- ICH GCP
- US Clinical Trials Registry
- Klinisk forsøg NCT07690254
Bispectral Index and First-Pass Intubation
Evaluation of First-Pass Endotracheal Intubation Success Using the Bispectral Index in Patients Undergoing Endotracheal Intubation in the Emergency Department
This prospective observational study aims to evaluate the association between anesthetic depth, measured by the Bispectral Index (BIS), and first-pass endotracheal intubation success in adult patients undergoing rapid sequence intubation (RSI) in the emergency department.
Endotracheal intubation is one of the most critical procedures performed in emergency medicine. Multiple intubation attempts are associated with an increased risk of adverse events, including hypoxemia, aspiration, cardiovascular complications, and death. Although several factors influence first-pass intubation success, the contribution of anesthetic depth at the time of intubation has not been adequately investigated in the emergency department.
Patients for whom the treating emergency physician decides to perform RSI will undergo noninvasive BIS monitoring using a disposable forehead sensor. The physician performing the intubation will remain blinded to the BIS value throughout the procedure and will perform intubation according to routine clinical practice. A separate investigator, who is not involved in patient management, will record the BIS value at the time of intubation.
The BIS measurements will not be disclosed to the treating team and will not influence clinical decision-making, medication dosing, or patient management. The study is purely observational and does not evaluate the performance or effectiveness of the BIS device itself.
The primary objective is to determine whether patients who achieve first-pass intubation success differ in anesthetic depth from those requiring multiple intubation attempts.
Studieoversigt
Status
Betingelser
Detaljeret beskrivelse
Endotracheal intubation is a life-saving intervention frequently performed in the emergency department for patients with airway compromise, respiratory failure, or critical illness. Despite its importance, emergency airway management is associated with a substantial risk of adverse events. Previous studies have consistently demonstrated that multiple intubation attempts significantly increase the likelihood of hypoxemia, aspiration, hypotension, cardiac arrest, airway trauma, and mortality. Therefore, achieving first-pass success has become one of the principal quality indicators in emergency airway management.
Several patient-, operator-, and procedure-related factors influence first-pass success, including anatomical difficulty, operator experience, type of laryngoscope, choice of sedative and neuromuscular blocking agents, and optimization of preoxygenation. Another potentially important but understudied factor is the depth of anesthesia at the moment of laryngoscopy and intubation.
Adequate anesthetic depth before laryngoscopy suppresses airway reflexes, minimizes patient movement, and improves intubating conditions. Previous studies comparing succinylcholine and rocuronium have suggested differences in intubation conditions; however, subsequent investigations demonstrated that these differences were largely attributable to dosing strategies rather than the neuromuscular blocking agent itself. Current evidence indicates that when appropriate doses are administered, both agents provide comparable intubating conditions. Nevertheless, regardless of the neuromuscular blocking agent used, insufficient hypnotic depth may adversely affect intubation conditions.
The Bispectral Index (BIS) is a processed electroencephalographic parameter ranging from 0 to 100 that provides an objective assessment of the patient's level of consciousness. BIS monitoring has been extensively validated in the operating room and has demonstrated a strong correlation with anesthetic depth. Randomized clinical trials have shown that BIS-guided anesthesia reduces the incidence of intraoperative awareness and may facilitate individualized anesthetic administration. However, BIS monitoring is not routinely incorporated into emergency airway management, and limited evidence exists regarding its relationship with first-pass intubation success in emergency department patients.
This study is designed as a prospective observational study. Adult patients undergoing rapid sequence intubation after a clinical decision by the treating emergency physician will be consecutively enrolled. A disposable BIS sensor will be placed on the patient's forehead immediately before induction, and BIS monitoring will be performed continuously. The emergency physician performing the intubation will remain blinded to the BIS values throughout the procedure. A separate investigator, independent of clinical management, will record the BIS value at the time of laryngoscopy and intubation.
Importantly, BIS measurements will not be communicated to the treating physician and will not influence the timing of intubation, medication selection, medication dosing, or any other aspect of patient care. Therefore, this investigation is not intended to evaluate or validate the BIS device as a medical device intervention. Rather, BIS is used solely as a noninvasive measurement tool to objectively quantify anesthetic depth. The monitor consists of a disposable adhesive forehead sensor and poses minimal risk to participants.
The primary analysis will compare BIS values between patients with successful first-pass intubation and those requiring more than one intubation attempt. The study seeks to determine whether inadequate anesthetic depth is associated with failure of first-pass intubation. The findings may improve understanding of the role of anesthetic depth during emergency airway management and may contribute to strategies aimed at increasing first-pass success while reducing intubation-related complications.
Undersøgelsestype
Tilmelding (Anslået)
Kontakter og lokationer
Studiekontakt
- Navn: Muhammed Türe, MD
- Telefonnummer: +905054691764
- E-mail: erkangoksu@akdeniz.edu.tr
Undersøgelse Kontakt Backup
- Navn: Erkan Göksu, MD
- Telefonnummer: 00905054691764
- E-mail: erkangoksu@akdeniz.edu.tr
Studiesteder
-
-
-
Antalya, Tyrkiet (Türkiye), 07059
- Rekruttering
- Akdeniz University Hospital
-
Kontakt:
- Erkan Göksu, MD
- Telefonnummer: 05054691764
- E-mail: erkangoksu@akdeniz.edu.tr
-
-
Deltagelseskriterier
Berettigelseskriterier
Aldre berettiget til at studere
- Voksen
- Ældre voksen
Tager imod sunde frivillige
Prøveudtagningsmetode
Studiebefolkning
Beskrivelse
Inclusion Criteria:
- Adults aged 18 years or older
- Patients undergoing rapid sequence intubation (RSI) in the emergency department
- No traumatic injury to the frontal region that would interfere with placement of the Bispectral Index (BIS) sensor
Exclusion Criteria:
- Age younger than 18 years
- Pregnant patients
- Patients with cardiac arrest or hemodynamic instability at the time of enrollment
- Patients intubated without the rapid sequence intubation (RSI) protocol
- Refusal to participate in the study (by the patient or, where applicable, a legally authorized representative)
Studieplan
Hvordan er undersøgelsen tilrettelagt?
Design detaljer
Kohorter og interventioner
Gruppe / kohorte |
|---|
|
Adults Undergoing Rapid Sequence Intubation
Adult patients (≥18 years) undergoing rapid sequence intubation (RSI) in the emergency department following the treating physician's clinical decision.
A noninvasive Bispectral Index (BIS) sensor is placed on the patient's forehead before induction to continuously monitor anesthetic depth.
The treating physician remains blinded to the BIS values, which are recorded by an independent investigator and are not used to guide patient management or influence the intubation procedure.
Participants are analyzed according to first-pass intubation success.
|
Hvad måler undersøgelsen?
Primære resultatmål
Resultatmål |
Foranstaltningsbeskrivelse |
Tidsramme |
|---|---|---|
|
First-pass endotracheal intubation success
Tidsramme: During the intubation procedure
|
Intubation at first attempt
|
During the intubation procedure
|
Samarbejdspartnere og efterforskere
Sponsor
Efterforskere
- Studieleder: Erkan Göksu, Akdeniz University School of Medicine Department of Emergency Medicine
Datoer for undersøgelser
Studer store datoer
Studiestart (Faktiske)
Primær færdiggørelse (Anslået)
Studieafslutning (Anslået)
Datoer for studieregistrering
Først indsendt
Først indsendt, der opfyldte QC-kriterier
Først opslået (Faktiske)
Opdateringer af undersøgelsesjournaler
Sidste opdatering sendt (Faktiske)
Sidste opdatering indsendt, der opfyldte kvalitetskontrolkriterier
Sidst verificeret
Mere information
Begreber relateret til denne undersøgelse
Andre undersøgelses-id-numre
- AUH (Aarhus University Hospital)
Plan for individuelle deltagerdata (IPD)
Planlægger du at dele individuelle deltagerdata (IPD)?
IPD-planbeskrivelse
IPD-delingstidsramme
IPD-delingsadgangskriterier
IPD-deling Understøttende informationstype
- SAP
- ANALYTIC_CODE
Lægemiddel- og udstyrsoplysninger, undersøgelsesdokumenter
Studerer et amerikansk FDA-reguleret lægemiddelprodukt
Studerer et amerikansk FDA-reguleret enhedsprodukt
produkt fremstillet i og eksporteret fra U.S.A.
Disse oplysninger blev hentet direkte fra webstedet clinicaltrials.gov uden ændringer. Hvis du har nogen anmodninger om at ændre, fjerne eller opdatere dine undersøgelsesoplysninger, bedes du kontakte register@clinicaltrials.gov. Så snart en ændring er implementeret på clinicaltrials.gov, vil denne også blive opdateret automatisk på vores hjemmeside .
Kliniske forsøg med Emergency Airway Management
-
Akdeniz University HospitalAfsluttetEmergency Airway Management | Gastric Inflation Risk During Bag-Valve-Mask Ventilation | Breathing EmergencyTyrkiet (Türkiye)
-
Insel Gruppe AG, University Hospital BernGaslini Children's HospitalTilmelding efter invitationAnæstesi | Trakeostomi komplikation | Emergency Front of Neck Airway hos børnSchweiz
-
University Children's Hospital, ZurichAfsluttetEmergency Front of Neck Airway hos børnSchweiz
-
Bezmialem Vakif UniversityAfsluttetLuftvejskomplikation af anæstesi | Luftvejsaspiration | Komplikation af anæstesi | Øsofagus skade | Supraglottisk luftvejseffektivitet | Endoskopisk ergonomi | ERCP Airway ManagementKalkun
-
The University of Texas Health Science Center,...AfsluttetEndotracheal intubation | Rapid Airway Management PositionerForenede Stater
-
University of Texas at AustinRekruttering
-
Hospital Israelita Albert EinsteinAfsluttetPatienter Emergency On-site Care ved Mobile Emergency UnitBrasilien
-
University of MalayaUkendtGenerel anæstesi | Supraglottic Airway Device | Larynx maske Airway
-
Sun Yat-Sen Memorial Hospital of Sun Yat-Sen UniversityAktiv, ikke rekrutterende
-
Masarykova Nemocnice v Usti nad Labem, Krajska...AfsluttetSupraglottic AirwayTjekkiet