EMG-Based Hand-Wrist Control: Study B Mirrored (PSICON-B)

keskiviikko 30. joulukuuta 2020 päivittänyt: Liberating Technologies, Inc.

EMG-Based Hand-Wrist Control: Study B: Comparison of Contralateral (Mirrored) EMG-

Study assessing four-channel prosthesis controller, that compares contralateral (mirrored) EMG-force training to ipsilateral EMG-target training with both limb-absent and able-bodied subjects

Tutkimuksen yleiskatsaus

Yksityiskohtainen kuvaus

System identification models relating forearm electromyogram (EMG) signals to phantom wrist radial-ulnar deviation force, pronation-supination moment and/or hand open-close force (EMG-force) are hampered by lack of supervised force/moment output signals in limb-absent subjects. In able-bodied and unilateral transradial limb-absent subjects, we studied three alternative supervised output sources in one degree of freedom (DoF) and 2-DoF target tracking tasks: (1) bilateral tracking with force feedback from the contralateral side (non-dominant for able-bodied/ sound for limb-absent subjects) with the contralateral force as the output, (2) bilateral tracking with force feedback from the contralateral side with the target as the output, and (3) dominant/limb-absent side unilateral target tracking without feedback and the target used as the output.

Opintotyyppi

Interventio

Ilmoittautuminen (Todellinen)

20

Vaihe

  • Varhainen vaihe 1

Yhteystiedot ja paikat

Tässä osiossa on tutkimuksen suorittajien yhteystiedot ja tiedot siitä, missä tämä tutkimus suoritetaan.

Opiskelupaikat

    • Massachusetts
      • Holliston, Massachusetts, Yhdysvallat, 01746
        • Liberating Technologies, Inc
      • Worcester, Massachusetts, Yhdysvallat, 01609
        • Worcester Polytechnic Institute

Osallistumiskriteerit

Tutkijat etsivät ihmisiä, jotka sopivat tiettyyn kuvaukseen, jota kutsutaan kelpoisuuskriteereiksi. Joitakin esimerkkejä näistä kriteereistä ovat henkilön yleinen terveydentila tai aiemmat hoidot.

Kelpoisuusvaatimukset

Opintokelpoiset iät

18 vuotta ja vanhemmat (Aikuinen, Vanhempi Aikuinen)

Hyväksyy terveitä vapaaehtoisia

Ei

Sukupuolet, jotka voivat opiskella

Kaikki

Kuvaus

Inclusion Criteria:

  • Limb-absent subjects: unilateral trans-radial limb absence or amputation
  • Be capable of completing the requested contractions on the affected dominant side

Exclusion Criteria:

  • Past injuries to the upper limbs that would limit their ability to complete the requested contractions
  • Scars that would impede the use of surface electrodes

Opintosuunnitelma

Tässä osiossa on tietoja tutkimussuunnitelmasta, mukaan lukien kuinka tutkimus on suunniteltu ja mitä tutkimuksella mitataan.

Miten tutkimus on suunniteltu?

Suunnittelun yksityiskohdat

  • Ensisijainen käyttötarkoitus: Perustiede
  • Jako: Ei satunnaistettu
  • Inventiomalli: Peräkkäinen tehtävä
  • Naamiointi: Ei mitään (avoin tarra)

Aseet ja interventiot

Osallistujaryhmä / Arm
Interventio / Hoito
Kokeellinen: Study B: Limb-Absent Subject
Ten unilateral transradial limb-absent subjects will each participate in one, half-day experimental session. Subjects will be seated and prepared in test apparatus seat (16 electrodes on the affected side, hand-wrist on the able side secured to load cells). Subjects will complete the 1-DoF and 2-DoF "dynamic" (force-varying) contractions (40-s duration, 0.75 Hz bandlimited, uniform random target). With 2 DoF contraction trials, hand Opn-Cls will always be one of the dimensions. The subject will then be released from the cuff and their able side not further involved in the experiment. The force feedback triangle cursor on the computer screen will be deleted such that only the target remains. Subjects will then repeat 1-DoF and 2-DoF trials in which the affected side attempts to produce hand-wrist effort that mimic movement of the target (with no feedback provided).
Test apparatus acquires 16 sEMG channels, measure four DoFs of force/moment at the hand-wrist [hand open-close (Opn-Cls); wrist extension-flexion (Ext-Flx), radial-ulnar deviation (Rad-Uln) and pronation-supination (Pro-Sup)]. sEMG system used 16 custom encased bipolar electrodes. A commercial LTI EMG amplifier (BE328) conditioned each EMG signal before A/D conversion on a PC. A thermoplastic hand cuff secured the wrist to a six-axis load cell (AMTI, model MC3A-100) for force/moment measurement. A separate hand-grasp one-axis load cell (Omega Engineering Inc., model LCR-150) was Velcro-secured between the fingers and the thumb, to measure power grip forces. The PC acquired and stored the sEMG and load cell data, and commanded a triangular screen target. The subject controlled a second triangular cursor. Left-right movement of the cursor was controlled by wrist Ext-Flx, up-down movement by Rad-Uln deviation, cursor rotation by Pro-Sup and cursor size by hand Opn-Cls forces.
Muut nimet:
  • load cell
  • electrode array
Kokeellinen: Study B: Able-bodied Subjects
Ten able-bodied subjects, the electrodes will be mounted on the dominant arm for the EMG-force and EMG-target tracking trials. In addition, the non-dominant arm will also be constrained and measured in a second load cell. This load cell will not be used for feedback during the experiment, but will compare (RMS error, off-line) the dominant vs. non-dominant forces. For EMG-target tracking, the dominant hand will remain in the wrist cuff (to prevent flailing during contractions), with the screen feedback disabled. These subjects will repeat the trials with the electrodes moved to the non-dominant side. EMG-force tracking will be repeated using mirrored contractions. The three training methods (EMG-force ipsilateral, EMG-force contralateral mirrored, EMG-target on the dominant side) will be contrasted to help understand the source of errors when training with limb-absent subjects.
Test apparatus acquires 16 sEMG channels, measure four DoFs of force/moment at the hand-wrist [hand open-close (Opn-Cls); wrist extension-flexion (Ext-Flx), radial-ulnar deviation (Rad-Uln) and pronation-supination (Pro-Sup)]. sEMG system used 16 custom encased bipolar electrodes. A commercial LTI EMG amplifier (BE328) conditioned each EMG signal before A/D conversion on a PC. A thermoplastic hand cuff secured the wrist to a six-axis load cell (AMTI, model MC3A-100) for force/moment measurement. A separate hand-grasp one-axis load cell (Omega Engineering Inc., model LCR-150) was Velcro-secured between the fingers and the thumb, to measure power grip forces. The PC acquired and stored the sEMG and load cell data, and commanded a triangular screen target. The subject controlled a second triangular cursor. Left-right movement of the cursor was controlled by wrist Ext-Flx, up-down movement by Rad-Uln deviation, cursor rotation by Pro-Sup and cursor size by hand Opn-Cls forces.
Muut nimet:
  • load cell
  • electrode array

Mitä tutkimuksessa mitataan?

Ensisijaiset tulostoimenpiteet

Tulosmittaus
Toimenpiteen kuvaus
Aikaikkuna
EMG-Force Models
Aikaikkuna: 1 day visit

Electromyogram waveform measurements (signal amplitude, frequency) will be related to force measurements (magnitude and moment) as related to the target tracking tasks. These signal-force data will be use to generate the models.

The specific outcome for each subject will be the RMS error measured in the tracking tasks as the difference between the generated control signal from the user versus the prompted target signal, as measured by an EMG array.

1 day visit

Yhteistyökumppanit ja tutkijat

Täältä löydät tähän tutkimukseen osallistuvat ihmiset ja organisaatiot.

Opintojen ennätyspäivät

Nämä päivämäärät seuraavat ClinicalTrials.gov-sivustolle lähetettyjen tutkimustietueiden ja yhteenvetojen edistymistä. National Library of Medicine (NLM) tarkistaa tutkimustiedot ja raportoidut tulokset varmistaakseen, että ne täyttävät tietyt laadunvalvontastandardit, ennen kuin ne julkaistaan ​​julkisella verkkosivustolla.

Opi tärkeimmät päivämäärät

Opiskelun aloitus (Todellinen)

Torstai 29. kesäkuuta 2017

Ensisijainen valmistuminen (Todellinen)

Maanantai 24. elokuuta 2020

Opintojen valmistuminen (Todellinen)

Maanantai 24. elokuuta 2020

Opintoihin ilmoittautumispäivät

Ensimmäinen lähetetty

Maanantai 28. joulukuuta 2020

Ensimmäinen toimitettu, joka täytti QC-kriteerit

Keskiviikko 30. joulukuuta 2020

Ensimmäinen Lähetetty (Todellinen)

Tiistai 5. tammikuuta 2021

Tutkimustietojen päivitykset

Viimeisin päivitys julkaistu (Todellinen)

Tiistai 5. tammikuuta 2021

Viimeisin lähetetty päivitys, joka täytti QC-kriteerit

Keskiviikko 30. joulukuuta 2020

Viimeksi vahvistettu

Tiistai 1. joulukuuta 2020

Lisää tietoa

Tähän tutkimukseen liittyvät termit

Muita asiaankuuluvia MeSH-ehtoja

Muut tutkimustunnusnumerot

  • 124039
  • IRB-17-155 (Muu tunniste: Worcester Polytechnic Institute)
  • 5R42HD076519-03 (Yhdysvaltain NIH-apuraha/sopimus)

Lääke- ja laitetiedot, tutkimusasiakirjat

Tutkii yhdysvaltalaista FDA sääntelemää lääkevalmistetta

Ei

Tutkii yhdysvaltalaista FDA sääntelemää laitetuotetta

Ei

Yhdysvalloissa valmistettu ja sieltä viety tuote

Ei

Nämä tiedot haettiin suoraan verkkosivustolta clinicaltrials.gov ilman muutoksia. Jos sinulla on pyyntöjä muuttaa, poistaa tai päivittää tutkimustietojasi, ota yhteyttä register@clinicaltrials.gov. Heti kun muutos on otettu käyttöön osoitteessa clinicaltrials.gov, se päivitetään automaattisesti myös verkkosivustollemme .