
Studijní populace
Patients with shock defined by one sign of hypoperfusion:
- systolic arterial pressure <90mmHg
- mean arterial pressure <65mmHg
- prolonged capillary refill time >3 sec
- mottling score > 2
- urine output < 0.1 mL/kg/h
Popis
Inclusion Criteria:
- Age ≥18 years old
- Hospitalisation in intensive care
- Haemodynamic monitoring with transpulmonary thermodilution already in place
Hemodynamic intervention deemed necessary by the clinician in charge of the patient
- Administration of a fluid bolus
- Change in norepinephrine dosage or performance of a vasopressor test
Exclusion Criteria:
- Pregnancy
- Other therapeutic modification during volume expansion or the vasopressor test
- Refusal to participate by relatives of the patient or the patient himself or herself
- Non availability of investigators
Studijní plán
Jak je studie koncipována?
Detaily designu
Kohorty a intervence
Skupina / kohorta |
Intervence / Léčba |
|---|---|
|
patients with shock
Patients with shock defined by one sign of hypoperfusion:
|
Giving a fluid bolus of 500mL of normal saline (NaCl 0.9%)
brief, transient increase in norepinephrine to achieve a 15mmHg increase in mean arterial pressure and evaluates if tissue perfusion improves
|
Co je měření studie?
Primární výstupní opatření
Měření výsledku |
Časové okno |
|---|---|
|
To compare the changes in cardiac output provided by the Argos system induced by changes in norepinephrine or fluid bolus, compared to cardiac output provided by the transpulmonary thermodilution device in critically ill patients.
Časové okno: 1 hour (before and after intervention)
|
1 hour (before and after intervention)
|
Sekundární výstupní opatření
Měření výsledku |
Popis opatření |
Časové okno |
|---|---|---|
|
Accuracy of the Argos system
Časové okno: 1 hour (before and after intervention)
|
To evaluate the accuracy of the Argos system to measure absolute values of cardiac output compared to transpulmonary thermodilution
|
1 hour (before and after intervention)
|
|
Precision of the Argos system
Časové okno: 1 hour (before and after intervention)
|
To evaluate the precision and least significant change of cardiac output provided by the Argos system
|
1 hour (before and after intervention)
|
Spolupracovníci a vyšetřovatelé
Sponzor
Termíny studijních záznamů
Hlavní termíny studia
Začátek studia (Aktuální)
Primární dokončení (Odhadovaný)
Dokončení studie (Odhadovaný)
Termíny zápisu do studia
První předloženo
První předloženo, které splnilo kritéria kontroly kvality
První zveřejněno (Aktuální)
Aktualizace studijních záznamů
Poslední zveřejněná aktualizace (Aktuální)
Odeslaná poslední aktualizace, která splnila kritéria kontroly kvality
Naposledy ověřeno
Více informací
Termíny související s touto studií
Klíčová slova
Další relevantní podmínky MeSH
Další identifikační čísla studie
- Argos study
- CE-SRLF 24-100 (Jiný identifikátor: SRLF)
Plán pro data jednotlivých účastníků (IPD)
Plánujete sdílet data jednotlivých účastníků (IPD)?
Informace o lécích a zařízeních, studijní dokumenty
Studuje lékový produkt regulovaný americkým FDA
Studuje produkt zařízení regulovaný americkým úřadem FDA
produkt vyrobený a vyvážený z USA
Tyto informace byly beze změn načteny přímo z webu clinicaltrials.gov. Máte-li jakékoli požadavky na změnu, odstranění nebo aktualizaci podrobností studie, kontaktujte prosím register@clinicaltrials.gov. Jakmile bude změna implementována na clinicaltrials.gov, bude automaticky aktualizována i na našem webu .
Klinické studie na Šokovat
-
Cairo UniversityDokončeno
-
Cairo UniversityNábor
-
Nepal Health Research CouncilFogarty International Center of the National Institute of HealthNábor
-
University of WashingtonFogarty International Center of the National Institute of HealthDokončeno
-
Arrowhead Regional Medical CenterDokončenoPopáleniny | Nedostatek kyseliny askorbové | Nerovnováha tekutin a elektrolytů | Druhý stupeň spálení | Třetí stupeň spalování | Burn ShockSpojené státy
Klinické studie na Fluid bolus
-
Bausch & Lomb IncorporatedDokončenoSyndrom suchého okaSpojené státy
-
Breath of Life International Pharma LtdStaženo
-
University of California, San FranciscoNational Heart, Lung, and Blood Institute (NHLBI); National Institute of Nursing...DokončenoSrdeční selhání, městnavéSpojené státy
-
Università Politecnica delle MarcheUniversità Politecnica delle Marche, Ospedali riuniti di AnconaNáborMikrocirkulace | Žilní kongesce | Rozšíření objemuItálie
-
Breath of Life International Pharma LtdStaženoDiabetes mellitus typu 1
-
Breath of Life International Pharma LtdUkončenoDiabetická neuropatieIzrael
-
Bausch & Lomb IncorporatedUkončenoBolest | Zánět | Šedý zákalSpojené státy
-
Bausch & Lomb IncorporatedDokončeno
-
Burst BiologicsNeznámýPoruchy chodidel a kotníků | Spojení kotníku, OsteotomieSpojené státy
-
Burst BiologicsNeznámýDegenerativní onemocnění plotének | Spondylitida | Spinální stenóza | Spondylolistéza | Herniated DiskSpojené státy