EMG-Based Hand-Wrist Control: Study B Mirrored (PSICON-B)

30 de diciembre de 2020 actualizado por: Liberating Technologies, Inc.

EMG-Based Hand-Wrist Control: Study B: Comparison of Contralateral (Mirrored) EMG-

Study assessing four-channel prosthesis controller, that compares contralateral (mirrored) EMG-force training to ipsilateral EMG-target training with both limb-absent and able-bodied subjects

Descripción general del estudio

Descripción detallada

System identification models relating forearm electromyogram (EMG) signals to phantom wrist radial-ulnar deviation force, pronation-supination moment and/or hand open-close force (EMG-force) are hampered by lack of supervised force/moment output signals in limb-absent subjects. In able-bodied and unilateral transradial limb-absent subjects, we studied three alternative supervised output sources in one degree of freedom (DoF) and 2-DoF target tracking tasks: (1) bilateral tracking with force feedback from the contralateral side (non-dominant for able-bodied/ sound for limb-absent subjects) with the contralateral force as the output, (2) bilateral tracking with force feedback from the contralateral side with the target as the output, and (3) dominant/limb-absent side unilateral target tracking without feedback and the target used as the output.

Tipo de estudio

Intervencionista

Inscripción (Actual)

20

Fase

  • Fase temprana 1

Contactos y Ubicaciones

Esta sección proporciona los datos de contacto de quienes realizan el estudio e información sobre dónde se lleva a cabo este estudio.

Ubicaciones de estudio

    • Massachusetts
      • Holliston, Massachusetts, Estados Unidos, 01746
        • Liberating Technologies, Inc
      • Worcester, Massachusetts, Estados Unidos, 01609
        • Worcester Polytechnic Institute

Criterios de participación

Los investigadores buscan personas que se ajusten a una determinada descripción, denominada criterio de elegibilidad. Algunos ejemplos de estos criterios son el estado de salud general de una persona o tratamientos previos.

Criterio de elegibilidad

Edades elegibles para estudiar

18 años y mayores (Adulto, Adulto Mayor)

Acepta Voluntarios Saludables

No

Géneros elegibles para el estudio

Todos

Descripción

Inclusion Criteria:

  • Limb-absent subjects: unilateral trans-radial limb absence or amputation
  • Be capable of completing the requested contractions on the affected dominant side

Exclusion Criteria:

  • Past injuries to the upper limbs that would limit their ability to complete the requested contractions
  • Scars that would impede the use of surface electrodes

Plan de estudios

Esta sección proporciona detalles del plan de estudio, incluido cómo está diseñado el estudio y qué mide el estudio.

¿Cómo está diseñado el estudio?

Detalles de diseño

  • Propósito principal: Ciencia básica
  • Asignación: No aleatorizado
  • Modelo Intervencionista: Asignación Secuencial
  • Enmascaramiento: Ninguno (etiqueta abierta)

Armas e Intervenciones

Grupo de participantes/brazo
Intervención / Tratamiento
Experimental: Study B: Limb-Absent Subject
Ten unilateral transradial limb-absent subjects will each participate in one, half-day experimental session. Subjects will be seated and prepared in test apparatus seat (16 electrodes on the affected side, hand-wrist on the able side secured to load cells). Subjects will complete the 1-DoF and 2-DoF "dynamic" (force-varying) contractions (40-s duration, 0.75 Hz bandlimited, uniform random target). With 2 DoF contraction trials, hand Opn-Cls will always be one of the dimensions. The subject will then be released from the cuff and their able side not further involved in the experiment. The force feedback triangle cursor on the computer screen will be deleted such that only the target remains. Subjects will then repeat 1-DoF and 2-DoF trials in which the affected side attempts to produce hand-wrist effort that mimic movement of the target (with no feedback provided).
Test apparatus acquires 16 sEMG channels, measure four DoFs of force/moment at the hand-wrist [hand open-close (Opn-Cls); wrist extension-flexion (Ext-Flx), radial-ulnar deviation (Rad-Uln) and pronation-supination (Pro-Sup)]. sEMG system used 16 custom encased bipolar electrodes. A commercial LTI EMG amplifier (BE328) conditioned each EMG signal before A/D conversion on a PC. A thermoplastic hand cuff secured the wrist to a six-axis load cell (AMTI, model MC3A-100) for force/moment measurement. A separate hand-grasp one-axis load cell (Omega Engineering Inc., model LCR-150) was Velcro-secured between the fingers and the thumb, to measure power grip forces. The PC acquired and stored the sEMG and load cell data, and commanded a triangular screen target. The subject controlled a second triangular cursor. Left-right movement of the cursor was controlled by wrist Ext-Flx, up-down movement by Rad-Uln deviation, cursor rotation by Pro-Sup and cursor size by hand Opn-Cls forces.
Otros nombres:
  • load cell
  • electrode array
Experimental: Study B: Able-bodied Subjects
Ten able-bodied subjects, the electrodes will be mounted on the dominant arm for the EMG-force and EMG-target tracking trials. In addition, the non-dominant arm will also be constrained and measured in a second load cell. This load cell will not be used for feedback during the experiment, but will compare (RMS error, off-line) the dominant vs. non-dominant forces. For EMG-target tracking, the dominant hand will remain in the wrist cuff (to prevent flailing during contractions), with the screen feedback disabled. These subjects will repeat the trials with the electrodes moved to the non-dominant side. EMG-force tracking will be repeated using mirrored contractions. The three training methods (EMG-force ipsilateral, EMG-force contralateral mirrored, EMG-target on the dominant side) will be contrasted to help understand the source of errors when training with limb-absent subjects.
Test apparatus acquires 16 sEMG channels, measure four DoFs of force/moment at the hand-wrist [hand open-close (Opn-Cls); wrist extension-flexion (Ext-Flx), radial-ulnar deviation (Rad-Uln) and pronation-supination (Pro-Sup)]. sEMG system used 16 custom encased bipolar electrodes. A commercial LTI EMG amplifier (BE328) conditioned each EMG signal before A/D conversion on a PC. A thermoplastic hand cuff secured the wrist to a six-axis load cell (AMTI, model MC3A-100) for force/moment measurement. A separate hand-grasp one-axis load cell (Omega Engineering Inc., model LCR-150) was Velcro-secured between the fingers and the thumb, to measure power grip forces. The PC acquired and stored the sEMG and load cell data, and commanded a triangular screen target. The subject controlled a second triangular cursor. Left-right movement of the cursor was controlled by wrist Ext-Flx, up-down movement by Rad-Uln deviation, cursor rotation by Pro-Sup and cursor size by hand Opn-Cls forces.
Otros nombres:
  • load cell
  • electrode array

¿Qué mide el estudio?

Medidas de resultado primarias

Medida de resultado
Medida Descripción
Periodo de tiempo
EMG-Force Models
Periodo de tiempo: 1 day visit

Electromyogram waveform measurements (signal amplitude, frequency) will be related to force measurements (magnitude and moment) as related to the target tracking tasks. These signal-force data will be use to generate the models.

The specific outcome for each subject will be the RMS error measured in the tracking tasks as the difference between the generated control signal from the user versus the prompted target signal, as measured by an EMG array.

1 day visit

Colaboradores e Investigadores

Aquí es donde encontrará personas y organizaciones involucradas en este estudio.

Fechas de registro del estudio

Estas fechas rastrean el progreso del registro del estudio y los envíos de resultados resumidos a ClinicalTrials.gov. Los registros del estudio y los resultados informados son revisados ​​por la Biblioteca Nacional de Medicina (NLM) para asegurarse de que cumplan con los estándares de control de calidad específicos antes de publicarlos en el sitio web público.

Fechas importantes del estudio

Inicio del estudio (Actual)

29 de junio de 2017

Finalización primaria (Actual)

24 de agosto de 2020

Finalización del estudio (Actual)

24 de agosto de 2020

Fechas de registro del estudio

Enviado por primera vez

28 de diciembre de 2020

Primero enviado que cumplió con los criterios de control de calidad

30 de diciembre de 2020

Publicado por primera vez (Actual)

5 de enero de 2021

Actualizaciones de registros de estudio

Última actualización publicada (Actual)

5 de enero de 2021

Última actualización enviada que cumplió con los criterios de control de calidad

30 de diciembre de 2020

Última verificación

1 de diciembre de 2020

Más información

Términos relacionados con este estudio

Términos MeSH relevantes adicionales

Otros números de identificación del estudio

  • 124039
  • IRB-17-155 (Otro identificador: Worcester Polytechnic Institute)
  • 5R42HD076519-03 (Subvención/contrato del NIH de EE. UU.)

Información sobre medicamentos y dispositivos, documentos del estudio

Estudia un producto farmacéutico regulado por la FDA de EE. UU.

No

Estudia un producto de dispositivo regulado por la FDA de EE. UU.

No

producto fabricado y exportado desde los EE. UU.

No

Esta información se obtuvo directamente del sitio web clinicaltrials.gov sin cambios. Si tiene alguna solicitud para cambiar, eliminar o actualizar los detalles de su estudio, comuníquese con register@clinicaltrials.gov. Tan pronto como se implemente un cambio en clinicaltrials.gov, también se actualizará automáticamente en nuestro sitio web. .

Suscribir