Safety and Efficacy Study of MGuard Stent After a Heart Attack (MASTER)
28. maj 2013 opdateret af: InspireMD
MASTER: MGUARD for Acute ST Elevation Reperfusion
The goal of this study is to demonstrate the superiority of the MGuard™ stent over commercially-approved bare-metal (BMS) /drug-eluting stents (DES) in achieving better myocardial reperfusion in primary angioplasty for the treatment of acute ST-elevation myocardial infarction (STEMI).
Studieoversigt
Status
Afsluttet
Betingelser
Intervention / Behandling
Undersøgelsestype
Interventionel
Tilmelding (Faktiske)
433
Fase
- Fase 4
Kontakter og lokationer
Dette afsnit indeholder kontaktoplysninger for dem, der udfører undersøgelsen, og oplysninger om, hvor denne undersøgelse udføres.
Studiesteder
-
-
-
Białystok, Polen
- Klinika Kardiologii Inwazyjnej Uniwersytetu Medycznego w Białymstoku
-
Krakow, Polen
- Gilghsin: John Paul II Hospital
-
Krakow, Polen
- Szpital Uniwersytecki w Krakowie
-
Nowy Targ, Polen
- Oddział Kardiologii Inwazyjnej
-
Oświęcim, Polen
- Centrum Kardiologii Inwazyjnej GVM Carint
-
Warsaw, Polen
- Centralny Szpital Kliniczny MSWiA w Warszawie
-
-
-
-
-
Johannesburg, Sydafrika
- Milpark Hospital
-
-
Deltagelseskriterier
Forskere leder efter personer, der passer til en bestemt beskrivelse, kaldet berettigelseskriterier. Nogle eksempler på disse kriterier er en persons generelle helbredstilstand eller tidligere behandlinger.
Berettigelseskriterier
Aldre berettiget til at studere
18 år og ældre (Voksen, Ældre voksen)
Tager imod sunde frivillige
Ingen
Køn, der er berettiget til at studere
Alle
Beskrivelse
Inclusion Criteria
- 18 years of age
- ST-segment elevation (more than 2mm in more than contiguous leads)
- MI with symptom onset less than 12h
- The patient is willing to comply with specified follow-up evaluations
- Signed ICF
- Single de novo lesion in the target (culprit) vessel
- Target lesion maximum length is 33 mm (by visual estimation)
- Reference vessel diameter must be more than 3.0 to less than 4.0 mm by visual estimation
- Randomization should occur as soon as Presence of TIMI 2 or 3 before randomization Exclusion Criteria
- Pregnant or nursing patients
- Left Bundle Branch Block (LBBB), paced rhythm, or other Electrocardiogram (ECG) abnormality
- Impaired renal function
- Prior coronary artery bypass graft surgery
- Bleeding diathesis
- Contraindication to aspirin
- cardiopulmonary resuscitation
- Cardiogenic shock
- chronic warfarin anticoagulation
- LVEF less than 20%
- other medical illness
- participation in another investigational drug or device study that has not reached its primary endpoint
- Left main coronary artery disease with 50% stenosis
- Ostial target lesion
- Failure to visualize vessel anatomy distal to the culprit lesion
- Moderate to heavily calcified target lesion or vessel
- excessive tortuosity
- bifurcation with a side branch more than 2.0 mm in diameter
- A significant (greater than 50%) stenosis proximal or distal to the target lesion is present that cannot be covered by same single stent
- Diffuse disease distal to target lesion with impaired runoff
- Any prior stent proximal to the target lesion, or within 10 mm distal of the target lesion
- PCI of another lesion performed within 6 months before the index procedure
- Target lesion located in a saphenous vein graft
Studieplan
Dette afsnit indeholder detaljer om studieplanen, herunder hvordan undersøgelsen er designet, og hvad undersøgelsen måler.
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Primært formål: Behandling
- Tildeling: Randomiseret
- Interventionel model: Parallel tildeling
- Maskning: Enkelt
Våben og indgreb
Deltagergruppe / Arm |
Intervention / Behandling |
|---|---|
|
Eksperimentel: MGuard
MGuard stent will be deployed
|
MGuard™ stent comprises a balloon-expandable, thin-strut stainless steel (316L) bare metal stent platform (strut width 100 µm) with mesh sleeve fibers of polyethyleneterephtalate (fiber width of 20 µm) attached to its outer surface.
These fibers act like a net (aperture size 150 x 180 µm) preventing distal embolization of the plaque debris/thrombus placed between the vessel wall and the stent.
|
|
Aktiv komparator: BMS or DES
A regular bare metal stent or drug-eluting stent will be deployed
|
Control BMS or DES
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Tidsramme |
|---|---|
|
The incidence of complete ST segment resolution (defined as ≥70% ST 1. The incidence of complete ST segment resolution
Tidsramme: 60 to 90 minutes after the last angiogram
|
60 to 90 minutes after the last angiogram
|
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
The incidence of Thrombolysis In Myocardial Infarction (TIMI) 3 flow at the end of the procedure.
Tidsramme: 60-90 minutes after last angiogram
|
The incidence of TIMI 3 flow at the end of the procedure.
|
60-90 minutes after last angiogram
|
|
Major Adverse Cardiac Events (MACE)at discharge, 30 days, 6 months and 12 months post-procedure
Tidsramme: discharge, 30 days, 6 and 12 months post-procedure
|
Major Adverse Cardiac Events (MACE): defined as cardiac death, reinfarction (Q wave and non-Q wave), or repeat ischemia-driven target lesion revascularization (TLR) by percutaneous or surgical methods at hospital discharge, 30 days, 6 months and 1 year post-procedure.
|
discharge, 30 days, 6 and 12 months post-procedure
|
Samarbejdspartnere og efterforskere
Det er her, du vil finde personer og organisationer, der er involveret i denne undersøgelse.
Sponsor
Efterforskere
- Ledende efterforsker: Alexandre Abizaid, MD, PhD, Inst Dante Pazzanese of Cardiology, Brazil
- Ledende efterforsker: Dariusz Dudek, MD, PhD, Cardiac Catheterization Laboratories, Krakow, Poland
- Ledende efterforsker: Sigmund Silber, MD, PhD, Heart Center at the Isar Academic Teaching Site of the University of Munich
- Studiestol: Gregg Stone, MD, Columbia University Medical Center The Cardiovascular Research Foundation
Publikationer og nyttige links
Den person, der er ansvarlig for at indtaste oplysninger om undersøgelsen, leverer frivilligt disse publikationer. Disse kan handle om alt relateret til undersøgelsen.
Generelle publikationer
- Dudek D, Dziewierz A, Brener SJ, Abizaid A, Merkely B, Costa RA, Bar E, Rakowski T, Kornowski R, Dressler O, Abizaid A, Silber S, Stone GW. Mesh-covered embolic protection stent implantation in ST-segment-elevation myocardial infarction: final 1-year clinical and angiographic results from the MGUARD for acute ST elevation reperfusion trial. Circ Cardiovasc Interv. 2015 Feb;8(2):e001484. doi: 10.1161/CIRCINTERVENTIONS.114.001484.
- Dudek D, Brener SJ, Rakowski T, Dziewierz A, Abizaid A, Silber S, Yaacoby E, Dizon JM, Costa RA, Maehara A, Dressler O, Stone GW. Efficacy of an Embolic Protection Stent as a Function of Delay to Reperfusion in ST-Segment Elevation Myocardial Infarction (from the MASTER Trial). Am J Cardiol. 2014 Nov 15;114(10):1485-9. doi: 10.1016/j.amjcard.2014.08.007. Epub 2014 Aug 27.
- Stone GW, Abizaid A, Silber S, Dizon JM, Merkely B, Costa RA, Kornowski R, Abizaid A, Wojdyla R, Maehara A, Dressler O, Brener SJ, Bar E, Dudek D. Prospective, Randomized, Multicenter Evaluation of a Polyethylene Terephthalate Micronet Mesh-Covered Stent (MGuard) in ST-Segment Elevation Myocardial Infarction: The MASTER Trial. J Am Coll Cardiol. 2012 Nov 6;60(19):1975-84. doi: 10.1016/j.jacc.2012.09.004.
Datoer for undersøgelser
Disse datoer sporer fremskridtene for indsendelser af undersøgelsesrekord og resumeresultater til ClinicalTrials.gov. Studieregistreringer og rapporterede resultater gennemgås af National Library of Medicine (NLM) for at sikre, at de opfylder specifikke kvalitetskontrolstandarder, før de offentliggøres på den offentlige hjemmeside.
Studer store datoer
Studiestart
1. juni 2011
Primær færdiggørelse (Faktiske)
1. juni 2012
Studieafslutning (Faktiske)
1. juni 2012
Datoer for studieregistrering
Først indsendt
17. februar 2011
Først indsendt, der opfyldte QC-kriterier
7. juni 2011
Først opslået (Skøn)
8. juni 2011
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Skøn)
29. maj 2013
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
28. maj 2013
Sidst verificeret
1. maj 2013
Mere information
Begreber relateret til denne undersøgelse
Nøgleord
Yderligere relevante MeSH-vilkår
Andre undersøgelses-id-numre
- IMD-07
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Ingen
Studerer et amerikansk FDA-reguleret enhedsprodukt
Ingen
produkt fremstillet i og eksporteret fra U.S.A.
Ingen
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .