The 4-SITE Field Following Study (4-SITE FF) (4-SITE FF)
RELIANCE QUADRIPOLAR DEFIBRILLATION LEADS (4-SITE)FIELD FOLLOWING The 4-SITE Field Following Study (4-SITE FF)
The purpose of this study is to evaluate and document appropriate clinical performance of the new 4-SITE Header / Lead interface in the TELIGEN 100 HE 4-SITE (VR and DR) implantable cardioverter defibrillator (ICD) PGs (Models F103; F111), the COGNIS 100 HE 4-SITE cardiac resynchronization therapy ICD (CRT-D) PG (Model P108) when connected to any of the RELIANCE Quadripolar (4-SITE) defibrillation leads.
Appropriate clinical performance of system components of the 4-SITE systems other than the new 4-SITE header / lead interface has been evaluated and documented already in the previously conducted COGNIS and TELIGEN 100 HE FIELD FOLLOWING STUDY, The COGENT-4 Field Following Study.
Data collected may be used to support international regulatory submissions including the FDA and the Japanese Ministry of Health, Labour and Welfare.
A tanulmány áttekintése
Állapot
Körülmények
Részletes leírás
The following devices will be used (all are CE-marked):
TELIGEN 100 HE 4-SITE (Single and Dual Chamber ICD)
- Model Number: F103 (Quadripolar, VR);
- Model Number: F111 (Quadripolar, IS-1, DR)
COGNIS 100 HE 4-SITE (CRT-D)
o Model Number: P108 (Quadripolar, IS-1, IS-1)
- RELIANCE 4-SITE leads (12 different model numbers in total as shown in the table below)
Tanulmány típusa
Beiratkozás (Tényleges)
Kapcsolatok és helyek
Tanulmányi helyek
-
-
Nordrhein-Westfalen
-
Bad Oeynhausen, Nordrhein-Westfalen, Németország, D - 32545
- Bad Oeynhausen Herz- und Diabeteszentrum Nordrhein-Westfalen
-
-
Részvételi kritériumok
Jogosultsági kritériumok
Tanulmányozható életkorok
Egészséges önkénteseket fogad
Tanulmányozható nemek
Mintavételi módszer
Tanulmányi populáció
Leírás
Inclusion Criteria:
Study Specific:
- ICD / CRT-D Indication according to normal clinical practice
- Patients receiving:
- a single or dual chamber 4-SITE compatible ICD
- or a 4-SITE compatible CRT-D
- one of the RELIANCE 4-SITE defibrillation leads
- Patients currently implanted with a pacemaker
- upgraded to a 4-SITE compatible ICD or CRT-D (4-SITE header)
- one of the RELIANCE 4-SITE defibrillation leads
General:
- Willing and capable of providing informed consent for
- undergoing a 4-SITE system implant,
- participating in all testing associated with this clinical investigation at an approved clinical investigational centre and at the intervals defined by this protocol
- Geographically stable patients who are available for follow-up at a study centre
- Age 18 or above, or of legal age to give informed consent specific to national law
Exclusion Criteria:
Study Specific:
- ICD and CRT-D Patients scheduled for a device replacement
- CRM Patients who have or who would need an lead adaptor
- All patients who have an active or non-active defibrillation lead other than 4-SITE
General:
- Not willing and not capable of providing informed consent, undergoing a device implant, participating in all testing associated with this clinical investigation (including VT/VF shock conversion) at an approved clinical investigational centre and at the intervals defined by this protocol
- Patients who were in NYHA Class IV during the last 3 month
- Patients with pre-existing diseases, which may confound study results
- Patients currently requiring dialysis,
- Cancer patients
- Patients with drug and/or alcohol abuse history
- Life expectancy < 12 months (or expected heart transplant within 12 months)
- Patients on a Heart Transplant List
- Women who are pregnant or plan to become pregnant. Method of assessment per physician discretion.
- Enrolled in any other concurrent study
Tanulási terv
Hogyan készül a tanulmány?
Tervezési részletek
- Időperspektívák: Leendő
Mit mér a tanulmány?
Elsődleges eredményintézkedések
Eredménymérő |
Intézkedés leírása |
Időkeret |
|---|---|---|
|
Appropriate Detection of induced VT
Időkeret: At Implant or during VT/VF testing up to one month after implant
|
Induced mean VT/VF detection time < 4.5 seconds
|
At Implant or during VT/VF testing up to one month after implant
|
|
Appropriate induced VT/VF Shock Conversion
Időkeret: At Implant or during testing within one month after implant
|
Successful VT/VF conversions (with 41 J, no external shock) will be 93% or higher
|
At Implant or during testing within one month after implant
|
|
Appropriate pacing thresholds at follow-up with the 4-SITE defibrillation leads
Időkeret: entire duration of the study
|
Mean pacing threshold post implant ≤ 1.5V
|
entire duration of the study
|
|
Appropriate Lead Impedances as a measure of lead integrity over 12 month time
Időkeret: entire study duration
|
|
entire study duration
|
|
Appropriate sensing and absence of artefacts
Időkeret: entire duration of study
|
Appropriate sensing and absence of artefacts during standardized pocket manipulations and standardized provocative maneuvers during FU
|
entire duration of study
|
Együttműködők és nyomozók
Szponzor
Tanulmányi rekorddátumok
Tanulmány főbb dátumok
Tanulmány kezdete
Elsődleges befejezés (Tényleges)
A tanulmány befejezése (Tényleges)
Tanulmányi regisztráció dátumai
Először benyújtva
Először nyújtották be, amely megfelel a minőségbiztosítási kritériumoknak
Első közzététel (Becslés)
Tanulmányi rekordok frissítései
Utolsó frissítés közzétéve (Becslés)
Az utolsó frissítés elküldve, amely megfelel a minőségbiztosítási kritériumoknak
Utolsó ellenőrzés
Több információ
A tanulmányhoz kapcsolódó kifejezések
Kulcsszavak
Egyéb vizsgálati azonosító számok
- 4-SITE FF 0209
Ezt az információt közvetlenül a clinicaltrials.gov webhelyről szereztük be, változtatás nélkül. Ha bármilyen kérése van vizsgálati adatainak módosítására, eltávolítására vagy frissítésére, kérjük, írjon a következő címre: register@clinicaltrials.gov. Amint a változás bevezetésre kerül a clinicaltrials.gov oldalon, ez a webhelyünkön is automatikusan frissül. .