- ICH GCP
- US Clinical Trials Registry
- Klinisk forsøg NCT07560085
Catheter-Oriented Unloading and Recovery With a Left Ventricular Assist Gear: Efficacy in Post Cardiotomy Cardiogenic Shock (COURAGE)
Efficacy and Safety of SynFlow Duro System in the Treatment of Post Cardiotomy Cardiogenic Shock: a Prospective, Multicenter, Single-arm Clinical Trial
Studieoversigt
Status
Betingelser
Intervention / Behandling
Detaljeret beskrivelse
Undersøgelsestype
Tilmelding (Anslået)
Fase
- Ikke anvendelig
Kontakter og lokationer
Studiekontakt
- Navn: Chen Jiang
- Telefonnummer: +86 13918357410
- E-mail: cjiang@forqaly.com
Studiesteder
-
-
Fujian
-
Fuzhou, Fujian, Kina
- Fujian Medical University Union Hospital
-
Kontakt:
- Liangwan Chen
- E-mail: chenliangwan@fjmu.edu.cn
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
- Voksen
- Ældre voksen
Tager imod sunde frivillige
Beskrivelse
Inclusion Criteria:
- 18≤Age≤80, any sex.
- Failure to wean from CPB or refractory cardiogenic shock after cardiac surgery.
Failure to wean from CPB is defined as failed to discontinue CPB or post-weaning CI <2.2L/min/m2 + PCWP >15mmHg, despite optimal pre-weaning management and receiving at least one high dose inotrope.
Refractory cardiogenic shock is defined as a condition that occurs despite adequate fluid status and meets all the following criteria:
① CI<2.2L/min/m2 + PCWP>15mmHg despite receiving at least one high dose inotrope or IABP; OR SBP <80mmHg or MAP<50mmHg despite receiving at least one high dose intrope + high dose norepinephrine or IABP.
② Presence of at least one signs indicating hypoperfusion: decreased mentation; cold, clammy, ashen or cyanotic extremities or skin or livedo reticularis; urine output<30ml/h or 0.5ml/kg/h; or lactate>2mmol/L or metabolic acidosis.
- Patient has signed the informed consent form.
Exclusion Criteria:
- Body surface area>2.5m2.
- Presence of any cardiac assist device (other than an IABP).
- Right ventricular failure.
- Evidence of any vascular disease or anatomy precluding placement or deployment of the device (e.g. severely calcified vessel).
- Evidence of ventricular thrombus.
- Moderate or severe aortic valve insufficiency/stenosis, or severe aortic valve calcification.
- Presence of mechanical aortic valve or cardiac contractile device.
- Obstructive, hypertrophic cardiomyopathy.
- Presence of uncorrected ventricular septal defect, atrial septal defect or patent foramen ovale.
- Mechanical manifestation of AMI (e.g. ventricular septal rupture, papillary muscle rupture, ventricular rupture).
- Any hematological disorder causing fragility of blood vessel or hemolysis.
- Known allergy or intolerant to heparin.
- Presence or suspicion of active systemic infection.
- Cardiopulmonary resuscitative maneuver lasting longer than 20 minutes before device placement.
- Sustained or non-sustained ventricular tachycardia or ventricular fibrillation unresponsive to medical therapy.
- Severe hepatic dysfunction, renal failure or severe respiratory failure
- Pregnant or lactating female.
- Concomitant enrollment in another investigational drug or device study where the primary endpoint has not yet been completed or that will clinically interfere with the endpoint of this investigation.
- Other conditions considered unsuitable for participating in this investigation by the investigators.
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
- Primært formål: Behandling
- Tildeling: N/A
- Interventionel model: Enkelt gruppeopgave
- Maskning: Ingen (Åben etiket)
Våben og indgreb
Deltagergruppe / Arm |
Intervention / Behandling |
|---|---|
|
Eksperimentel: SynFlow Duro
|
a novel micro-axial flow pump for temporary left ventricular assistance
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Device support success
Tidsramme: Up to 30 days post-explant or hospital discharge (whichever is longer)
|
Cardiac recovery (survival to 30 days post-explant or hospital discharge [which is longer] without other mechanical support including IABP) or briging to other durable cardiac therapies (heart transplantation or implantation of approved durable ventricular assist device)
|
Up to 30 days post-explant or hospital discharge (whichever is longer)
|
Sekundære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
Technical success rate
Tidsramme: Through device explantation, up to 14 days
|
Technical success is defined as successful device placement, deployment, startup and explantation without any device malfunction
|
Through device explantation, up to 14 days
|
|
Device support duration
Tidsramme: Through device explantation, up to 14 days
|
Time from device startup to device explanation
|
Through device explantation, up to 14 days
|
|
Time to recovery
Tidsramme: From device placement to 30 days post-explant or hospital discharge (whichever is longer)
|
Recovery is defined as survival to 30 days post-explant or hospital discharge (which is longer) without other mechanical support including IABP
|
From device placement to 30 days post-explant or hospital discharge (whichever is longer)
|
|
Major adverse event
Tidsramme: From device placement to 30 days post-explant or hospital discharge or bridging to other durable cardiac therapies
|
Defined as stroke or death
|
From device placement to 30 days post-explant or hospital discharge or bridging to other durable cardiac therapies
|
|
Minor adverse event
Tidsramme: From device placement to 30 days post-explant or hospital discharge or bridging to other durable cardiac therapies
|
Defined as minor or major bleeding, hemolysis, limb ischemia, requiring continuous renal placement therapy, ventricular tachycardia or fibrillation requiring cardioversion, sepsis, mechanical complications and major access vascular complications.
|
From device placement to 30 days post-explant or hospital discharge or bridging to other durable cardiac therapies
|
|
Adverse event
Tidsramme: From device placement to 30 days post-explant or hospital discharge or bridging to other durable cardiac therapies
|
Defined as untoward medical event occurring during the course of clinical investigation, whether or not related to the medical device.
|
From device placement to 30 days post-explant or hospital discharge or bridging to other durable cardiac therapies
|
|
Serious adverse event
Tidsramme: From device placement to 30 days post-explant or hospital discharge or bridging to other durable cardiac therapies
|
Defined as the event occurring during the course of clinical investigation, which leads to death or serious deterioration in the health, including a life-threatening illness or injury; a permanent impairment of a body structure or a body function; in-patient or prolonged hospitalization; medical or surgical intervention to prevent permanent impairment to a body structure or a body function; foetal distress, foetal death, congenital abnormality or birth defect, etc.
|
From device placement to 30 days post-explant or hospital discharge or bridging to other durable cardiac therapies
|
|
Device deficiency
Tidsramme: Through device explantation, up to 14 days
|
Defined as the unreasonable risk that may endanger human health and life safety in the normal use of medical devices during clinical investigations, such as labeling errors, quality problems, malfunctions, etc.
|
Through device explantation, up to 14 days
|
|
Cardiac index change versus baseline
Tidsramme: Within 24 hours after device startup
|
Within 24 hours after device startup
|
|
|
Cardiac output change versus baseline
Tidsramme: Within 24 hours after device startup
|
Within 24 hours after device startup
|
|
|
Pulmonary artery pressure change versus baseline
Tidsramme: Within 24 hours after device startup
|
Within 24 hours after device startup
|
|
|
Mean arterial pressure change versus baseline
Tidsramme: Within 24 hours after device startup
|
Within 24 hours after device startup
|
|
|
Central venous pressure change versus baseline
Tidsramme: Within 24 hours after device startup
|
Within 24 hours after device startup
|
Samarbejdspartnere og efterforskere
Efterforskere
- Studiestol: Liangwan Chen, Fujian Medical University Union Hospital
Datoer for undersøgelser
Studer store datoer
Studiestart (Anslået)
Primær færdiggørelse (Anslået)
Studieafslutning (Anslået)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Faktiske)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Yderligere relevante MeSH-vilkår
Andre undersøgelses-id-numre
- CSPR001
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Studerer et amerikansk FDA-reguleret enhedsprodukt
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med Kardiogent stød
-
National Institute of Allergy and Infectious Diseases...Afsluttet
-
Sun Yat-Sen Memorial Hospital of Sun Yat-Sen UniversityRekrutteringShock Wave Terapi for PatellartendinitKina
-
Biomedizinische Forschungs gmbHMedical University of ViennaAfsluttet
-
Biomedizinische Forschungs gmbHMedical University of ViennaAfsluttetToxic Shock Syndrome | Vaccination; SepsisØstrig
-
Biomedizinische Forschungs gmbHMedical University of ViennaAfsluttetSepsis | Toksisk-chok syndrom
-
Hospices Civils de LyonAfsluttetToxic Shock SyndromeFrankrig
-
B. Braun Melsungen AGAfsluttetNedsat og uspecifikt blodtryksforstyrrelser og shockKina
-
Cairo UniversityAfsluttet
-
Cairo UniversityRekruttering
-
The Children's Hospital of Zhejiang University...Ikke rekrutterer endnuRandomiseret kontrolleret forsøg | Ekstrakorporeal Shock Wave Lithotripsi | Børne-urolithiasis | Optisk Bevægelsesoptagelse | Ultrasound Lokalisering | LokaliseringseffektivitetKina
Kliniske forsøg med SynFlow Duro
-
ForQaly Medical (Shanghai) Co., LtdAfsluttet