EMG-Based Hand-Wrist Control: Study B Mirrored (PSICON-B)

30. Dezember 2020 aktualisiert von: Liberating Technologies, Inc.

EMG-Based Hand-Wrist Control: Study B: Comparison of Contralateral (Mirrored) EMG-

Study assessing four-channel prosthesis controller, that compares contralateral (mirrored) EMG-force training to ipsilateral EMG-target training with both limb-absent and able-bodied subjects

Studienübersicht

Detaillierte Beschreibung

System identification models relating forearm electromyogram (EMG) signals to phantom wrist radial-ulnar deviation force, pronation-supination moment and/or hand open-close force (EMG-force) are hampered by lack of supervised force/moment output signals in limb-absent subjects. In able-bodied and unilateral transradial limb-absent subjects, we studied three alternative supervised output sources in one degree of freedom (DoF) and 2-DoF target tracking tasks: (1) bilateral tracking with force feedback from the contralateral side (non-dominant for able-bodied/ sound for limb-absent subjects) with the contralateral force as the output, (2) bilateral tracking with force feedback from the contralateral side with the target as the output, and (3) dominant/limb-absent side unilateral target tracking without feedback and the target used as the output.

Studientyp

Interventionell

Einschreibung (Tatsächlich)

20

Phase

  • Frühphase 1

Kontakte und Standorte

Dieser Abschnitt enthält die Kontaktdaten derjenigen, die die Studie durchführen, und Informationen darüber, wo diese Studie durchgeführt wird.

Studienorte

    • Massachusetts
      • Holliston, Massachusetts, Vereinigte Staaten, 01746
        • Liberating Technologies, Inc
      • Worcester, Massachusetts, Vereinigte Staaten, 01609
        • Worcester Polytechnic Institute

Teilnahmekriterien

Forscher suchen nach Personen, die einer bestimmten Beschreibung entsprechen, die als Auswahlkriterien bezeichnet werden. Einige Beispiele für diese Kriterien sind der allgemeine Gesundheitszustand einer Person oder frühere Behandlungen.

Zulassungskriterien

Studienberechtigtes Alter

18 Jahre und älter (Erwachsene, Älterer Erwachsener)

Akzeptiert gesunde Freiwillige

Nein

Studienberechtigte Geschlechter

Alle

Beschreibung

Inclusion Criteria:

  • Limb-absent subjects: unilateral trans-radial limb absence or amputation
  • Be capable of completing the requested contractions on the affected dominant side

Exclusion Criteria:

  • Past injuries to the upper limbs that would limit their ability to complete the requested contractions
  • Scars that would impede the use of surface electrodes

Studienplan

Dieser Abschnitt enthält Einzelheiten zum Studienplan, einschließlich des Studiendesigns und der Messung der Studieninhalte.

Wie ist die Studie aufgebaut?

Designdetails

  • Hauptzweck: Grundlegende Wissenschaft
  • Zuteilung: Nicht randomisiert
  • Interventionsmodell: Sequenzielle Zuweisung
  • Maskierung: Keine (Offenes Etikett)

Waffen und Interventionen

Teilnehmergruppe / Arm
Intervention / Behandlung
Experimental: Study B: Limb-Absent Subject
Ten unilateral transradial limb-absent subjects will each participate in one, half-day experimental session. Subjects will be seated and prepared in test apparatus seat (16 electrodes on the affected side, hand-wrist on the able side secured to load cells). Subjects will complete the 1-DoF and 2-DoF "dynamic" (force-varying) contractions (40-s duration, 0.75 Hz bandlimited, uniform random target). With 2 DoF contraction trials, hand Opn-Cls will always be one of the dimensions. The subject will then be released from the cuff and their able side not further involved in the experiment. The force feedback triangle cursor on the computer screen will be deleted such that only the target remains. Subjects will then repeat 1-DoF and 2-DoF trials in which the affected side attempts to produce hand-wrist effort that mimic movement of the target (with no feedback provided).
Test apparatus acquires 16 sEMG channels, measure four DoFs of force/moment at the hand-wrist [hand open-close (Opn-Cls); wrist extension-flexion (Ext-Flx), radial-ulnar deviation (Rad-Uln) and pronation-supination (Pro-Sup)]. sEMG system used 16 custom encased bipolar electrodes. A commercial LTI EMG amplifier (BE328) conditioned each EMG signal before A/D conversion on a PC. A thermoplastic hand cuff secured the wrist to a six-axis load cell (AMTI, model MC3A-100) for force/moment measurement. A separate hand-grasp one-axis load cell (Omega Engineering Inc., model LCR-150) was Velcro-secured between the fingers and the thumb, to measure power grip forces. The PC acquired and stored the sEMG and load cell data, and commanded a triangular screen target. The subject controlled a second triangular cursor. Left-right movement of the cursor was controlled by wrist Ext-Flx, up-down movement by Rad-Uln deviation, cursor rotation by Pro-Sup and cursor size by hand Opn-Cls forces.
Andere Namen:
  • load cell
  • electrode array
Experimental: Study B: Able-bodied Subjects
Ten able-bodied subjects, the electrodes will be mounted on the dominant arm for the EMG-force and EMG-target tracking trials. In addition, the non-dominant arm will also be constrained and measured in a second load cell. This load cell will not be used for feedback during the experiment, but will compare (RMS error, off-line) the dominant vs. non-dominant forces. For EMG-target tracking, the dominant hand will remain in the wrist cuff (to prevent flailing during contractions), with the screen feedback disabled. These subjects will repeat the trials with the electrodes moved to the non-dominant side. EMG-force tracking will be repeated using mirrored contractions. The three training methods (EMG-force ipsilateral, EMG-force contralateral mirrored, EMG-target on the dominant side) will be contrasted to help understand the source of errors when training with limb-absent subjects.
Test apparatus acquires 16 sEMG channels, measure four DoFs of force/moment at the hand-wrist [hand open-close (Opn-Cls); wrist extension-flexion (Ext-Flx), radial-ulnar deviation (Rad-Uln) and pronation-supination (Pro-Sup)]. sEMG system used 16 custom encased bipolar electrodes. A commercial LTI EMG amplifier (BE328) conditioned each EMG signal before A/D conversion on a PC. A thermoplastic hand cuff secured the wrist to a six-axis load cell (AMTI, model MC3A-100) for force/moment measurement. A separate hand-grasp one-axis load cell (Omega Engineering Inc., model LCR-150) was Velcro-secured between the fingers and the thumb, to measure power grip forces. The PC acquired and stored the sEMG and load cell data, and commanded a triangular screen target. The subject controlled a second triangular cursor. Left-right movement of the cursor was controlled by wrist Ext-Flx, up-down movement by Rad-Uln deviation, cursor rotation by Pro-Sup and cursor size by hand Opn-Cls forces.
Andere Namen:
  • load cell
  • electrode array

Was misst die Studie?

Primäre Ergebnismessungen

Ergebnis Maßnahme
Maßnahmenbeschreibung
Zeitfenster
EMG-Force Models
Zeitfenster: 1 day visit

Electromyogram waveform measurements (signal amplitude, frequency) will be related to force measurements (magnitude and moment) as related to the target tracking tasks. These signal-force data will be use to generate the models.

The specific outcome for each subject will be the RMS error measured in the tracking tasks as the difference between the generated control signal from the user versus the prompted target signal, as measured by an EMG array.

1 day visit

Mitarbeiter und Ermittler

Hier finden Sie Personen und Organisationen, die an dieser Studie beteiligt sind.

Studienaufzeichnungsdaten

Diese Daten verfolgen den Fortschritt der Übermittlung von Studienaufzeichnungen und zusammenfassenden Ergebnissen an ClinicalTrials.gov. Studienaufzeichnungen und gemeldete Ergebnisse werden von der National Library of Medicine (NLM) überprüft, um sicherzustellen, dass sie bestimmten Qualitätskontrollstandards entsprechen, bevor sie auf der öffentlichen Website veröffentlicht werden.

Haupttermine studieren

Studienbeginn (Tatsächlich)

29. Juni 2017

Primärer Abschluss (Tatsächlich)

24. August 2020

Studienabschluss (Tatsächlich)

24. August 2020

Studienanmeldedaten

Zuerst eingereicht

28. Dezember 2020

Zuerst eingereicht, das die QC-Kriterien erfüllt hat

30. Dezember 2020

Zuerst gepostet (Tatsächlich)

5. Januar 2021

Studienaufzeichnungsaktualisierungen

Letztes Update gepostet (Tatsächlich)

5. Januar 2021

Letztes eingereichtes Update, das die QC-Kriterien erfüllt

30. Dezember 2020

Zuletzt verifiziert

1. Dezember 2020

Mehr Informationen

Begriffe im Zusammenhang mit dieser Studie

Zusätzliche relevante MeSH-Bedingungen

Andere Studien-ID-Nummern

  • 124039
  • IRB-17-155 (Andere Kennung: Worcester Polytechnic Institute)
  • 5R42HD076519-03 (US NIH Stipendium/Vertrag)

Arzneimittel- und Geräteinformationen, Studienunterlagen

Studiert ein von der US-amerikanischen FDA reguliertes Arzneimittelprodukt

Nein

Studiert ein von der US-amerikanischen FDA reguliertes Geräteprodukt

Nein

Produkt, das in den USA hergestellt und aus den USA exportiert wird

Nein

Diese Informationen wurden ohne Änderungen direkt von der Website clinicaltrials.gov abgerufen. Wenn Sie Ihre Studiendaten ändern, entfernen oder aktualisieren möchten, wenden Sie sich bitte an register@clinicaltrials.gov. Sobald eine Änderung auf clinicaltrials.gov implementiert wird, wird diese automatisch auch auf unserer Website aktualisiert .

Klinische Studien zur Amputation

Klinische Studien zur PSICON Measurement Apparatus

3